Skip to content

n/a 8000-040-001Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · Patient Tracker Electromagnetic · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 07613327413229), received between Jul 10, 2023 and Jan 23, 2026 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement14
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2026Incorrect Measurement

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31 2025. THIS REPORT SUMMARIZES 5 EVENTS FOR THE FAILURE: INCORRECT MEASUREMENT. 4 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. 1 EVENT HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163396 ↗
  • MalfunctionOct 28, 2025Incorrect Measurement; Incorrect Measurement

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1 ¿ SEPTEMBER 30 2025. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: INCORRECT MEASUREMENT. - 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399851 ↗
  • MalfunctionJul 29, 2025Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: INCORRECT MEASUREMENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22646530 ↗
  • MalfunctionApr 21, 2025Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. - 3 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21875901 ↗
  • MalfunctionJan 21, 2025Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. - 1 EVENT HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21201192 ↗
  • MalfunctionJul 24, 2024Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 8 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 7 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823145 ↗
  • MalfunctionJul 24, 2024Fracture; Fracture

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE OR CUTTING ACCESSORY FRACTURED. 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19822988 ↗
  • MalfunctionApr 22, 2024Incorrect Measurement; Incorrect Measurement

    WHILE BEING USED IN A CASE, THE DEVICE WAS INACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE SAME DEVICE WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19153082 ↗
  • MalfunctionMar 22, 2024Incorrect Measurement; Incorrect Measurement

    PER THE ACCOUNT, THE INTERFERENCE PERCENTAGE WOULD NOT GO OVER 85% DURING THE START OF A FUNCTIONAL ENDOSCOPIC SINUS PROCEDURE. LOWER INTERFACE PERCENTAGES DURING THE PROCEDURE COULD POTENTIALLY LEAD TO INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18959601 ↗
  • MalfunctionFeb 1, 2024Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER, THE ACCURACY OF THE DEVICE FAILED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18623719 ↗
  • MalfunctionDec 27, 2023Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER, DURING A CASE THE TRACKING QUALITY WAS NOT ACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18398025 ↗
  • MalfunctionDec 27, 2023Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER, DURING A CASE THE TRACKING QUALITY WAS NOT ACCURATE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18397851 ↗
  • MalfunctionOct 9, 2023Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER, THE PATIENT TRACKERS WOULD NOT GO ABOVE 85% INTERFERENCE CAUSING CONCERN WITH ACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17895023 ↗
  • MalfunctionOct 9, 2023Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER, THE PATIENT TRACKERS WOULD NOT GO ABOVE 85% INTERFERENCE CAUSING CONCERN WITH ACCURACY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17895021 ↗
  • MalfunctionJul 10, 2023Incorrect Measurement; Incorrect Measurement

    PER THE CUSTOMER, WHILE NAVIGATING WITH A SCOPIS PRODUCT, THE DEVICE WAS CLEARLY INACCURATE IN THE LEFT MAXILLARY SINUS. THE SURGEON POINTED THIS OUT AND RE-REGISTERED THE PATIENT TO .45. THE PROBLEM CONTINUED AND A BACKUP DEVICE WAS USED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17290002 ↗

Other Ear, Nose, and Throat Stereotaxic Instrument devices

All Ear, Nose, and Throat Stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page