Triathlon 5537-G-213-ERecalls & reported problems
Howmedica Osteonics Corp. · Total Stabilizer+ Tibial Insert · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327287967), received between May 8, 2024 and Dec 18, 2025 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Mechanical Problem” (2 reports).
7 other recalls cover devices cleared under the same FDA submission (K172634); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Mechanical Problem2
- Detachment of Device or Device Component1
Patient problems
- Loss of Range of Motion2
- Fall1
- No Clinical Signs, Symptoms or Conditions1
- Rupture1
- Scar Tissue1
- Unspecified Infection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryDec 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS PROVIDED: RIGHT TOTAL KNEE. REVISION SECONDARY TO A FALL AND PATELLA TENDON RUPTURE. THE INSERT WAS REVISED TO A 19MM TO REGAIN STABILITY LOST DUE TO THE INJURY. THE USE OF MAKO DURING THE INDEX PROCEDURE WAS NOT THOUGHT TO HAVE CONTRIBUTED TO THIS COMPLICATION.
Patient problems: Fall; Rupture
MAUDE report 23837029 ↗InjuryJun 30, 2025Mechanical Problem; Mechanical Problem
AS REPORTED: "POLY THICKNESS DOWNSIZED DUE TO LIMITED RANGE OF MOTION. NO OTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY." A 2X13 TS INSERT WAS REVISED TO A 2X9 TS INSERT.
Patient problems: Loss of Range of Motion
MAUDE report 22356847 ↗InjuryMar 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "REV LTKA- SINGLE STAGE REVISION FOR INFECTION." REP CONFIRMED NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 21676340 ↗MalfunctionNov 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
PRIMARY PROCEDURE, RIGHT KNEE. IT WAS REPORTED THAT WHEN THE SURGEON PRESSED THE INSERT INTO THE BASEPLATE, THE LOCKING WIRE APPEARED RAISED. AFTER IMPACTION, THE LOCKING WIRE STILL APPEARED RAISED. ON REMOVAL, IT WAS CONFIRMED THE LOCKING WIRE SEPARATED FROM THE INSERT. INSERT WAS WASTED AND A SECOND IMPLANT WAS OBTAINED. THE SECOND IMPLANT LOCKED INTO THE SAME IN-SITU BASEPLATE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A DELAY OF APPROXIMATELY 1 MINUTE. WASTED IMPLANT WAS DISPOSED BEFORE THE REP COULD OBTAIN PICTURES. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20756199 ↗InjuryMay 8, 2024Mechanical Problem; Mechanical Problem
AS REPORTED: "REVISION LTKA. REVISING POLY TODAY SO DOCTORS CAN REMOVE HO BONE AND TRY TO GET THE PATIENT MORE FLEXION AND A SCAR REVISION. NO HARM TO POLY. 2X13MM TS POLY CAME OUT AND THE A29 PATELLA. 2X9MM TS POLY BACK IN TO HELP PATIENT GET BETTER FLEXION." REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Loss of Range of Motion; Scar Tissue
MAUDE report 19268490 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172634. One submission often covers many models, so check the product and lot details in each record.
- Z-1263-2024 · Feb 6, 2024 · Open, ClassifiedNO 3 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM
- Z-1265-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MM
- Z-1262-2024 · Feb 6, 2024 · Open, ClassifiedX3 TRIATHLON CS INSERT NO 7 9 MM
- Z-1261-2024 · Feb 6, 2024 · Open, ClassifiedX3 TRIATHLON CS INSERT NO 6 11 MM
- Z-1266-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MM
- Z-1264-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM
- Z-1654-2023 · Apr 13, 2023 · Open, ClassifiedTRIATHLON X3 TIBIAL BEARING INSERT PS #4 13MM Intended for knee replacement Catalog Number: 5532-G-413-E
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page