MV3 5900Recalls & reported problems
Stryker Corporation · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Stryker MV3 5900 is a bariatric bed. Its 41 adverse event reports describe malfunctions—mainly incorrect measurements, positioning failures, and missing audible alarms—with no deaths or injuries. There is one open recall, started October 2025, because certain MV3 beds and Arise EX mattresses were sold close together without compatibility testing. Reports dropped sharply, from 10 to 0 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 41
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 07613327419221), received between Jan 28, 2021 and Oct 27, 2025 — classified as 41 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (18 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Oct 31, 2025Open, Classified
Stryker has issued a voluntary recall of certain MV3 beds and Arise 1000EX mattresses because some of these products have been sold in close proximity, suggesting a potential intent for them to be used together. These products have not been tested for compatibility and thus are not designed to be used together. The combination of both is not recommended by Stryker.
Z-0918-2026 · Root cause: Release of Material/Component prior to receiving test results · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect Measurement18
- Positioning Failure11
- No Audible Alarm6
- Device Fell3
- Patient Device Interaction Problem2
- Electrical /Electronic Property Problem1
- Unintended System Motion1
Patient problems
- No Clinical Signs, Symptoms or Conditions41
- Bone Fracture(s)1
- Fall1
- Pain1
Recent adverse event reports
The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 27, 2025Positioning Failure; Positioning Failure
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393620 ↗MalfunctionOct 27, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED PINCH/CRUSH POINT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393619 ↗MalfunctionOct 27, 2025No Audible Alarm; No Audible Alarm
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES, EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393059 ↗MalfunctionJul 24, 2025No Audible Alarm; No Audible Alarm
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22589824 ↗MalfunctionJul 24, 2025Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE LATCH OF SIDERAIL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22589742 ↗MalfunctionApr 28, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INCORRECT MEASUREMENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21922041 ↗MalfunctionApr 28, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED PINCH/CRUSH POINT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: Bone Fracture(s); No Clinical Signs, Symptoms or Conditions; Pain
MAUDE report 21919976 ↗MalfunctionJan 30, 2025Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FALSE LATCH OF THE SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21275575 ↗MalfunctionJan 28, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INACCURATE SCALE READINGS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21255284 ↗MalfunctionJan 28, 2025Incorrect Measurement; Device Fell; Incorrect Measurement; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INACCURATE SCALE READINGS AND UNEXPECTED SIDERAIL COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21254409 ↗MalfunctionOct 29, 2024Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20556480 ↗MalfunctionOct 29, 2024Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INCORRECT MEASUREMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20556479 ↗MalfunctionJul 29, 2024No Audible Alarm; No Audible Alarm
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19850330 ↗MalfunctionJul 26, 2024Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19840625 ↗MalfunctionJul 26, 2024Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19840192 ↗MalfunctionApr 22, 2024Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19159791 ↗MalfunctionJan 24, 2024Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 8 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18573176 ↗MalfunctionJan 24, 2024Unintended System Motion; Unintended System Motion
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED UN-COMMANDED MOTION OF THE ELECTRONIC POSITIONING FUNCTIONS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18573174 ↗MalfunctionOct 25, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
HIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18007331 ↗MalfunctionOct 24, 2023Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17999044 ↗MalfunctionOct 24, 2023Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SCALE MEASURED WEIGHT INACCURATELY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998850 ↗MalfunctionOct 23, 2023Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17990487 ↗MalfunctionJul 19, 2023Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF THE SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17356158 ↗MalfunctionJul 19, 2023Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED INACCURATE SCALE READINGS. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; THERE WAS NO ADVERSE CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17356125 ↗MalfunctionApr 25, 2023Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FALSE LATCH OF SIDERAIL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807255 ↗
Other Bariatric bed devices
- Medline MDR107004Medline Industries, Inc. · Bariatric bed110 reports
- Medline FCEALTMAXXMedline Industries, Inc. · Bariatric bed84 reports
- Synergy® Surface P6881CAPHill-Rom, Inc. · Bariatric bed2 reports
- Drive P503Drive Devilbiss Healthcare · Bariatric bed2 reports
- Joerns Bari10AJoerns Healthcare, LLC · Bariatric Bed2 reports
- Drive 15300BV-PKGDrive Devilbiss Healthcare · Bariatric bed2 reports
- Synergy® Surface P004651Hill-Rom, Inc. · Bariatric bed1 report
- drive PLTCABLDrive Devilbiss Healthcare · Bariatric bed1 death1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record