Xia 3 System - Serrato 482616550Recalls & reported problems
Stryker Corporation · Polyaxial Screw · Class 2 · Implant
All models of Xia 3 System - Serrato (36)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327351552), received between Mar 29, 2019 and Nov 14, 2024 — classified as 7 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration5
- Appropriate Device Problem Term/Code Not Available2
- Fracture2
- Loose or Intermittent Connection2
- Material Protrusion/Extrusion2
- Unintended Movement2
- Device Slipped1
Patient problems
- Pain7
- Injury6
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryNov 14, 2024Migration; Migration
IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION SURGERY APPROXIMATELY TWO YEARS POST-OPERATIVELY TO ADDRESS A MIGRATED XIA SERRATO POLYAXIAL SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20690935 ↗MalfunctionJan 27, 2023Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT EXPERIENCED POST-OPERATIVE FRACTURE OF TWO SERRATO SCREWS. THE PATIENT HAS NOT BEEN REVISED AND REVISION IS NOT CURRENTLY SCHEDULED. THIS REPORT CAPTURES THE FIRST OF TWO FRACTURED SCREWS.
Patient problems: Pain
MAUDE report 16258308 ↗MalfunctionJan 27, 2023Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT EXPERIENCED POST-OPERATIVE FRACTURE OF TWO SERRATO SCREWS. THE PATIENT HAS NOT BEEN REVISED AND REVISION IS NOT CURRENTLY SCHEDULED. THIS REPORT CAPTURES THE SECOND OF TWO FRACTURED SCREWS.
Patient problems: Pain
MAUDE report 16258224 ↗InjuryJul 9, 2019Migration
IT WAS REPORTED THAT A SCREW MIGRATED POST-OPERATIVELY. PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2019. DURING REVISION SURGERY, THE SURGEON REMOVED THE SAME SCREW THAT WAS IMPLANTED INTO S1 AND RE-IMPLANTED AT A DIFFERENT ANGLE.
Patient problems: Injury
MAUDE report 8771793 ↗InjuryMar 30, 2019Material Protrusion/Extrusion; Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE PATIENT WAS PRESENTED WITH A LEG PAIN POST-OPERATIVELY. A POST-OP X-RAY SHOWED A COMPROMISED STRUCTURE AND POTENTIAL DISLODGEMENT OF THE CAGE. FURTHER COMMUNICATION WITH SALES REP AND CT SCAN REVIEW CONFIRMED THAT THE CAGE HAD NOT BEEN DISLODGED. AN X-RAY SHOWED THE SCREW HAS PROTRUDED THROUGH THE VERTEBRAL BODY INTO DISC SPACE. A REVISION SURGERY WAS PERFORMED TO REMOVE SCREW AND CONSTRUCT EXTENSION FROM L3-S1.
Patient problems: Injury; Pain
MAUDE report 8467486 ↗InjuryMar 29, 2019Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available; Migration; Migration
IT WAS REPORTED THAT THE SCREW WAS FOUND TO BE LOOSE. THE PATIENT WAS IN PAIN AND REVISION SURGERY WAS PERFORMED FOR REPLACEMENT OF HARDWARE, UPSIZE SCREWS AND FUSE POSTERIOR LATERAL.
Patient problems: Injury; Pain
MAUDE report 8467390 ↗InjuryMar 29, 2019Material Protrusion/Extrusion; Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE PATIENT WAS PRESENTED WITH A LEG PAIN POST-OPERATIVELY. A POST-OP X-RAY SHOWED A COMPROMISED STRUCTURE AND POTENTIAL DISLODGEMENT OF CAGE. FURTHER COMMUNICATION WITH SALES REP AND CT SCAN REVIEW CONFIRMED THAT THE CAGE HAD NOT BEEN DISLODGED. AN X-RAY SHOWED THE SCREW HAS PROTRUDED THROUGH THE VERTEBRAL BODY INTO DISC SPACE. A REVISION SURGERY WAS PERFORMED TO REMOVE SCREW AND CONSTRUCT EXTENSION FROM L3-S1.
Patient problems: Injury; Pain
MAUDE report 8467389 ↗InjuryMar 29, 2019Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available; Migration; Migration
IT WAS REPORTED THAT THE SCREW WAS FOUND TO BE LOOSE. THE PATIENT WAS IN PAIN AND REVISION SURGERY WAS PERFORMED FOR REPLACEMENT OF HARDWARE, UPSIZE SCREWS AND FUSE POSTERIOR LATERAL.
Patient problems: Injury; Pain
MAUDE report 8467159 ↗InjuryMar 29, 2019Device Slipped; Migration; Migration
IT WAS REPORTED THAT THE SCREW WAS FOUND TO BE LOOSE. THE PATIENT WAS IN PAIN AND REVISION SURGERY WAS PERFORMED FOR REPLACEMENT OF HARDWARE, UPSIZE SCREWS AND FUSE POSTERIOR LATERAL.
Patient problems: Injury; Pain
MAUDE report 8467120 ↗
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page