n/a 8000-040-001Recalls & reported problems
Stryker Leibinger GmbH & Co. KG · Patient Tracker Electromagnetic · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 07613327413229), received between Jul 10, 2023 and Jan 23, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement14
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Incorrect Measurement
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 5 EVENTS FOR THE FAILURE: INCORRECT MEASUREMENT. 4 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24163396 ↗MalfunctionOct 28, 2025Incorrect Measurement; Incorrect Measurement
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: INCORRECT MEASUREMENT. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23399851 ↗MalfunctionJul 29, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: INCORRECT MEASUREMENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22646530 ↗MalfunctionApr 21, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. - 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21875901 ↗MalfunctionJan 21, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21201192 ↗MalfunctionJul 24, 2024Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 8 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 7 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19823145 ↗MalfunctionJul 24, 2024Fracture; Fracture
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19822988 ↗MalfunctionApr 22, 2024Incorrect Measurement; Incorrect Measurement
WHILE BEING USED IN A CASE, THE DEVICE WAS INACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19153082 ↗MalfunctionMar 22, 2024Incorrect Measurement; Incorrect Measurement
PER THE ACCOUNT, THE INTERFERENCE PERCENTAGE WOULD NOT GO OVER 85% DURING THE START OF A FUNCTIONAL ENDOSCOPIC SINUS PROCEDURE. LOWER INTERFACE PERCENTAGES DURING THE PROCEDURE COULD POTENTIALLY LEAD TO INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18959601 ↗MalfunctionFeb 1, 2024Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE ACCURACY OF THE DEVICE FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18623719 ↗MalfunctionDec 27, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, DURING A CASE THE TRACKING QUALITY WAS NOT ACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18398025 ↗MalfunctionDec 27, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, DURING A CASE THE TRACKING QUALITY WAS NOT ACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18397851 ↗MalfunctionOct 9, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE PATIENT TRACKERS WOULD NOT GO ABOVE 85% INTERFERENCE CAUSING CONCERN WITH ACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17895023 ↗MalfunctionOct 9, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE PATIENT TRACKERS WOULD NOT GO ABOVE 85% INTERFERENCE CAUSING CONCERN WITH ACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17895021 ↗MalfunctionJul 10, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, WHILE NAVIGATING WITH A SCOPIS PRODUCT, THE DEVICE WAS CLEARLY INACCURATE IN THE LEFT MAXILLARY SINUS. THE SURGEON POINTED THIS OUT AND RE-REGISTERED THE PATIENT TO .45. THE PROBLEM CONTINUED AND A BACKUP DEVICE WAS USED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17290002 ↗
Other Ear, Nose, and Throat Stereotaxic Instrument devices
- TruDi NAV Suction TDNS000ZAcclarent, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall61 reports
- Scopis 8000-021-002Stryker Leibinger GmbH & Co. KG · Ear, Nose, and Throat Stereotaxic Instrument1 recall19 reports
- TruDi™ Navigation System 1Biosense Webster (Israel) Ltd. · Ear, Nose, and Throat Stereotaxic Instrument1 recall14 reports
- FUSION Compact ™ 9735602Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument64 reports
- StealthStation 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- Medtronic Navigation 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- StealthStation Cranial Software 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- Scopis ENT 8000-020-002Stryker Leibinger GmbH & Co. KG · Ear, Nose, and Throat Stereotaxic Instrument33 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page