Skip to content

GAMMA 8525-0360SRecalls & reported problems

Stryker GmbH · Long Nail, LEFT · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327474619), received between Feb 27, 2024 and Mar 26, 2026 — classified as 5 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Adverse Event Without Identified Device or Use Problem2
  • Break2
  • Malposition of Device1

Patient problems

  • Nonunion/Delayed-union Bone Fracture2
  • Pain2
  • Bursitis1
  • Failure of Implant1
  • Implant Pain1
  • No Clinical Signs, Symptoms or Conditions1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 26, 2026Fracture

    IT WAS REPORTED THAT THE PATIENT HAD A LONG GAMMA 4 NAIL IMPLANTED FOLLOWING A LEFT INTERTROCHANTETIC HIP FRACTURE IN (B)(6) OF 2025. THE PATIENT LATER REPORTED EXPERIENCING INCREASED PAIN. A FOLLOW UP X-RAY SEEMED TO INDICATE A FRACTURED NAIL. THE DEVICE WAS SUBSEQUENTLY REVISED, AND A RESTORATION MODULAR TOTAL HIP PROSTHESIS WAS IMPLANTED. IT WAS ALSO REPORTED THAT THERE WAS S A POTENTIAL NON-UNION.

    Patient problems: Nonunion/Delayed-union Bone Fracture; Pain

    MAUDE report 24698842 ↗
  • MalfunctionFeb 6, 2026Malposition of Device

    AS REPORTED: "DURING THE GAMMA4 LONG NAIL TECHNIQUE, AFTER INSERTING THE NAIL, WHEN ATTEMPTING TO INSERT THE LAG SCREW, THE PRECISION PIN WAS INSERTED BUT COULD NOT BE POSITIONED CORRECTLY. DESPITE CAREFULLY DRILLING FOR THE LAG SCREW WITHOUT A GUIDE PIN, IT WAS IMPOSSIBLE TO DRILL TO THE APPROPRIATE DEPTH BEYOND THE LAG SCREW HOLE. AS THE DRILL SIMPLY WOULD NOT PASS THROUGH, THE PRECISE LENGTH OF THE LAG SCREW COULD NOT BE DETERMINED. AN APPROXIMATE LENGTH WAS ESTIMATED AND THE LAG SCREW INSERTED, BUT IT TOO STOPPED MIDWAY. AFTER SWITCHING TO A NAIL WITH A DIFFERENT NECK ANGLE AND REINSERTING IT, WE WERE ABLE TO PLACE IT IN THE OPTIMAL POSITION. POSTOPERATIVELY, CALIBRATION USING THE TARGETING DEVICE, NAIL, AND LAG SCREW EMPLOYED FOR INSERTION REVEALED NO ISSUES. CONSEQUENTLY, THE SURGEON CONCLUDED THAT THE DEVICE HAD DEFLECTED DUE TO THE PATIENT'S BONE QUALITY AND SHAPE, PREVENTING INSERTION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24277550 ↗
  • InjuryJun 17, 2025Break; Fracture; Break; Fracture

    AS REPORTED: "POOR BONE UNION AFTER INITIAL SURGERY WITH GAMMA4. REVISION BHA SURGERY WAS PERFORMED ON (B)(6) 2025, DUE TO G4 NAIL FRACTURE. DURING THE INITIAL SURGERY, THE BONE WAS FIXED IN A SLIGHTLY VARUS POSITION. NON-UNION OCCURRED, RESULTING IN A G4 NAIL FRACTURE AT THE LAG SCREW POSITION."

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 22242379 ↗
  • InjuryMay 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "SUBJECT: (B)(6). STRYKER PEGASUS STUDY. DEEP VEIN THROMBOSIS. ACUTE DVT (LEFT PERONEAL VEIN) DIAGNOSED AT ED ON 23/AUG/2023 AFTER A FEW DAYS OF WORSENING LEG SWELLING AND PAIN. DIAGNOSIS VIA ULTRASOUND. PATIENT MEDICATION CHANGED FROM HEPARIN TO ELIQUIS AND DISCHARGED SAME DAY. BY ORTHOPEDIC F/U ON 18/SEP/2023 PATIENT WAS SYMPTOM-FREE."

    Patient problems: Pain; Thrombosis/Thrombus

    MAUDE report 21993350 ↗
  • InjuryApr 8, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED: "A GAMMA4 NAIL IMPLANTED ABOUT 3 MONTHS AGO FAILED AT THE LAG SCREW SITE. THE NAIL WAS COMPLETELY BROKEN INTO 2 PIECES AT THIS POINT IN THE NAIL. OCCURRED WHILE THE PATIENT WAS USING A STATIONARY BIKE. PATIENT HAD TO HAVE A REOPERATION DUE TO THE IMPLANT FAILURE. IMPLANT WAS REMOVED AND REPLACED WITH A TROCH NAIL FROM A DIFFERENT COMPANY."

    Patient problems: Failure of Implant

    MAUDE report 19068083 ↗
  • InjuryFeb 27, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: CLINICAL STUDY PEGASUS SUBJECT (B)(4). ADVERSE EVENT: PAIN. PATIENT EXPERIENCING SYMPTOMS OF GREATER TROCHANTERIC BURSITIS, POSSIBLY FROM HER PROMINENT LAG SCREW ON THE LATERAL BORDER OF THE FEMUR FOLLOWING FX COMPRESSION PATIENT PREVIOUSLY ACHIEVED BONE CONSOLIDATION AT 3 MONTH VISIT ((B)(6) 2023) BOTH CLINICALLY AND RADIOGRAPHICALLY.

    Patient problems: Bursitis; Implant Pain

    MAUDE report 18790254 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates