Skip to content

Medline FCEALTMAXXRecalls & reported problems

Medline Industries, Inc. · ALTERRA MAXX UNIVERSAL BED · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medline Alterra Maxx bariatric bed has 204 adverse event reports, with 4 injuries and no deaths. Most describe electrical faults, such as motors or the control box stopping, and some describe bent or fractured parts and loud noise. Reports rose to 200 in the last 12 months from 1 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
84
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 84 adverse event reports that list this exact device (DI 10888277651838), received between Aug 15, 2018 and Sep 30, 2026 — classified as 4 injuries, 80 malfunctions.

The most frequently reported specific device problem is “Fracture” (33 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 7 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 69 reportsOct 2026: 0 reports20242025202669
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture33
  • Electrical /Electronic Property Problem24
  • Material Twisted/Bent13
  • Break3
  • Crack3
  • Dent in Material3
  • Material Separation3
  • Material Split, Cut or Torn3

Patient problems

  • No Clinical Signs, Symptoms or Conditions77
  • Fall3
  • Laceration(s)3
  • Discomfort1
  • Sleep Dysfunction1

Recent adverse event reports

The 25 most recent of 84 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Fracture; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952197 ↗
  • MalfunctionSep 30, 2026Fracture

    It was reported that, "Frame is cracked. The bed will not raise all the way up."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950866 ↗
  • MalfunctionSep 30, 2026Fracture; Dent in Material; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950842 ↗
  • MalfunctionSep 30, 2026Fracture

    This VMSR report summarizes 4 malfunction events that occurred between October 1 and December 31, 2025. The account reported that beds (FCEALTMAXX) exhibited detached head deck metal frames. Customer-provided photographs showed fractured actuator brackets, fractured welds, and a fractured head actuator on the affected beds. No patient information was available. No patient or user injuries were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945740 ↗
  • MalfunctionSep 30, 2026Failure to Run on Battery

    It was reported that the backup battery failed to kick in when the end user lost power. This is the first time the backup has failed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945480 ↗
  • MalfunctionSep 29, 2026Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938350 ↗
  • MalfunctionSep 29, 2026Fracture; Material Twisted/Bent

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937563 ↗
  • MalfunctionSep 29, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934125 ↗
  • MalfunctionSep 29, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933219 ↗
  • MalfunctionSep 29, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932830 ↗
  • MalfunctionSep 28, 2026Electrical /Electronic Property Problem

    This summary report covers malfunction events that occurred between October 1 and December 31, 2025. This VMSR summarizes one malfunction event involving four Medline Alterra Maxx Universal Beds (catalog number FCEALTMAXX). The account reported that the hi/lo motors moved only about one inch at a time and that the foot-section hi/lo motor was sticking and not moving properly. The devices were not returned, so Medline could not confirm the reported malfunction. The issue did not occur during use; no patient information was available. No patient or user injuries were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914850 ↗
  • MalfunctionSep 25, 2026Material Separation; Material Protrusion/Extrusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905817 ↗
  • MalfunctionSep 25, 2026Fracture; Detachment of Device or Device Component; Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50905403 ↗
  • MalfunctionSep 25, 2026Electrical /Electronic Property Problem; Fracture

    This summary report covers malfunction events that occurred between July 1 and September 30, 2024. This VMSR summarizes one malfunction event involving three Medline Alterra Maxx Universal Beds (catalog number FCEALTMAXX). The account reported broken control boxes and, during troubleshooting, reported a fractured head-motor shaft on one bed, a physically fractured control box on a second bed, and a clicking or intermittently stopping head function associated with a T-cable on a third bed. No devices, photographs, or videos were returned for evaluation; therefore, the reported malfunctions could not be confirmed. No patient information was available. No patient or user injuries were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898812 ↗
  • MalfunctionSep 24, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889034 ↗
  • MalfunctionSep 24, 2026Fracture

    It was reported that height actuator broke.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883298 ↗
  • MalfunctionSep 23, 2026Electrical /Electronic Property Problem

    It was reported that "The light on the back on the pendant is not working. Not the light on the bed."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870397 ↗
  • MalfunctionSep 23, 2026Fracture; Dent in Material

    It was reported that "BROKEN MOTOR/CASTER."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870257 ↗
  • MalfunctionSep 23, 2026Electrical /Electronic Property Problem; Noise, Audible

    It was reported that "The bed is making a clicking noise. Customer stated it may be coming from the motor. Bed is functioning but the noise is very loud. The noise seems to appear when customer is on the bed and the noise comes and goes. Noise started about 3 months ago."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869358 ↗
  • MalfunctionSep 23, 2026Fracture

    It was reported that "THE BED ARRIVED WITH A DAMAGED MOTOR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867054 ↗
  • MalfunctionSep 22, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50859634 ↗
  • MalfunctionSep 22, 2026Electrical /Electronic Property Problem

    It was reported that "Pendent has a couple broken buttons." and "the head section will go up, but doesn't go down all the way, the foot section works fine."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50854058 ↗
  • MalfunctionSep 21, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50848494 ↗
  • MalfunctionSep 21, 2026Fracture; Dent in Material

    It was reported that "CUSTOMER WAS LOWERING THE BED AND SOME HOW IT SMASHED THE CONTROL BOX AND NOW THE BED IS NOT WORKING AT ALL."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846416 ↗
  • MalfunctionSep 21, 2026Key or Button Unresponsive/not Working

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50843059 ↗