Skip to content

ADAPT 0101-2900Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · PC Unit · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327312393), received between Dec 5, 2019 and Dec 5, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Image Display Error/Artifact” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Image Display Error/Artifact1
  • Incorrect Measurement1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 5, 2019Image Display Error/Artifact; Incorrect Measurement; Image Display Error/Artifact; Incorrect Measurement

    DURING TESTING OF THE DEVICE, SIGNIFICANT MISREPRESENTATIONS AND MISMEASUREMENTS OCCURRED. THE DEVICE IS REPORTED TO BE NEW. SURGICAL DELAY OF 30MIN. NOT LONGER. THIS RESULTED IN TRIPLE INCREASE IN RADIATION EXPOSURE, NORMALLY RADIATION TIME FOR GAMMA3 NAIL IN THIS HOSPITAL IS 12-13 SECONDS, IN THIS SURGERY 40 SECONDS. NO OTHER PATIENT IMPACT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9417005 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page