T2 2353-3907Recalls & reported problems
Stryker GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07613327411843), received between Sep 10, 2021 and Nov 24, 2025 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Failure to Align” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Align3
- Mechanical Jam2
- Break1
- Detachment of Device or Device Component1
- Device Difficult to Setup or Prepare1
- Difficult to Insert1
- Difficult to Remove1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 24, 2025Failure to Align; Failure to Align
AS REPORTED: "WHEN INSERTING THE DISTAL ADVANCE LOCKING SCREW, THE DISTAL TARGETING DEVICE WAS LOOSE, AND THE SCREW WAS DRIVEN IN AT AN INCORRECT ANGLE BEYOND THE ACCEPTABLE INSERTION RANGE. AS A RESULT, THE SCREW BECAME STUCK WITH THE NAIL AND COULD NOT BE REMOVED. ALTHOUGH IT WAS EVENTUALLY EXTRACTED WITH EFFORT, THE PRODUCTS WERE DAMAGED." 11/06/2025 - ADDITIONAL INFORMATION RECEIVED: HOW WAS THE ISSUE SOLVED? LONGER SCREW DRIVER WAS USED AS AN ALTERNATIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23632949 ↗MalfunctionJun 18, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED: "GREY BUTTON FELL OUT OF THE TARGETER, NO LONGER ABLE TO LOCK THE TRIPLE SLEEVE INTO THE TARGETER, UNABLE TO PLACE SCREW IN PREFERRED DISTAL LOCKING HOLE".
Patient problems: Insufficient Information
MAUDE report 22288219 ↗MalfunctionNov 23, 2024Difficult to Remove; Mechanical Jam; Difficult to Remove; Mechanical Jam
AS REPORTED: "WHEN REMOVING THE DISTAL AND PROXIMAL DEVICES AFTER SURGERY, THEY WERE FIRMLY FIXED AND COULD NOT BE REMOVED MANUALLY, SO DR. TRIED TO REMOVE THEM BY HITTING THEM WITH A HAMMER AND DAMAGED THEM".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20766494 ↗InjuryJul 10, 2024Mechanical Jam; Mechanical Jam
AS REPORTED: "T2 ALPHA FEMUR RETROGRADE DISTAL TARGETING ATTACHMENT MALFUNCTIONED AND WOULD NOT ATTACH TO NAIL HOLDING JIG UNLESS GREAT FORCE WAS APPLIED. ONCE ATTACHED, WITH GREAT EFFORT, IT COULD NOT BE REMOVED. COULD NOT UTILIZE PROXIMAL TARGETING ATTACHMENT, ALSO HAD TO ACQUIRE VARIOUS ADDITIONAL X-RAY ANGLES TO PROPERLY ASCERTAIN NAIL PLACEMENT, SINCE WE COULD NOT REMOVE DISTAL TARGETING ATTACHMENT. BOTH OF THESE CAUSED A PROLONGED PROCEDURE OF 30 MIN."
Patient problems: Insufficient Information
MAUDE report 19710015 ↗MalfunctionApr 20, 2023Break; Fracture; Break; Fracture
AS REPORTED: "THE PIECE THAT HOLDS THE SLEEVE BROKE OFF AND NOW IT WON'T HOLD THE SLEEVES IN PLACE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16778166 ↗MalfunctionDec 20, 2022Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
AS REPORTED: "WHEN CHECKING THE DRILL PASSAGE BEFORE INSERTING THE NAIL, DISTAL TARGETING DEVICE COULD NOT BE ATTACHED TO THE NAIL ADAPTER. THE LOCKING LEVER ON THE TARGETING DEVICE DID NOT MOVE AT ALL. DUE TO POOR FUNCTION, THE PROCEDURE WAS CHANGED TO THE USE OF A RADIOLUCENT DRILL AND THE SURGERY WAS COMPLETED.".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16014778 ↗MalfunctionSep 14, 2021Difficult to Insert; Failure to Align; Difficult to Insert; Failure to Align
THE END USER REPORTED THEY HAD A COLINEAR CLAMP ON LATERALLY AND WHEN THE TARGETING WING WAS ATTACHED IT DID HIT THAT CLAMP. THEY ALSO REPORTED THAT WITH ANTEGRADE NAILS, SOME ROC HAVE MIS-MATCHED THE PATIENT AND MADE INSERTION DIFFICULT. NO ADVERSE CONSEQUENCES WERE REPORTED TO THE PATIENT AND NO CLINICALLY SIGNIFICANT DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12470746 ↗MalfunctionSep 10, 2021Failure to Align; Failure to Align
IT WAS REPORTED THAT ONE ADVANCED LOCKER MISSED THE LATERAL OBLIQUE HOLE. THE FIRST DRILL PASS HIT A TINE OF A LARGE WEBER CLAMP POSSIBLY AFFECTING THE TRAJECTORY OF THE SCREW. THE SURGEON USED A REGULAR LOCKER AND INSERTION WENT COMPLETELY NORMAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12456415 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page