Triathlon 5531-G-611-ERecalls & reported problems
Howmedica Osteonics Corp. · Tibial Bearing Insert - CS · Class 2 · Implant
All models of Triathlon (608)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 15
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 07613327337341), received between Jun 15, 2020 and Feb 10, 2026 — classified as 15 injuries.
The most frequently reported specific device problem is “Unstable” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K172634); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Feb 6, 2024Open, Classified
Potential packaging breaches of inner blister and outer sterile blister.
Z-1261-2024 · Root cause: Packaging · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem8
- Unstable4
- Insufficient Device Problem Information2
- Detachment of Device or Device Component1
Patient problems
- Unspecified Infection9
- Insufficient Information2
- Joint Laxity2
- Ambulation or Postural Difficulties1
- Bacterial Infection1
- Injury1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
InjuryFeb 10, 2026Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS PROVIDED: THE REVISION TODAY OF A LEFT TOTAL KNEE WAS SECONDARY TO INFECTION. ALL ORIGINAL IMPLANTS WERE REMOVED, THE JOINT WAS WASHED OUT, AND NEW IMPLANTS WERE PLACED. THE USE OF MAKO IN THE INDEX SURGERY WAS NOT THOUGHT TO HAVE CONTRIBUTED TO THIS COMPLICATION.
Patient problems: Unspecified Infection
MAUDE report 24303032 ↗InjuryOct 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
RIGHT TOTAL KNEE. THE REVISION TODAY WAS SECONDARY TO INFECTION. 6X11 CS INSERT WAS EXCHANGED FOR ANOTHER. THE USE OF MAKO IN THE INDEX SURGERY WAS NOT THOUGHT TO HAVE CONTRIBUTED TO THIS COMPLICATION. NO FURTHER INFORMATION IS AVAILABLE FROM THE DOCTOR OR HOSPITAL.
Patient problems: Unspecified Infection
MAUDE report 20356151 ↗InjuryMay 2, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO INFECTION. A 6X11 CS INSERT WAS EXCHANGED FOR ANOTHER OF THE SAME CATALOG NUMBER. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 19234312 ↗InjuryApr 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED: "REVISION RTKA WITH ARTICULATING SPACER PLACEMENT BECAUSE OF INFECTION." REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 19114346 ↗InjuryApr 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO INFECTION. A FIRST OF A TWO STAGE REVISION WAS PERFORMED ON (B)(6) 2024 WITH PLACEMENT OF A SPACER. THE 2ND STAGE REMOVAL OF THE SPACER AND PLACEMENT OF PERMANENT IMPLANTS WAS PERFORMED ON (B)(6) 2024.
Patient problems: Unspecified Infection
MAUDE report 19059009 ↗InjuryMar 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO POSSIBLE INFECTION. A 6X11 CS INSERT WAS EXCHANGED FOR ANOTHER 6X11 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 16551053 ↗InjuryJan 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PATIENT'S KNEE WAS REVISED. APPROXIMATELY 1 MONTH POST-IMPLANT, X-RAYS SHOWED THAT THE INSERT WAS NOT LOCKED INTO THE BASEPLATE. A 6X11 CS INSERT WAS REVISED TO A 6X10 CS INSERT. REP CAN PROVIDE SOME ADDITIONAL INFORMATION (REFER TO COMMUNICATIONS LOG), AND CONFIRMED THAT OTHERWISE, NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Insufficient Information
MAUDE report 16213512 ↗InjuryDec 20, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT REPORTED KNEE INFECTION AFTER PRIMARY. PATIENT NOTED THAT INFECTION WAS TREATED BY A DENTIST WITH AN ANTIBIOTIC. PATIENT WOULD LIKE TO KNOW IF IMPLANTS ARE SUBJECT TO RECALL. UPDATE PER PATIENT RESPONSE: "I HAVE NO COMPLAINTS, JUST TRYING TO MAKE SURE ALL PARTS ARE GOOD TO GO SO I CAN REQUEST THE EXACT SAME MODEL FOR THE LEFT KNEE AND HAVE A SUCCESSFUL, QUICK RECOVERY."
Patient problems: Unspecified Infection
MAUDE report 16018180 ↗InjurySep 9, 2022Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO INSTABILITY (REP CONFIRMED THAT NOTHING WAS REPORTED WORN, BROKEN, OR LOOSE). A 6X11 CS INSERT WAS REVISED TO A 6X14 CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Joint Laxity
MAUDE report 15394002 ↗InjuryJun 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A STAGE 1 REVISION WAS PERFORMED ON THE PATIENT'S RIGHT KNEE DUE TO INFECTION. ALL IMPLANTS WERE REMOVED AND A SPACER WAS PLACED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Unspecified Infection
MAUDE report 14839374 ↗InjuryMar 13, 2022Unstable; Unstable
AS REPORTED: "PT. PRESENTED WITH MID-FLEXION INSTABILITY. OPTED TO REVISE THE FEMUR COMPONENT AND INSERT OPTION, LEAVING THE TIBIAL-BASEPLATE COMPONENT IN-SITU." RIGHT KNEE. A CR FEMORAL COMPONENT AND CS INSERT WERE REVISED TO A TS FEMORAL COMPONENT WITH STEM, EXTENDER, AND AUGMENTS, AND A TS INSERT. REP PROVIDED REVISION USAGE SHEET. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Joint Laxity
MAUDE report 13752424 ↗InjuryFeb 18, 2021Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO LIGAMENT INSTABILITY. THE FEMORAL COMPONENT AND INSERT WERE REVISED. REP REPORTED HE CAN POSSIBLY PROVIDE PRIMARY AND REVISION USAGE, AND CONFIRMED THAT OTHERWISE NO FURTHER INFORMATION WOULD BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Insufficient Information
MAUDE report 11343551 ↗InjuryDec 21, 2020Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED DUE TO INSTABILITY. SURGEON REPORTED THAT THE PATIENT'S FEMORAL COMPONENT DISLOCATED POSTERIORLY FROM THE INSERT. A 6X11 CS INSERT WAS REVISED TO A THICKER CS INSERT. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.
Patient problems: Ambulation or Postural Difficulties
MAUDE report 11044536 ↗InjuryOct 22, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
THE FOLLOWING WAS REPORTED: PATIENT HAS CLEAR LIQUID OOZING FROM WOUND. SURGEON PREFORMED I&D AND POLY SWAP.
Patient problems: Bacterial Infection
MAUDE report 10718993 ↗InjuryJun 15, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
DR (B)(6) REPORTS PATIENT (B)(6) A STATUS POST PRIMARY LEFT TOTAL KNEE DONE ON (B)(6)2020 APPEARS INFECTED AND REQUESTS TO REMOVE THE IMPLANTS AND PUT IN AN ANTIBIOTIC SPACER. DR (B)(6) CAREFULLY REMOVED THE IMPLANTS USING FLEXIBLE OSTEOTOMES AND A SAW. HE THEN PROCEEDED TO DEBRIDE AND WASH THE BONE. ONCE REMOVED AND WASHED HE IMPLANTED AN ANTIBIOTIC SPACER MADE OF CEMENT LADEN WITH EXTRA ANTIBIOTICS. THE PROCEDURE WAS PERFORMED WITHOUT INCIDENT. NO FURTHER INFORMATION IS AVAILABLE. CORRECTION 01/JUNE/2020 WG: BRANCH PROVIDED IMPLANT SHEETS FOR THE ROBOTIC PRIMARY ON (B)(6) 2019 AND A POLY SWAP PERFORMED ON (B)(6) 2020.
Patient problems: Injury; Unspecified Infection
MAUDE report 10152143 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K172634. One submission often covers many models, so check the product and lot details in each record.
- Z-1265-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MM
- Z-1263-2024 · Feb 6, 2024 · Open, ClassifiedNO 3 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM
- Z-1262-2024 · Feb 6, 2024 · Open, ClassifiedX3 TRIATHLON CS INSERT NO 7 9 MM
- Z-1266-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 16 MM
- Z-1264-2024 · Feb 6, 2024 · Open, ClassifiedNO 4 TRIATHLON TS PLUS TIB INS X3 POLY 11 MM
- Z-1654-2023 · Apr 13, 2023 · Open, ClassifiedTRIATHLON X3 TIBIAL BEARING INSERT PS #4 13MM Intended for knee replacement Catalog Number: 5532-G-413-E
Other PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER devices
- ATTUNE 1512-14-050Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER106 reports
- ATTUNE 1504-40-205Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall53 reports
- ATTUNE 1504-40-105Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall46 reports
- ATTUNE 1547-06-001Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall37 reports
- ATTUNE 1513-12-060Depuy (Ireland) · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER72 reports
- Triathlon 5531-G-709-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5521-B-300Howmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall19 reports
- Triathlon 5531-G-610-EHowmedica Osteonics Corp. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page