Skip to content

Navigation Enabled 48702681Recalls & reported problems

Stryker Corporation · Serrato-ES2-Xia 3 Driver, Cannulated · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327460964), received between Sep 27, 2024 and Mar 8, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Difficult to Insert” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert3
  • Detachment of Device or Device Component1
  • Fracture1
  • Naturally Worn1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Spinal Column Injury1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 8, 2026Fracture; Difficult to Insert; Naturally Worn

    INSPECTION OF THE RETURNED DEVICE INDICATES THAT SEVERAL PRONGS ON THE DEVICE TIP HAD FRACTURED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24542057 ↗
  • MalfunctionSep 3, 2025Difficult to Insert; Difficult to Insert

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22954502 ↗
  • MalfunctionDec 19, 2024Difficult to Insert; Detachment of Device or Device Component; Difficult to Insert; Detachment of Device or Device Component

    IT WAS REPORTED THAT A XIA NAVIGATION ENABLED DRIVER WAS DISENGAGING FROM SCREWS INTRA-OPERATIVELY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20984366 ↗
  • InjurySep 27, 2024Unstable; Unstable

    IT WAS REPORTED THAT A NAVIGATION ENABLED SERRATO/ES2/XIA 3 CANNULATED DRIVER DISENGAGED FROM THE SCREW DURING PLACEMENT AT S1. THIS ALTERED THE TRAJECTORY OF THE SCREW AND MADE THE SCREW HOLE TOO LARGE. THE SURGEON OPTED TO PLACE A SCREW IN THE LEFT ILIAC BONE INSTEAD. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 20-MINUTE SURGICAL DELAY AND NO REPORTED ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: Spinal Column Injury

    MAUDE report 20322654 ↗

FDA clearance and classification

The FDA regulates product code OLO (orthopedic stereotaxic instrument) under 21 CFR 882.4560 as a Class II device. 21 CFR 882.4560 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Navigation Enabled 48702681

Has the Navigation Enabled 48702681 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613327460964), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Navigation Enabled 48702681?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Sep 27, 2024 and Mar 8, 2026. In those reports, the most common specific device problems are “Difficult to Insert” (3), “Detachment of Device or Device Component” (1) and “Fracture” (1); the most common patient problem is “Spinal Column Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Navigation Enabled 48702681?

None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Navigation Enabled 48702681, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand Navigation Enabled, model 48702681). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page