NA 4505000000Recalls & reported problems

Stryker Corporation · System 8 Cordless Driver Handpiece · Class 1

Quick summary

This Stryker System 8 cordless driver is a surgical motor handpiece. Reports mostly describe it staying on, leaking fluid or blood, parts detaching or breaking apart, and unintended movement, with no reported injuries or deaths. Of 91 adverse event reports, all were malfunctions. There are no recalls for this device. Reports fell from 17 in the prior 12 months to 1 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
91
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 91 adverse event reports that list this exact device (DI 07613327304459), received between Jul 27, 2018 and Jan 23, 2026 — classified as 91 malfunctions.

The most frequently reported specific device problem is “Device Remains Activated” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 6 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 6 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 5 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 5 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 5 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 4 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Remains Activated25
  • Fluid/Blood Leak24
  • Detachment of Device or Device Component9
  • Material Disintegration8
  • Fail-Safe Did Not Operate7
  • Unintended Movement6
  • Physical Resistance/Sticking4
  • Fracture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions81
  • No Patient Involvement8
  • No Known Impact Or Consequence To Patient3
  • Insufficient Information2
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 25 most recent of 91 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Fluid/Blood Leak

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 0 EVENTS FOR THE FAILURE: FLUID/BLOOD LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163762 ↗
  • MalfunctionOct 28, 2025Device Remains Activated; Device Remains Activated

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 4 EVENTS FOR THE FAILURE: DEVICE REMAINS ACTIVATED. - 4 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399827 ↗
  • MalfunctionOct 28, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: FLUID/BLOOD LEAK - 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399789 ↗
  • MalfunctionOct 23, 2025Unintended Movement; Unintended Movement

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: UNINTENDED MOVEMENT. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368699 ↗
  • MalfunctionOct 23, 2025Material Disintegration; Material Disintegration

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: MATERIAL DISINTEGRATION - 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368693 ↗
  • MalfunctionJul 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22647412 ↗
  • MalfunctionJul 29, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: DEVICE REMAINS ACTIVATED. 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22647046 ↗
  • MalfunctionJul 24, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FAIL-SAFE DID NOT OPERATE. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22586854 ↗
  • MalfunctionApr 21, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 3 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876277 ↗
  • MalfunctionApr 21, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A STICKY SAFETY DEVICE THAT COULD LEAD TO UNINTENDED POWER-UP. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876232 ↗
  • MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21875929 ↗
  • MalfunctionApr 21, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 2 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21875714 ↗
  • MalfunctionApr 9, 2025Fracture; Adverse Event Without Identified Device or Use Problem; Fracture; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE DRILL TIP HIT A NAIL, CAUSING THE TIP OF THE DRILL TO SHEAR OFF AND BECOME EMBEDDED IN THE BONE. ADDITIONAL ATTEMPTS ARE BEING MADE TO THE CUSTOMER FOR ADDITIONAL INFORMATION.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21808670 ↗
  • MalfunctionJan 22, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. 6 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21207533 ↗
  • MalfunctionJan 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21207532 ↗
  • MalfunctionJan 21, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. - 5 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21201056 ↗
  • MalfunctionJan 21, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200915 ↗
  • MalfunctionJan 21, 2025Material Disintegration; Material Disintegration

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200316 ↗
  • MalfunctionOct 22, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 8 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. - 8 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20502882 ↗
  • MalfunctionOct 21, 2024Material Disintegration; Material Disintegration

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. -THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20501311 ↗
  • MalfunctionOct 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20500934 ↗
  • MalfunctionOct 21, 2024Unintended Power Up; Unintended Power Up

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE RAN WITHOUT USER ACTIVATION. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20500440 ↗
  • MalfunctionOct 21, 2024Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. 7 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20497796 ↗
  • MalfunctionJul 24, 2024Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823260 ↗
  • MalfunctionJul 24, 2024Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 9 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. 9 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823141 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →