GAMMA 8430-1440SRecalls & reported problems
Stryker GmbH · Long Nail, RIGHT · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327476514), received between Sep 11, 2025 and Sep 11, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Fracture1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 11, 2025Break; Fracture; Break; Fracture
AS REPORTED: "THE GAMMA 4 11 × 440 MM (130°) NAIL WAS FOUND TO HAVE FRACTURED AT THE SUPERIOR ASPECT OF THE LAG SCREW ENTRY HOLE. THE NAIL WAS ORIGINALLY IMPLANTED APPROXIMATELY 11 WEEKS AGO. NO INTRAOPERATIVE CONCERNS WERE RAISED AT THE TIME OF INITIAL SURGERY. THE PATIENT REPORTED NO HISTORY OF FALLS OR SIGNIFICANT TRAUMA. MECHANISM OF INJURY (MOI) REMAINS UNKNOWN. INTRAOPERATIVE FINDINGS: FRACTURE OF THE NAIL IDENTIFIED AT THE SUPERIOR ASPECT OF THE LAG SCREW ENTRY HOLE. NAIL EXTRACTION WAS CARRIED OUT WITH MINIMAL DIFFICULTY. NO BROKEN FRAGMENTS WERE FOUND TO REMAIN WITHIN THE PATIENT. CORRECTIVE PROCEDURE: A NEW GAMMA 4 NAIL (11 × 420 MM, 130°) WAS IMPLANTED SUCCESSFULLY DURING THE SAME SURGICAL SESSION."
Patient problems: Insufficient Information
MAUDE report 23029995 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page