Stryker Corporation · Class 2

Quick summary

This Stryker wheeled stretcher is a Class II device used to move patients. Its reports mostly describe mechanical problems: parts jamming, the mattress or positioning failing, the stretcher tipping or falling, and fluid leaking onto the floor, along with side rails that latch incorrectly and brakes that fail to engage. There are 75 reports, 1 death and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
75
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 75 adverse event reports that list this exact device (DI 07613327278262), received between Apr 23, 2018 and Jan 23, 2026 — classified as 1 death, 74 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 4 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 4 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 4 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam18
  • Device Fell17
  • Positioning Failure17
  • Fluid/Blood Leak12
  • Mechanical Problem4
  • Sharp Edges4
  • Device Slipped1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions59
  • No Patient Involvement15
  • Insufficient Information2
  • Fall1
  • Limb Fracture1
  • Pain1

Recent adverse event reports

The 25 most recent of 75 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Device Slipped

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED MATTRESS SLIDES OFF LITTER; UNSTABLE MATTRESS SUPPORT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24168583 ↗
  • MalfunctionOct 24, 2025Positioning Failure; Positioning Failure

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23383839 ↗
  • MalfunctionOct 24, 2025Mechanical Jam; Mechanical Jam

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23383716 ↗
  • MalfunctionApr 25, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WERE NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21909427 ↗
  • MalfunctionJan 29, 2025Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE LATCH OF THE SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21267021 ↗
  • MalfunctionJan 24, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 11 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKING ONTO THE FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21229475 ↗
  • MalfunctionOct 31, 2024Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE CUSTOMER IS HAVING ISSUES LOCKING THE BRAKE WHICH CAN CAUSE THE BRAKES TO NOT BE ENGAGED; FALSE ENGAGEMENT OF BRAKES. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20579457 ↗
  • MalfunctionOct 25, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20537539 ↗
  • MalfunctionOct 25, 2024Device Fell; Device Fell

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20535271 ↗
  • MalfunctionOct 25, 2024Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20533770 ↗
  • MalfunctionJul 24, 2024Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19819169 ↗
  • MalfunctionJul 24, 2024Mechanical Problem; Mechanical Problem

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19819150 ↗
  • MalfunctionJul 24, 2024Device Fell; Device Fell

    THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19819144 ↗
  • MalfunctionJul 24, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19819143 ↗
  • MalfunctionApr 23, 2024Device Fell; Device Fell

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19163286 ↗
  • MalfunctionApr 22, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; NO ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19155522 ↗
  • MalfunctionJan 24, 2024Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18573903 ↗
  • MalfunctionJan 22, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18552263 ↗
  • MalfunctionJan 22, 2024Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED BRAKES CANNOT BE ENGAGED OR BRAKES DIFFICULT TO ENGAGE/DISENGAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18552256 ↗
  • MalfunctionJan 22, 2024Device Fell; Device Fell

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SIDERAIL AND FOWLER UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18552244 ↗
  • MalfunctionOct 24, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17998810 ↗
  • MalfunctionOct 24, 2023Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17998808 ↗
  • MalfunctionOct 24, 2023Device Fell; Device Fell

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17998769 ↗
  • MalfunctionJul 27, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FLUID LEAKS ONTO FLOOR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17407268 ↗
  • MalfunctionJul 24, 2023Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES DIFFICULT TO ENGAGE OR BRAKES CANNOT BE ENGAGED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17379339 ↗