Skip to content

Xia 3 System - Serrato 482618570Recalls & reported problems

Stryker Corporation · Polyaxial Screw · Class 2 · Implant

All models of Xia 3 System - Serrato (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327351736), received between Jul 22, 2019 and May 11, 2021 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Material Separation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Injury1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2021Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TULIPS OF 3 XIA SERRATO POLYAXIAL SCREWS POPPED OFF THE SHANKS INTRA-OPERATIVELY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH AN UNSPECIFIED SURGICAL DELAY BY USING ALTERNATIVELY AVAILABLE IMPLANTS. THIS REPORT CAPTURES THE SECOND OF THE THREE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11807135 ↗
  • MalfunctionMay 11, 2021Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE TULIPS OF 3 XIA SERRATO POLYAXIAL SCREWS POPPED OFF THE SHANKS INTRA-OPERATIVELY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH AN UNSPECIFIED SURGICAL DELAY BY USING ALTERNATIVELY AVAILABLE IMPLANTS. THIS REPORT CAPTURES THE FIRST OF THE THREE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11807120 ↗
  • InjuryOct 16, 2020Material Separation; Material Separation

    IT WAS REPORTED THAT THE TULIP OF A XIA SERRATO POLYAXIAL SCREW LOCATED AT S5 "CAME APART" FROM THE SHAFT POST-OPERATIVELY. REVISION SURGERY HAS BEEN PERFORMED WHERE THE SCREW WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10693098 ↗
  • InjuryJul 22, 2019Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION SURGERY(B)(6) 2019 FOR THE REMOVAL OF AN S1 SERRATO SCREW WHICH "CAME OFF" POST-OPERATIVELY. THIS RECORD IS RELATED TO MANUFACTURER'S INCIDENT REPORT 3005525032-2019-00069.

    Patient problems: Injury

    MAUDE report 8812182 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page