NA 8208000000Recalls & reported problems

Stryker Corporation · System 8 Sagittal Saw · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker sagittal saw is a powered bone saw. Reports describe fluid or blood leaks, parts detaching, the saw staying activated and backup shutoffs not operating. There are 114 reports with no deaths or injuries, and reports in the last 12 months fell to 0 from 18. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
114
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 114 adverse event reports that list this exact device (DI 07613327304541), received between Jul 24, 2018 and Oct 23, 2025 — classified as 114 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (27 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 6 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 6 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 6 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 5 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 7 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak27
  • Detachment of Device or Device Component26
  • Device Remains Activated19
  • Fracture11
  • Fail-Safe Did Not Operate10
  • Material Fragmentation8
  • Unintended Power Up4
  • Fail-Safe Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions95
  • No Patient Involvement14
  • No Consequences Or Impact To Patient6
  • Insufficient Information4
  • No Known Impact Or Consequence To Patient3

Recent adverse event reports

The 25 most recent of 114 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 23, 2025Device Remains Activated; Device Remains Activated

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: DEVICE REMAINS ACTIVATED. - 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. - 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368580 ↗
  • MalfunctionOct 23, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FLUID/BLOOD LEAK. 3 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368549 ↗
  • MalfunctionOct 23, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FAIL-SAFE DID NOT OPERATE. - 3 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23368538 ↗
  • MalfunctionJul 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 10 EVENTS FOR THE FAILURE: DETACHMENT OF DEVICE OR DEVICE COMPONENT. 7 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22646738 ↗
  • MalfunctionJul 24, 2025Material Fragmentation; Material Fragmentation

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: MATERIAL FRAGMENTATION. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587166 ↗
  • MalfunctionJul 24, 2025Fracture; Fracture

    "THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN APRIL 1 ¿ JUNE 30, 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: FRACTURE. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587160 ↗
  • MalfunctionJul 24, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 3 EVENTS FOR THE FAILURE: FLUID/BLOOD LEAK. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587116 ↗
  • MalfunctionJul 24, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 5 EVENTS FOR THE FAILURE: FAIL-SAFE DID NOT OPERATE. - 5 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587087 ↗
  • MalfunctionJul 24, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 2 EVENTS FOR THE FAILURE: DEVICE REMAINS ACTIVATED. - 2 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22587051 ↗
  • MalfunctionJul 24, 2025Material Disintegration; Material Disintegration

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: MATERIAL DISINTEGRATION. - 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22586866 ↗
  • MalfunctionApr 21, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 13 MALFUNCTION EVENTS IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. 13 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876285 ↗
  • MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 11 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 3 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 5 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED REGARDING HEALTH IMPACT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876282 ↗
  • MalfunctionApr 21, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 3 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876253 ↗
  • MalfunctionJan 22, 2025Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 30 MALFUNCTION EVENTS IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. - 30 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21208611 ↗
  • MalfunctionJan 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 15 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 6 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 6 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21207667 ↗
  • MalfunctionJan 21, 2025Fluid/Blood Leak; Fluid/Blood Leak

    THIS REPORT SUMMARIZES 38 MALFUNCTION EVENTS IN WHICH THE DEVICE WAS REPORTEDLY LEAKING. 29 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 2 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. 7 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21201047 ↗
  • MalfunctionJan 21, 2025Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE HAD RUN-ON. 1 EVENT HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200916 ↗
  • MalfunctionJan 21, 2025Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21200502 ↗
  • MalfunctionOct 22, 2024Material Disintegration; Material Disintegration

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE WAS SHEDDING METAL DEBRIS. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20502585 ↗
  • MalfunctionOct 21, 2024Fracture; Fracture

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20499734 ↗
  • MalfunctionOct 21, 2024Material Fragmentation; Material Fragmentation

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD SMALL PIECES BREAKING OFF. - 2 EVENTS HAD NO PATIENT INVOLVEMENT: NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20497368 ↗
  • MalfunctionOct 21, 2024Device Remains Activated; Device Remains Activated

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD RUN-ON. - 3 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED REGARDING HEALTH IMPACT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20496998 ↗
  • MalfunctionOct 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 21 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. - 20 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE. -THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD PATIENT INVOLVEMENT, NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20494314 ↗
  • MalfunctionOct 21, 2024Fail-Safe Did Not Operate; Fail-Safe Did Not Operate

    THIS REPORT SUMMARIZES 20 MALFUNCTION EVENTS IN WHICH THE DEVICE EXPERIENCED A FAIL-SAFE PROBLEM THAT COULD RESULT IN UNINTENDED ACTIVATION. - 20 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20494075 ↗
  • MalfunctionJul 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS REPORT SUMMARIZES 24 MALFUNCTION EVENTS IN WHICH THE DEVICE HAD A COMPONENT DETACH. 22 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823271 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices