Skip to content

T2 ALPHA 2333-0932SRecalls & reported problems

Stryker GmbH · Femoral Nail GT, RIGHT · Class 2 · Implant

All models of T2 ALPHA (123)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327286489), received between Feb 1, 2022 and Jul 2, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Fracture2
  • Positioning Failure1

Patient problems

  • Failure of Implant2
  • No Clinical Signs, Symptoms or Conditions1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 2, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED, ORIF WAS PERFORMED ON (B)(6) 2023 AND THE T2 ALPHA GT NAIL WAS IMPLANTED IN THE RIGHT FEMUR OF THE PATIENT. ON (B)(6) 2024, THE NAIL BREAKAGE WAS CONFIRMED. THE DOCTOR'S OPINION IS THAT THE NAIL WAS BROKEN BECAUSE THE LOAD MAY HAVE BEEN APPLIED ON THE NAIL DUE TO THE PSEUDARTHROSIS. THE REVISION SURGERY WILL BE PERFORMED ON (B)(6) 2024.

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 19652778 ↗
  • InjurySep 8, 2022Break; Fracture; Break; Fracture

    AS REPORTED: "(B)(6) 2022 ORIF WITH T2ALPHAGT NAIL WAS PERFORMED. 3 MONTHS AFTER SURGERY, THE NAIL WAS BROKEN. ON (B)(6) 2022, THE BROKEN NAIL WAS REMOVED AND THE FRACTURE WAS RE-FIXED WITH T2ALPHAGT. THE SURGEON'S OPINION WAS THAT THE INITIAL SURGERY NAIL WAS TOO THIN AND SHORT FOR THE PATIENT'S BONE, WHICH MAY HAVE LED TO THE IMPLANT FRACTURE."

    Patient problems: Failure of Implant

    MAUDE report 15377749 ↗
  • MalfunctionFeb 1, 2022Positioning Failure; Positioning Failure

    AS REPORTED: "DURING VERIFYING CORRECT ALIGNMENT BY INSERTING A DRILL THROUGH THE SLEEVE AND TARGETING ARM ASSEMBLY, THE DRILL DID NOT PASS THROUGH THE APPROPRIATE HOLE OF THE NAIL (FIRST NAIL). AFTER THAT SOME ADJUSTMENT WAS DONE BUT THE ISSUE DID NOT SOLVE. THE PHYSICIAN CHANGED TO NEW DIFFERENT SIZE NAIL (SECOND NAIL) AND THE VERIFICATION OF CORRECT ALIGNMENT WAS SUCCESSFUL. THE PROCEDURE WAS COMPLETED WITHOUT PROBLEM USING SECOND NAIL WHICH WAS DIFFERENT SIZE AS THE PLANED SIZE NAIL."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13418978 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates