Medline MDT24A10Recalls & reported problems
Medline Industries, Inc. · MATTRESS,SYSTEM,A10,LAL/ALTERNATING PRES · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medline alternating pressure air mattress is a hospital bed overlay that inflates and deflates under patients to prevent pressure sores. The 48 reports describe air escaping, the mattress losing air or not inflating and cycling properly, with pumps showing air-loss warning lights and leaking valves. There are 48 malfunctions, no deaths and no injuries, and all 48 reports come from the past 12 months, with none in the year before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 118
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 118 adverse event reports that list this exact device (DI 10193489135671), received between Apr 14, 2026 and Sep 30, 2026 — classified as 2 injuries, 116 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak33
- Inflation Problem32
- Unintended Deflation21
- Deflation Problem15
- Electrical /Electronic Property Problem5
- Decrease in Pressure4
- Failure to Power Up3
- Material Integrity Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions116
- Discomfort1
- Peripheral Edema1
- Pressure Sore/Ulcer1
Recent adverse event reports
The 25 most recent of 118 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Electrical /Electronic Property Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952876 ↗MalfunctionSep 30, 2026Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948465 ↗MalfunctionSep 30, 2026Unintended Deflation
It was reported that "Air mattress is getting deflated from center. Pump shows normal pressure but as soon as patient lies on the mattress it deflates from the center."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50947130 ↗MalfunctionSep 29, 2026Infusion or Flow Problem
It was reported that air in mattress is not getting alternated and is stagnant although the pump is makig the regular noise which it does when it is working. There is no visible damage on the pump or the mattress.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50934866 ↗MalfunctionSep 28, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50925259 ↗MalfunctionSep 28, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915884 ↗MalfunctionSep 28, 2026Inflation Problem
It was reported that ¿the mattress was not staying inflated and that it would slowly lose air.¿ It was also reported that ¿the pump was showing the green light for normal pressure, but once he got in bed it would eventually light up for low pressure and it wouldn¿t return to normal pressure.¿
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915627 ↗MalfunctionSep 28, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50915170 ↗MalfunctionSep 25, 2026Gas/Air Leak
It was reported that mattress wasn't holding air.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899075 ↗MalfunctionSep 25, 2026Inflation Problem
It was reported, "low pressure light keeps coming on and it doesn't turn back on for 20 minutes, so you don't hear air or mattress filling up" and is causing "alarm issues." The customer contact stated, "the mattress sinks in the middle when his wife is on the mattress."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898220 ↗MalfunctionSep 24, 2026Gas/Air Leak
It was reported that the mattress is not inflating properly.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885129 ↗MalfunctionSep 23, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873931 ↗MalfunctionSep 22, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50860407 ↗MalfunctionSep 22, 2026Gas/Air Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859612 ↗MalfunctionSep 21, 2026Unintended Deflation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50849003 ↗MalfunctionSep 21, 2026Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50848998 ↗MalfunctionSep 21, 2026Unintended Deflation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50845340 ↗MalfunctionSep 21, 2026Gas/Air Leak
It was reported that mattress has air leak.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50841947 ↗MalfunctionSep 21, 2026Deflation Problem; Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50841765 ↗MalfunctionSep 21, 2026Pressure Problem; Pumping Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50841661 ↗MalfunctionSep 19, 2026Unintended Deflation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50831853 ↗MalfunctionSep 17, 2026Electrical /Electronic Property Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50809220 ↗InjurySep 17, 2026Deflation Problem; Unintended Deflation
It was reported that beginning on (b)(6) 2026 the "air mattress is not holding air and it leaks out slowly over a few hours of time." According to the customer, "the patient had a small pressure wound which has since been slowly increasing in size" due to the lack of "pressure relief needed from the air mattress."
Patient problems: Peripheral Edema; Pressure Sore/Ulcer
MAUDE report 50804269 ↗MalfunctionSep 17, 2026Electrical /Electronic Property Problem
It was reported that "Static switch no longer working." The customer further stated, "The static switch quit working again. It seems to be working in alternating mode, but not sure."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50801337 ↗MalfunctionSep 16, 2026Inflation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50789994 ↗
Other MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE devices
- Medline MDT24A20Medline Industries, Inc. · MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE267 reports
- Guardian GS-2.5-CTurncare Inc · Mattress, Air Flotation, Alternating Pressure2 recalls15 reports
- IsoAir 2941Stryker Corporation · MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE1 recall1 report
- Arise 1000 EX AREXLAL-CUJoerns Healthcare, LLC · Mattress, Air Flotation, Alternating Pressure1 recall0 reports
- Medline MDT24SUPRAAPLMedline Industries, Inc. · MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE42 reports
- Medline MDT24SUPRADPSMedline Industries, Inc. · MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE40 reports
- Synergy® Surface P6881Hill-Rom, Inc. · MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE28 reports
- NIMBUS 650001DARArjohuntleigh AB · MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE2 deaths17 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record