Stryker GmbH · Trochanteric Nail · Class 2 · Implant

All models of GAMMA (249)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
1
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327479348), received between Aug 28, 2024 and Mar 23, 2026 — classified as 1 death, 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Malposition of Device1
  • Mechanical Jam1
  • Positioning Failure1

Patient problems

  • Perforation3
  • Fall1
  • Insufficient Information1
  • Necrosis1
  • No Clinical Signs, Symptoms or Conditions1
  • Respiratory Insufficiency1
  • Stroke/CVA1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • DeathMar 23, 2026Mechanical Jam

    AS REPORTED: "UNABLE TO UNSCREW BOLT FROM THE NAIL- TRIED EVERYTHING, HAD TO REMOVE ALL IMPLANTS FROM PATIENT AND PUT IN GAMMA 3 NAIL. A GAMMA 3 11X180X125 NAIL WAS USED TO COMPLETE THE SURGERY".

    Patient problems: Respiratory Insufficiency; Stroke/CVA

    MAUDE report 24670353 ↗
  • MalfunctionJan 22, 2026Positioning Failure

    AS REPORTED: "MIS-DRILLING OCCURRED WHEN DRILLING INTO THE DISTAL SCREW HOLE. THE ACCURACY OF THE TARGET DEVICE WAS CHECKED AFTER SURGERY, AND NO PROBLEMS WERE FOUND."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24152038 ↗
  • InjuryOct 20, 2025Malposition of Device; Malposition of Device

    AS REPORTED: "THE INITIAL SURGERY TOOK PLACE ON (B)(6) 2025, XRAY IMAGE WAS THEN TAKEN ON (B)(6) 2025, ISSUE WAS FOUND THAT THE SHS MIGRATED MEDIALLY. EACH OPERATION WAS PERFORMED BY A DIFFERENT DOCTOR/CLINICAL TEAM; REVISION SURGERY PERFORMED ON (B)(6) 2025. RE-REVISION SURGERY PLANNED FOR (B)(6) 2025." THIS COMPLAINT CAPTURES THE 2ND PLANNED REVISION.

    Patient problems: Perforation

    MAUDE report 23342887 ↗
  • InjuryOct 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "THE LAG SCREW CUT OUT. THE PATIENT HAD BEEN FOLLOWING INSTRUCTIONS. HOWEVER, THIS TIME IT WAS TRAUMATIC AVASCULAR NECROSIS CAUSED BY A FALL. THE EVENT WAS DISCOVERED ON (B)(6) 2025, AND THE NAIL WAS REMOVED AND AN ARTIFICIAL HEAD WAS IMPLEMENTED ON (B)(6) 2025."

    Patient problems: Fall; Necrosis; Perforation

    MAUDE report 23219834 ↗
  • InjurySep 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "AFTER THE OPERATION, IT WAS CUT OUT."

    Patient problems: Perforation

    MAUDE report 23174948 ↗
  • InjuryAug 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED FROM A TROCHANTERIC NAIL GAMMA4 NAIL TO A TOTAL HIP DUE TO THE ROTATION OF LAG SCREW NOT BEING MAINTAINED.

    Patient problems: Insufficient Information

    MAUDE report 20091631 ↗