Quick summary

This Stryker Prime wheeled stretcher has 80 adverse event reports, no deaths, 4 injuries and 76 malfunctions. Reports mostly describe mechanical jams, brake or steering problems, fluid leaks, mattress slipping or positioning failures, and exposed sharp edges. No recalls name this device. Reports fell from 8 to 4 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
80
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 80 adverse event reports that list this exact device (DI 07613327278231), received between Apr 23, 2018 and Apr 23, 2026 — classified as 4 injuries, 76 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (34 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 4 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam34
  • Fluid/Blood Leak12
  • Mechanical Problem12
  • Device Slipped6
  • Positioning Failure6
  • Sharp Edges5
  • Patient Device Interaction Problem2
  • Device Fell1

Patient problems

  • No Clinical Signs, Symptoms or Conditions51
  • No Patient Involvement20
  • Injury3
  • No Consequences Or Impact To Patient3
  • Bruise/Contusion2
  • Pain2
  • Fall1

Recent adverse event reports

The 25 most recent of 80 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2026Mechanical Jam

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24965132 ↗
  • MalfunctionApr 23, 2026Sharp Edges

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP EDGES EXPOSED (METAL). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24965128 ↗
  • MalfunctionJan 26, 2026Device Slipped

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED MATTRESS SLIDES OFF LITTER; UNSTABLE MATTRESS SUPPORT . NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24180202 ↗
  • MalfunctionJan 26, 2026Mechanical Jam

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24178590 ↗
  • MalfunctionOct 27, 2025Mechanical Jam; Mechanical Jam

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393622 ↗
  • MalfunctionOct 27, 2025Sharp Edges; Sharp Edges

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP EDGES EXPOSED (METAL). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23393060 ↗
  • MalfunctionApr 25, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21912123 ↗
  • MalfunctionApr 25, 2025Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES 22 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES DIFFICULT TO ENGAGE/DISENGAGE OR STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21911580 ↗
  • MalfunctionApr 25, 2025Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES DO NOT REMAIN ENGAGED. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21911540 ↗
  • MalfunctionJan 30, 2025Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE LATCH OF THE SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21273856 ↗
  • MalfunctionJan 27, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FLUID LEAKING ONTO THE FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21241355 ↗
  • MalfunctionJan 27, 2025Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED BRAKES OR STEER MECHANISMS WHICH WERE DIFFICULT TO ENGAGE OR DISENGAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21241352 ↗
  • MalfunctionOct 29, 2024Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES 16 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE, BRAKES DIFFICULT TO ENGAGE/DISENGAGE OR BRAKES CANNOT BE ENGAGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20558727 ↗
  • MalfunctionOct 25, 2024Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20538960 ↗
  • MalfunctionJul 25, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19832300 ↗
  • MalfunctionJul 25, 2024Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19832210 ↗
  • MalfunctionJul 25, 2024Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES CANNOT BE ENGAGED OR STEER MECHANISM IS DIFFICULT TO ENGAGE/DISENGAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19832155 ↗
  • MalfunctionJul 25, 2024Device Slipped; Device Slipped

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED MATTRESS SLIDES OFF LITTER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19832151 ↗
  • MalfunctionApr 22, 2024Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES (B)(4) MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES DIFFICULT TO ENGAGE/DISENGAGE OR BRAKES CANNOT BE ENGAGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19156078 ↗
  • MalfunctionApr 22, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19155156 ↗
  • MalfunctionApr 22, 2024Sharp Edges; Sharp Edges

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19154950 ↗
  • MalfunctionJan 24, 2024Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 0 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18576165 ↗
  • MalfunctionJan 23, 2024Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED BRAKES DIFFICULT TO ENGAGE/DISENGAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18563158 ↗
  • MalfunctionJan 23, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 11 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18563156 ↗
  • MalfunctionOct 25, 2023Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18007349 ↗

Other STRETCHER, WHEELED devices