Stryker GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07613327411843), received between Sep 10, 2021 and Nov 24, 2025 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Failure to Align” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align3
  • Mechanical Jam2
  • Break1
  • Detachment of Device or Device Component1
  • Device Difficult to Setup or Prepare1
  • Difficult to Insert1
  • Difficult to Remove1
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information2

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2025Failure to Align; Failure to Align

    AS REPORTED: "WHEN INSERTING THE DISTAL ADVANCE LOCKING SCREW, THE DISTAL TARGETING DEVICE WAS LOOSE, AND THE SCREW WAS DRIVEN IN AT AN INCORRECT ANGLE BEYOND THE ACCEPTABLE INSERTION RANGE. AS A RESULT, THE SCREW BECAME STUCK WITH THE NAIL AND COULD NOT BE REMOVED. ALTHOUGH IT WAS EVENTUALLY EXTRACTED WITH EFFORT, THE PRODUCTS WERE DAMAGED." 11/06/2025 - ADDITIONAL INFORMATION RECEIVED: HOW WAS THE ISSUE SOLVED? LONGER SCREW DRIVER WAS USED AS AN ALTERNATIVE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23632949 ↗
  • MalfunctionJun 18, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED: "GREY BUTTON FELL OUT OF THE TARGETER, NO LONGER ABLE TO LOCK THE TRIPLE SLEEVE INTO THE TARGETER, UNABLE TO PLACE SCREW IN PREFERRED DISTAL LOCKING HOLE".

    Patient problems: Insufficient Information

    MAUDE report 22288219 ↗
  • MalfunctionNov 23, 2024Difficult to Remove; Mechanical Jam; Difficult to Remove; Mechanical Jam

    AS REPORTED: "WHEN REMOVING THE DISTAL AND PROXIMAL DEVICES AFTER SURGERY, THEY WERE FIRMLY FIXED AND COULD NOT BE REMOVED MANUALLY, SO DR. TRIED TO REMOVE THEM BY HITTING THEM WITH A HAMMER AND DAMAGED THEM".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20766494 ↗
  • InjuryJul 10, 2024Mechanical Jam; Mechanical Jam

    AS REPORTED: "T2 ALPHA FEMUR RETROGRADE DISTAL TARGETING ATTACHMENT MALFUNCTIONED AND WOULD NOT ATTACH TO NAIL HOLDING JIG UNLESS GREAT FORCE WAS APPLIED. ONCE ATTACHED, WITH GREAT EFFORT, IT COULD NOT BE REMOVED. COULD NOT UTILIZE PROXIMAL TARGETING ATTACHMENT, ALSO HAD TO ACQUIRE VARIOUS ADDITIONAL X-RAY ANGLES TO PROPERLY ASCERTAIN NAIL PLACEMENT, SINCE WE COULD NOT REMOVE DISTAL TARGETING ATTACHMENT. BOTH OF THESE CAUSED A PROLONGED PROCEDURE OF 30 MIN."

    Patient problems: Insufficient Information

    MAUDE report 19710015 ↗
  • MalfunctionApr 20, 2023Break; Fracture; Break; Fracture

    AS REPORTED: "THE PIECE THAT HOLDS THE SLEEVE BROKE OFF AND NOW IT WON'T HOLD THE SLEEVES IN PLACE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16778166 ↗
  • MalfunctionDec 20, 2022Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare

    AS REPORTED: "WHEN CHECKING THE DRILL PASSAGE BEFORE INSERTING THE NAIL, DISTAL TARGETING DEVICE COULD NOT BE ATTACHED TO THE NAIL ADAPTER. THE LOCKING LEVER ON THE TARGETING DEVICE DID NOT MOVE AT ALL. DUE TO POOR FUNCTION, THE PROCEDURE WAS CHANGED TO THE USE OF A RADIOLUCENT DRILL AND THE SURGERY WAS COMPLETED.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16014778 ↗
  • MalfunctionSep 14, 2021Difficult to Insert; Failure to Align; Difficult to Insert; Failure to Align

    THE END USER REPORTED THEY HAD A COLINEAR CLAMP ON LATERALLY AND WHEN THE TARGETING WING WAS ATTACHED IT DID HIT THAT CLAMP. THEY ALSO REPORTED THAT WITH ANTEGRADE NAILS, SOME ROC HAVE MIS-MATCHED THE PATIENT AND MADE INSERTION DIFFICULT. NO ADVERSE CONSEQUENCES WERE REPORTED TO THE PATIENT AND NO CLINICALLY SIGNIFICANT DELAY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12470746 ↗
  • MalfunctionSep 10, 2021Failure to Align; Failure to Align

    IT WAS REPORTED THAT ONE ADVANCED LOCKER MISSED THE LATERAL OBLIQUE HOLE. THE FIRST DRILL PASS HIT A TINE OF A LARGE WEBER CLAMP POSSIBLY AFFECTING THE TRAJECTORY OF THE SCREW. THE SURGEON USED A REGULAR LOCKER AND INSERTION WENT COMPLETELY NORMAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12456415 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →