Triathlon 5531-G-409-ERecalls & reported problems
Howmedica Osteonics Corp. · Tibial Bearing Insert - CS · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Howmedica Osteonics Triathlon tibial bearing insert is the plastic spacer fitted into a total knee replacement. The 37 adverse event reports mostly describe knee instability, infection, pain and limited movement that led to revision surgery to swap or resize the insert, with36 injuries and no deaths. One report describes a locking wire coming loose from the insert in its package before it touched the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 37
- Deaths reported
- 0
- Injuries reported
- 36
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 07613327337150), received between Mar 30, 2020 and Aug 26, 2026 — classified as 36 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (7 reports).
7 other recalls cover devices cleared under the same FDA submission (K172634); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem21
- Insufficient Device Problem Information7
- Unstable7
- Detachment of Device or Device Component2
- Device Dislodged or Dislocated2
- Fracture1
- Material Erosion1
- Mechanical Problem1
Patient problems
- Unspecified Infection14
- Pain7
- Joint Laxity6
- Ambulation or Postural Difficulties4
- Loss of Range of Motion4
- Fall2
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.
InjuryAug 26, 2026Unstable
Left knee. Patient had their insert revised due to instability in both flexion and extension. A 4x9 CS insert was exchanged for a 4x11 CS insert. No other information available due to Hospital policy.
Patient problems: Joint Laxity
MAUDE report 50461231 ↗InjuryMay 5, 2026Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS PROVIDED: LEFT TOTAL KNEE REPLACEMENT BECAME INFECTED, REPLACED POLY.
Patient problems: Unspecified Infection
MAUDE report 25086941 ↗InjuryMay 1, 2026Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS PROVIDED: IN PROVIDING INFORMATION FOR (REVISION OF PATIENT'S RIGHT KNEE ON (B)(6) 2026, REP PROVIDED INFORMATION INDICATING THAT THE PATIENT'S RIGHT KNEE WAS REVISED ON (B)(6) 2025. REASON FOR REVISION IS UNKNOWN. A 4X9 CS INSERT WAS REVISED TO A 4X10 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Insufficient Information
MAUDE report 25054626 ↗InjuryJul 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S KNEE WAS REVISED DUE TO POSSIBLE INFECTION. AN I&D WITH POLY EXCHANGE WAS PERFORMED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 22660005 ↗InjuryApr 2, 2025Unstable; Unstable
IT WAS REPORTED THAT THE SURGEON REVISED A TIBIAL INSERT OF A RIGHT KNEE DUE TO INSTABILITY. HE UPSIZED FROM A 4X9 TO 4X11 CS. NO FURTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY.
Patient problems: Joint Laxity
MAUDE report 21751034 ↗InjuryFeb 5, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO PAIN AND LIMITED ROM. A 4X9 INSERT WAS REVISED TO ANOTHER 4X9 INSERT, AND THE PATIENT'S NATIVE PATELLA WAS RESURFACED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Loss of Range of Motion; Pain
MAUDE report 21315168 ↗InjuryDec 10, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
RIGHT TOTAL KNEE. THE REVISION TODAY WAS SECONDARY TO A FALL AND INJURY TO THE PATELLA TENDON. THE 9 MM INSERT WAS REMOVED TO ALLOW FOR WASHING OUT OF THE JOINT AS IT WAS AN OPEN INJURY. A NEW 9MM WAS PLACED AND IT WAS DETERMINED THAT THE JOINT WAS UNSTABLE. AFTER TRIALING A 14MM INSERT WAS IMPLANTED TO STABILIZE THE JOINT. NO FURTHER INFORMATION IS AVAILABLE FROM THE HOSPITAL OR SURGEON.
Patient problems: Fall
MAUDE report 20896469 ↗MalfunctionSep 12, 2024Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
PRIMARY PROCEDURE, LEFT KNEE. IT WAS REPORTED THAT IMMEDIATELY AFTER REMOVING THE INSERT FROM THE STERILE BLISTER, THE LOCKING WIRE FELL OFF OF THE INSERT. THE INSERT DID NOT COME IN CONTACT WITH THE PATIENT. A SECOND IMPLANT WAS USED TO COMPLETE THE SURGERY SUCCESSFULLY WITH NO DELAY. REP WAS PRESENT FOR AND WITNESS TO THIS EVENT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20199669 ↗InjuryApr 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
LEFT TOTAL KNEE. THE REVISION TODAY WAS SECONDARY TO PAIN AND ¿CATCHING¿ IN THE POSTERIOR LATERAL PART OF THE KNEE. ON X-RAY A FRAGMENT OF BONE WAS IDENTIFIED AND BELIEVED TO BE CAUSING THE PROBLEM. IN SURGERY THE 9MM INSERT WAS REMOVED TO MAKE ACCESS TO THE BONE FRAGMENT EASIER. THE BONE FRAGMENT WAS REMOVED. AFTER TRIALING SEVERAL INSERTS IT WAS DECIDED THAT AN 11MM INSERT PROVIDED THE BEST STABILITY. THE USE OF MAKO IN THE INDEX PROCEDURE WAS NOT BELIEVED TO HAVE CONTRIBUTED TO THIS PROBLEM. NO FURTHER INFORMATION AVAILABLE FROM THE SURGEON OR HOSPITAL.
Patient problems: Bone Fracture(s); Pain
MAUDE report 19211377 ↗InjuryApr 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A LEFT KNEE REVISION. PATIENT PRESENTED WITH DRAINAGE OF HER LEFT-KNEE FOLLOWING TKA SURGERY ON (B)(6) 2024. DR. PERFORMED AN EXTENSIVE I&D AND INSERT-SWAP, AND DETERMINED THE DRAINAGE RESULTED FROM A SUPERFICIAL ABSCESS. NO COMPLAINTS WERE FILED AGAINST THE COMPONENTS THEMSELVES, NO FURTHER INFO AVAILABLE.
Patient problems: Unspecified Infection
MAUDE report 19090630 ↗InjuryFeb 7, 2024Material Erosion; Fracture; Unstable; Naturally Worn; Material Erosion; Fracture; Unstable; Naturally Worn
IT WAS REPORTED: PATIENT CAME BACK DUE TO KNEE PAIN. CASE WENT SMOOTH, ISSUE IS THE POLY HAS WORN OUT ON THE MEDIAL SIDE IN A FEW YEARS. WE ARE REVISING THIS IN TWO WEEKS.
Patient problems: Joint Laxity; Pain
MAUDE report 18659692 ↗InjuryJan 3, 2024Unstable; Unstable
REVISED A LEFT TRIATHLON TKA DUE TO INSTABILITY.
Patient problems: Joint Laxity
MAUDE report 18437846 ↗InjuryDec 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO SUSPECTED INFECTION. A POLY EXCHANGE WAS PERFORMED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 18369883 ↗InjuryDec 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISED A TRIATHLON TIBIAL INSERT OF A RIGHT KNEE DUE TO PATIENT FALLING. HE HAD TO PERFORM A QUAD REPAIR SO IN THAT PROCESS EXPLANTED THE INSERT FOR ACCESS.
Patient problems: Fall
MAUDE report 18362401 ↗InjuryNov 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO LACK OF MOTION (SURGEON REPORTED CAUSE AS PATIENT DID NOT UNDERGO PHYSICAL THERAPY). A 4X9 CS INSERT WAS EXCHANGED FOR ANOTHER 4X9 CS INSERT, AND SOFT TISSUE DEBRIDEMENT WAS PERFORMED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Loss of Range of Motion
MAUDE report 18136272 ↗InjuryOct 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PERFORMED AN I&D WITH INSERT SWAP OF A LEFT TOTAL KNEE DUE TO POSSIBLE INFECTION.
Patient problems: Unspecified Infection
MAUDE report 17948668 ↗InjuryOct 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
EXPLANTED A PRIMARY TRIATHLON LEFT KNEE DUE TO INFECTION. HE PLACED AN ANTIBIOTIC SPACER.
Patient problems: Unspecified Infection
MAUDE report 17910724 ↗InjuryOct 9, 2023Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT'S KNEE WAS REVISED DUE TO INSTABILITY. A 4X9 CS INSERT WAS EXCHANGED FOR A 4X13 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Joint Laxity
MAUDE report 17893045 ↗InjurySep 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
I&D ON THE LEFT KNEE OF A PATIENT THAT LAST HAD AN I&D ON (B)(6) 2023 DUE TO INFECTION. DR. INTENDED TODAY TO DO THIS I&D AS A SHORT TERM SOLUTION AND PLANS TO FULLY EXPLANT WITHIN A FEW WEEKS FOR AN ANTIBIOTIC SPACER.
Patient problems: Unspecified Infection
MAUDE report 17805595 ↗InjuryJun 30, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DR. PERFORMED AN I&D INSERT SWAP OF A LEFT TKA DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 17234965 ↗InjuryJan 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE SALES REP THAT THERE HAS BEEN AN EXCHANGE OF INSERT DUE TO UNKNOWN INFECTION.
Patient problems: Unspecified Infection
MAUDE report 16190361 ↗InjuryDec 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: " RIGHT TOTAL KNEE REPLACEMENT WAS PERFORMED DUE TO STIFFNESS IN THE JOINT."
Patient problems: Loss of Range of Motion
MAUDE report 15985682 ↗InjuryNov 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RIGHT KNEE MANIPULATION.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 15840913 ↗InjuryNov 21, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
LEFT KNEE MANIPULATION ON (B)(6) 2022.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 15832898 ↗InjurySep 20, 2022Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated
THE FOLLOWING WAS REPORTED: TRIATHLON X3 POLY LOCKING WIRE FAILURE. TKA LEFT.
Patient problems: Insufficient Information
MAUDE report 15459647 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172634. One submission often covers many models, so check the product and lot details in each record.
- Z-1261-2024 · Feb 6, 2024 · Open, ClassifiedX3 TRIATHLON CS INSERT NO 6 11 MM
- Z-1263-2024 · Feb 6, 2024 · Open, ClassifiedNO 3 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM
- Z-1266-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MM
- Z-1265-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MM
- Z-1264-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM
- Z-1262-2024 · Feb 6, 2024 · Open, ClassifiedX3 TRIATHLON CS INSERT NO 7 9 MM
- Z-1654-2023 · Apr 13, 2023 · Open, ClassifiedTRIATHLON X3 TIBIAL BEARING INSERT PS #4 13MM Intended for knee replacement Catalog Number: 5532-G-413-E
Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices
- ATTUNE 1512-14-050Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER106 reports
- ATTUNE 1504-40-205Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall53 reports
- ATTUNE 1504-40-105Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall46 reports
- ATTUNE 1547-06-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall37 reports
- ATTUNE 1513-12-060Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER72 reports
- Triathlon 5531-G-709-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5521-B-300Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5531-G-610-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.