Skip to content

Xia 3 System - Serrato 482618550Recalls & reported problems

Stryker Corporation · Polyaxial Screw · Class 2 · Implant

All models of Xia 3 System - Serrato (36)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327352344), received between May 26, 2021 and Feb 26, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Fracture1
  • Material Deformation1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2026Device Dislodged or Dislocated

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24455764 ↗
  • InjuryNov 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    A COMPANY REPRESENTATIVE REPORTED THAT THE TULIP OF A XIA SERRATO POLYAXIAL SCREW DISENGAGED FROM THE SCREW SHANK INTRA-OPERATIVELY. IT WAS FURTHER REPORTED THAT THE SCREW WAS REMOVED AND A LONGER SCREW WAS INSERTED INSTEAD. THE PROCEDURE WAS COMPLETED WITH A 10 MINUTE SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20689648 ↗
  • MalfunctionMay 26, 2021Fracture; Material Separation; Material Deformation; Fracture; Material Separation; Material Deformation

    A COMPANY REPRESENTATIVE REPORTED THAT AN 8.5 X 50 MM SERRATO POLYAXIAL SCREW TULIP SPLAYED DURING INSERTION OF THE BLOCKER. THE TULIP THEN DISENGAGED FROM THE SCREW SHAFT WHEN THE SURGEON WENT TO REMOVE AND REPLACE THE BLOCKER. THE PHYSICIAN CHANGED OUT THE SCREW AND SURGERY WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY OR ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11885445 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates