GAMMA 8425-2320SRecalls & reported problems

Stryker GmbH · Long Nail, RIGHT · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327476415), received between Aug 20, 2025 and Aug 20, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2025Malposition of Device; Malposition of Device

    AS REPORTED: "AFTER PERFORMING A CALIBRATION CHECK AND INSERTING THE GAMMA LONG NAIL ACCORDING TO THE MANUAL, THE LAG SCREW WAS INSERTED. WHEN ATTEMPTING TO ATTACH THE U-GUIDE TO THE LAG SCREW, IT COULD NOT BE ATTACHED. THE SURGEONS CONFIRMED WITH X-RAYS THAT THE LAG SCREW ITSELF HAD SLIPPED OUT OF THE LAG SCREW HOLE AND WAS INSERTED THROUGH THE FRONT OF THE NAIL. (AFTER INSERTING THE NAIL, THE NAIL WAS RETIGHTENED WITH A BALL-TIP SCREWDRIVER, AND A CALIBRATION CHECK WAS ALSO PERFORMED BEFORE INSERTION. ALL PROCEDURES WERE PERFORMED ACCORDING TO THE MANUAL, AND THE PRECISION PIN WAS ALSO INSERTED IN THE CORRECT ORDER THROUGH THE SLEEVE BEFORE PERFORMING THE LAG SCREW DRILL.) THE LAG SCREW THAT COULD NOT BE INSERTED IN THE CORRECT POSITION WAS REMOVED. AFTER SLIGHTLY DEEPENING THE NAIL, THE PRECISION PIN WAS REINSERTED, AND IT ENTERED WHILE INTERFERING WITH THE LAG SCREW HOLE OF THE NAIL. THE LAG SCREW DRILL WAS ALSO RE-DRILLED WHILE INTERFERING WITH THE LAG SCREW HOLE. AFTER THAT, THE REMOVED LAG SCREW WAS REINSERTED, AND THE OPERATION WAS COMPLETED. ALTHOUGH THE CONNECTION IS UNCLEAR, WHEN THE PRECISION PIN WAS FIRST INSERTED AND DRILLING WAS PERFORMED, THERE WAS A BONE FRAGMENT P

    Patient problems: Insufficient Information

    MAUDE report 22848762 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page