Xia 3 System - Serrato 482619590Recalls & reported problems
Stryker Corporation · Polyaxial Screw · Class 2 · Implant
All models of Xia 3 System - Serrato (36)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327351576), received between Dec 4, 2017 and Jan 19, 2021 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Separation” (3 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jun 18, 2019Class IITerminated
Discrepant length; Lot B88751 90mm screws were manufactured with a 100mm length instead of the required 90mm length; lot B90243 100mm screws were manufactured with a 90mm length instead of the required 100mm length.
Z-2160-2019 · Root cause: Labeling mix-ups · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation3
- Break1
- Detachment Of Device Component1
- Device Markings/Labelling Problem1
- Fracture1
- Manufacturing, Packaging or Shipping Problem1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient2
- Injury1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 19, 2021Material Separation; Material Separation
IT WAS REPORTED THAT DURING A SCHEDULED REVISION, THE TULIP OF A XIA SERRATO POLYAXIAL SCREW BROKE OFF DURING SCREW REMOVAL. WHILE REPLACING THE SERRATO SCREW, THE TULIP OF A XIA 3 POLYAXIAL SCREW ALSO BROKE OFF. ALL PIECES OF THE FIRST SCREW WERE REMOVED; HOWEVER, THE SHAFT OF THE SECOND SCREW REMAINS IN THE PATIENT. SURGERY WAS SUCCESSFULLY COMPLETED WITH A 30 MINUTE DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS REPORT WILL REPRESENT THE SERRATO SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11194637 ↗InjuryJul 10, 2020Material Separation; Material Separation
IT WAS REPORTED THAT WHILE FINAL TIGHTENING TWO XIA SERRATO POLYAXIAL SCREWS AT S2AI, THE TULIPS SEPARATED FROM THE SHAFTS. ONE OF THE SCREW SHAFTS WAS REMOVED ENTIRELY AND ONE OF THE SCREW SHAFTS REMAINED IN THE PATIENT¿S BONE. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF THREE-HOURS AND THE CONSTRUCT WAS COMPLETED WITH THE USE OF ILIAC BOLTS. THIS RECORD REPRESENTS THE SCREW WHOSE SHAFT REMAINS IMPLANTED IN THE PATIENT¿S BONE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10261167 ↗InjuryJul 10, 2020Material Separation; Material Separation
IT WAS REPORTED THAT WHILE FINAL TIGHTENING TWO XIA SERRATO POLYAXIAL SCREWS AT S2AI, THE TULIPS SEPARATED FROM THE SHAFTS. ONE OF THE SCREW SHAFTS WAS REMOVED ENTIRELY AND ONE OF THE SCREW SHAFTS REMAINED IN THE PATIENT¿S BONE. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF THREE-HOURS AND THE CONSTRUCT WAS COMPLETED WITH THE USE OF ILIAC BOLTS. THIS RECORD REPRESENTS THE SCREW WHICH WAS FULLY REMOVED FROM THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10261164 ↗MalfunctionJul 10, 2019Manufacturing, Packaging or Shipping Problem; Nonstandard Device; Device Markings/Labelling Problem; Manufacturing, Packaging or Shipping Problem
A SCREW LABELLED AS 95 MM, WAS ACTUALLY 100 MM IN LENGTH. NO PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES TO A PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8778453 ↗InjuryMar 30, 2019Break; Fracture; Fracture
IT WAS REPORTED THAT THE SCREW WAS BROKEN POST-OPERATIVELY. A REVISION SURGERY WILL BE REQUIRED TO REPLACE THE SCREW.
Patient problems: Injury; No Known Impact Or Consequence To Patient
MAUDE report 8467480 ↗MalfunctionDec 4, 2017Detachment Of Device Component
IT WAS REPORTED THAT THE POLY HEAD POPPED OFF OF THE SCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7079544 ↗
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page