Skip to content

IsoAir 2941 (REF 2941-001-100)Recalls & reported problems

Stryker Corporation · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327503999), received between Mar 19, 2022 and Mar 19, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Aug 12, 2020Terminated

    Connector between support surface and pump may be out of tolerance resulting in difficulty in removing the CPR connector causing a delay in resuscitation efforts for patients experiencing cardiac arrest and needing CPR

    Z-3015-2020 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Patient Device Interaction Problem1

Patient problems

  • Pressure Sore/Ulcer1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2022Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT SUSTAINED A STAGE 4 PRESSURE INJURY ON AN ISOAIR SUPPORT SURFACE. ATTEMPTS ARE BEING MADE TO GATHER MORE DETAILS FROM THE USER FACILITY.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 13826147 ↗

Other MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE devices

All MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.