Skip to content

Xia 3 System - Serrato 482678580Recalls & reported problems

Stryker Corporation · Medial Biased-Angle Polyaxial Screw · Class 2 · Implant

All models of Xia 3 System - Serrato (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327351873), received between Mar 4, 2019 and Jan 13, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Device Dislodged or Dislocated2
  • Break1
  • Fracture1
  • Misassembled1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • Spinal Column Injury1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 13, 2026Detachment of Device or Device Component; Device Dislodged or Dislocated

    A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WAS REVISED TO ADDRESS A XIA SERRATO POLYAXIAL SCREW TULIP DISENGAGEMENT, WHICH WAS IDENTIFIED ON POST-OPERATIVE IMAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24064055 ↗
  • InjuryMay 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    NO NEW INFORMATION.

    Patient problems: Spinal Column Injury

    MAUDE report 22032638 ↗
  • MalfunctionDec 16, 2020Misassembled; Misassembled

    IT WAS REPORTED THAT A XIA SERRATO POLYAXIAL SCREW WAS NOT ABLE TO BE FULLY FINAL TIGHTENED INTRA-OPERATIVELY. THE SURGEON LEFT THE SCREW AS IS AND COMPLETED THE PROCEDURE. NO ADVERSE CONSEQUENCES OR SURGICAL DELAYS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11016218 ↗
  • MalfunctionMar 4, 2019Break; Fracture; Device Dislodged or Dislocated; Fracture; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE TULIP OF SCREW DISSOCIATED WHILE PLACING BLOCKER. A SURGICAL DELAY OF 8 MINUTE WAS REPORTED TO REPLACE THE SCREW. ALL BROKEN FRAGMENTS WERE RETRIEVED. THE SURGERY WAS COMPLETED USING THE SAME BLOCKER. NO OTHER CONSEQUENCE WAS REPORTED TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 8385948 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page