NA 5407-350-000Recalls & reported problems
Stryker Corporation · PI DRIVE 2 PLUS MOTOR · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 07613327504057), received between May 7, 2022 and Feb 19, 2026 — classified as 20 malfunctions.
The most frequently reported specific device problem is “Fracture” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture15
- Detachment of Device or Device Component3
- Fluid/Blood Leak2
Patient problems
- No Clinical Signs, Symptoms or Conditions20
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT A PIECE OF THE DEVICE WAS MISSING DURING TESTING AT THE USER FACILITY. NO PATIENT INVOLVEMENT, DELAY, MEDICAL INTERVENTION, OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24390994 ↗MalfunctionFeb 4, 2026Detachment of Device or Device Component
A COMPONENT WAS NOTED TO BE MISSING FROM THE DEVICE DURING TESTING AT THE MANUFACTURING FACILITY. NO PATIENT INVOLVEMENT, DELAY, MEDICAL INTERVENTION, OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24257117 ↗MalfunctionNov 19, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY A ROUTER BROKE OFF IN THE DRILL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, MEDICAL INTERVENTION, OR ADVERSE CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23597291 ↗MalfunctionNov 18, 2025Fracture; Fracture
IT WAS REPORTED THAT A BROKEN CUTTING ACCESSORY WAS STUCK IN THE MOTOR AFTER A PROCEDURE AT THE USER FACILITY. NO PATIENT INVOLVEMENT, DELAY, MEDICAL INTERVENTION, OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23581061 ↗MalfunctionJul 7, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22414286 ↗MalfunctionJul 1, 2025Fracture; Fracture
IT WAS REPORTED THAT A PIECE OF BUR WAS STUCK IN THE DEVICE DURING SERVICE AT THE MANUFACTURER FACILITY. NO PATIENT INVOLVEMENT, DELAY, MEDICAL INTERVENTION, OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22376803 ↗MalfunctionJun 27, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING SERVICE CONDUCTED AT THE MANUFACTURER A BROKEN BUR WAS FOUND INSIDE THE MOTOR. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22347240 ↗MalfunctionJun 4, 2025Fracture; Fracture
IT WAS REPORTED THAT THERE WAS A PIECE OF A BUR STUCK IN THE DEVICE DURING EQUIPMENT TESTING CONDUCTED BY A SERVICE TECHNICIAN AT THE MANUFACTURER FACILITY. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22143798 ↗MalfunctionMay 7, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING SERVICE CONDUCTED AT THE MANUFACTURER A BUR PIECE WAS FOUND INSIDE THE MOTOR. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE; NO DELAY OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21975890 ↗MalfunctionMay 7, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING SERVICE CONDUCTED AT THE MANUFACTURER A BUR PIECE WAS FOUND INSIDE THE MOTOR. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE; NO DELAY OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21975820 ↗MalfunctionFeb 24, 2025Fracture; Fracture
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21449678 ↗MalfunctionFeb 10, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING SERVICE CONDUCTED AT THE MANUFACTURER A BROKEN BUR WAS FOUND INSIDE THE DEVICE. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE; NO DELAY OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21343421 ↗MalfunctionAug 21, 2024Fracture; Fracture
IT WAS REPORTED THAT A PROBE BROKE AND THAT IT WAS DISCOVERED PRIOR TO A CASE AND OCCURRED DURING THE STERILIZATION PROCESS. THE THIN SILVER WIRE BROKE AT THE BASE. IT WAS KINKED FOR SEVERAL CYCLES AND THEN IT FINALLY BROKE OFF WHEN IT WAS BEING CLEANED. IT SHOULD BE NOTED THERE WAS NO PROCEDURE OR PATIENT INVOLVEMENT ON THIS RECORD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20021235 ↗MalfunctionAug 14, 2024Fracture; Fracture
THE USER FACILITY REPORTED THAT THE DRILL BIT BROKE DURING REPROCESSING AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19978880 ↗MalfunctionMay 9, 2024Fracture; Fracture
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT THE BIT BROKE OFF INSIDE THE DRILL DURING A CASE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19281070 ↗MalfunctionMar 4, 2024Fracture; Fracture
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THERE WAS A BROKEN BUR IN THE DEVICE WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18832506 ↗MalfunctionSep 19, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT GREASE WAS LEAKING FROM THE DEVICE DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17773078 ↗MalfunctionSep 19, 2023Fluid/Blood Leak; Fluid/Blood Leak
THE MANUFACTURER SALES REPRESENTATIVE REPORTED THAT GREASE WAS LEAKING FROM THE DEVICE DURING A PROCEDURE AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17773076 ↗MalfunctionDec 8, 2022Fracture; Fracture
THE USER FACILITY REPORTED THAT THE BUR BROKE OFF IN THE DRILL DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15944770 ↗MalfunctionMay 7, 2022Fracture; Fracture
A BROKEN CUTTING ACCESSORY WAS FOUND WITHIN THE DEVICE DURING EQUIPMENT TESTING CONDUCTED BY A SERVICE TECHNICIAN AT THE MANUFACTURER FACILITY. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14322526 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page