Stryker Corporation · System 8 Dual Trigger Rotary Handpiece · Class 1

Quick summary

This Stryker dual trigger handpiece is a powered orthopedic surgical tool. Reports describe the handpiece staying activated, parts detaching, fluid or blood leaking in and backup shutoffs not operating, and some note that no patient was involved. There are 114 reports, including 1 injury and no deaths, and reports in the last 12 months fell to 1 from 17. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
114
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 114 adverse event reports that list this exact device (DI 07613327304466), received between Oct 29, 2018 and Jan 23, 2026 — classified as 1 injury, 113 malfunctions.

The most frequently reported specific device problem is “Device Remains Activated” (25 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 5 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 6 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 5 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 4 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 4 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 6 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Remains Activated25
  • Detachment of Device or Device Component23
  • Fluid/Blood Leak23
  • Fail-Safe Problem9
  • Fail-Safe Did Not Operate7
  • Unintended Power Up7
  • Material Disintegration6
  • Material Fragmentation6

Patient problems

  • No Clinical Signs, Symptoms or Conditions98
  • No Patient Involvement12
  • No Consequences Or Impact To Patient5
  • No Known Impact Or Consequence To Patient5
  • Insufficient Information2
  • Damage to Ligament(s)1
  • Limb Fracture1
  • Sprain1

Recent adverse event reports

The 25 most recent of 114 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Fluid/Blood Leak

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FLUID/BLOOD LEAK. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163731 ↗
  • MalfunctionOct 28, 2025Material Disintegration; Material Disintegration

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: MATERIAL DISINTEGRATION. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399967 ↗
  • MalfunctionOct 28, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    NO CHANGE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399921 ↗
  • MalfunctionOct 28, 2025Device Remains Activated; Device Remains Activated

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: DEVICE REMAINS ACTIVATED. - 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399881 ↗
  • MalfunctionOct 28, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FLUID/BLOOD LEAK. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399842 ↗
  • MalfunctionOct 23, 2025Fail-Safe Problem; Fail-Safe Problem

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FAIL-SAFE PROBLEM - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368333 ↗
  • MalfunctionOct 23, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FAIL-SAFE DID NOT OPERATE - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368313 ↗
  • MalfunctionJul 29, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 5 EVENTS FOR THE FAILURE: FLUID/BLOOD LEAK. 5 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22648151 ↗
  • MalfunctionJul 29, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DEVICE REMAINS ACTIVATED. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22647282 ↗
  • MalfunctionJul 24, 2025Mechanical Jam; Mechanical Jam

    THIS REPORT SUMMARIZES 0 EVENTS FOR THE FAILURE: MECHANICAL JAM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587174 ↗
  • MalfunctionJul 24, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) ¿ (B)(6) 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587158 ↗
  • MalfunctionApr 21, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 4 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876276 ↗
  • MalfunctionApr 21, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876240 ↗
  • MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21875950 ↗
  • MalfunctionJan 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 7 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21207662 ↗
  • MalfunctionJan 21, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. - 5 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200911 ↗
  • MalfunctionJan 21, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200756 ↗
  • MalfunctionJan 21, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. -THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200318 ↗
  • MalfunctionOct 22, 2024Material Fragmentation; Material Fragmentation

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD SMALL PIECES BREAKING OFF. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20504581 ↗
  • MalfunctionOct 21, 2024Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20499498 ↗
  • MalfunctionOct 21, 2024Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 5 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20497338 ↗
  • MalfunctionOct 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 12 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 11 EVENTS HAD NO PATIENT INVOLVEMENT: NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20494143 ↗
  • MalfunctionOct 21, 2024Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 8 MALFUNCTION EVENTS IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 8 EVENTS HAD NO PATIENT INVOLVEMENT: NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20493930 ↗
  • MalfunctionAug 1, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE GUIDE PIN IS MISSING FROM THE DEVICE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19884036 ↗
  • MalfunctionJul 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 12 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 11 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823265 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.