Skip to content

Xia 3 System - Serrato 482615545Recalls & reported problems

Stryker Corporation · Polyaxial Screw · Class 2 · Implant

All models of Xia 3 System - Serrato (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327352054), received between Jan 28, 2020 and Jan 28, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Paralysis1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 28, 2020Migration; Migration

    PHYSICIAN REPORTED THAT AFTER INTERVERTEBRAL POSTERIOR LUMBAR INTERBODY FUSION (PLIF) OF L4-L5/S A PATIENT BECAME PARALYZED AND THE L5 XIA SERRATO 5.5X45 POLYAXIAL SCREW LOOKED AS IF IT HAD COME OUT OF THE PEDICLE IN A POST-OPERATIVE X-RAY. UPON REVISION SURGERY, IT WAS OBSERVED THAT THE NERVE ROOT HAD BEEN DAMAGED. THE SCREW WAS REMOVED AND REPLACED WITH A DIFFERENT-SIZED SCREW AT A DIFFERENT INSERTION ANGLE. PARALYSIS RESOLVED UPON REMOVAL OF SCREW.

    Patient problems: Paralysis

    MAUDE report 9638653 ↗