Skip to content

N/A LF1923Recalls & reported problems

Stryker Sustainability Solutions, Inc. · LigaSure Maryland Jaw, Without Nano-coating, 23cm, Open/Lap Sealer/Divider · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 07613327504880), received between Jul 16, 2020 and Jul 3, 2024 — classified as 3 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem6
  • Energy Output Problem5
  • Activation Problem3
  • Failure to Seal3
  • Difficult to Open or Close1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Hematoma3
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 3, 2024Failure to Seal; Failure to Seal

    IT WAS REPORTED THAT THE LIGASURE WAS NOT SEALING PROPERLY. ALSO THE TISSUE WAS STICKING TO THE JAWS OF THE DEVICE. NO INFORMATION WAS PROVIDED REGARDING PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19668249 ↗
  • MalfunctionJul 3, 2024Failure to Seal; Failure to Seal

    IT WAS REPORTED THAT THE LIGASURE WAS NOT SEALING PROPERLY. ALSO THE TISSUE WAS STICKING TO THE JAWS OF THE DEVICE. NO INFORMATION WAS PROVIDED REGARDING PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19667933 ↗
  • MalfunctionJun 21, 2024Activation Problem; Activation Problem

    IT WAS REPORTED THAT THE LIGASURE DEVICE FAILED TO PERFORM AS INTENDED DURING THE CASE. IT WAS REPLACED WITH ANOTHER LIGASURE DEVICE. THERE WAS NO PATIENT INJURY REPORTED, THE EXTENDED PROCEDURE TIME REPORTED WAS FIVE MINUTES, AND THE COMPLAINANT IS NOT AWARE OF ANY MEDICAL INTERVENTION. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19582288 ↗
  • MalfunctionJun 21, 2024Activation Problem; Activation Problem

    IT WAS REPORTED THAT THE LIGASURE DEVICE FAILED TO PERFORM AS INTENDED DURING THE CASE. IT WAS REPLACED WITH ANOTHER LIGASURE DEVICE. THERE WAS NO PATIENT INJURY REPORTED, THE EXTENDED PROCEDURE TIME REPORTED WAS FIVE MINUTES, AND THE COMPLAINANT IS NOT AWARE OF ANY MEDICAL INTERVENTION. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19582079 ↗
  • MalfunctionJun 21, 2024Activation Problem; Activation Problem

    IT WAS REPORTED THAT THE LIGASURE DEVICE FAILED TO PERFORM AS INTENDED DURING THE CASE. IT WAS REPLACED WITH ANOTHER LIGASURE DEVICE. THERE WAS NO PATIENT INJURY REPORTED, THE EXTENDED PROCEDURE TIME REPORTED WAS FIVE MINUTES, AND THE COMPLAINANT IS NOT AWARE OF ANY MEDICAL INTERVENTION. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19581733 ↗
  • MalfunctionMar 12, 2024Failure to Seal; Failure to Seal

    IT WAS REPORTED THE DEVICE HAD TO BE REPLACED DUE TO THE DEVICE NOT WORKING PROPERLY, EMITTING AN ERROR CODE, AND NOT SEALING TISSUE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18888119 ↗
  • MalfunctionJan 11, 2024Energy Output Problem; Use of Device Problem; Energy Output Problem; Use of Device Problem

    IT WAS REPORTED THE SURGEON WENT TO CLAMP ON, HELD BUTTON, BUT THE DEVICE WOULD NOT SEAL. NO ERROR CODES WERE REPORTED TO HAVE BEEN SEEN. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND THE EXTENDED PROCEDURE TIME REPORTED WAS MINIMAL ONLY TO REPLACE THE DEVICE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18499489 ↗
  • MalfunctionJan 11, 2024Energy Output Problem; Use of Device Problem; Energy Output Problem; Use of Device Problem

    I WAS REPORTED THE DEVICE SEALED FINE TWICE, BUT WHEN USED THREE MINUTES LATER IT WOULD NOT SEAL THREE TIMES IN A ROW. THERE WAS NO FLUID BUILDUP OR TISSUE REPORTED TO HAVE BEEN PREVENTING THE DEVICE FROM SEALING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND THE EXTENDED PROCEDURE TIME REPORTED WAS MINIMAL. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18499049 ↗
  • MalfunctionJan 5, 2024Energy Output Problem; Use of Device Problem; Energy Output Problem; Use of Device Problem

    IT WAS REPORTED THE DEVICE WAS REMOVED FROM THE CASE BECAUSE IT FAILED MULTIPLE TIMES TO COMPLETE CYCLE AND SEEMED TO NOT GET HOT ENOUGH. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18455421 ↗
  • InjurySep 17, 2021Energy Output Problem; Use of Device Problem; Energy Output Problem; Use of Device Problem

    IT WAS REPORTED THERE WAS CONCERN OVER PATIENT HAVING HEMATOMA 10-24 HOURS POST SURGERY AND CONCERNS AT THAT POINT WITH THE SEALING OF THE DEVICE. NO DEVICE WAS SAVED BECAUSE THE DEVICE WORKED PROPERLY DURING THE CASES. THE PATIENT WAS BROUGHT BACK WITHIN THE SAME DAY FOR EVACUATION OF THE HEMATOMA. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Hematoma

    MAUDE report 12492913 ↗
  • InjurySep 17, 2021Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THERE WAS CONCERN OVER PATIENT HAVING HEMATOMA 10-24 HOURS POST SURGERY AND CONCERNS AT THAT POINT WITH THE SEALING OF THE DEVICE. NO DEVICE WAS SAVED BECAUSE THE DEVICE WORKED PROPERLY DURING THE CASES. THE PATIENT WAS BROUGHT BACK WITHIN THE SAME DAY FOR EVACUATION OF THE HEMATOMA. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Hematoma

    MAUDE report 12492912 ↗
  • InjurySep 17, 2021Energy Output Problem; Use of Device Problem; Energy Output Problem; Use of Device Problem

    IT WAS REPORTED THERE WAS CONCERN OVER PATIENT HAVING HEMATOMA 10-24 HOURS POST SURGERY AND CONCERNS AT THAT POINT WITH THE SEALING OF THE DEVICE. NO DEVICE WAS SAVED BECAUSE THE DEVICE WORKED PROPERLY DURING THE CASES. THE PATIENT WAS BROUGHT BACK WITHIN THE SAME DAY FOR EVACUATION OF THE HEMATOMA. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Hematoma

    MAUDE report 12492911 ↗
  • MalfunctionJul 16, 2020Retraction Problem; Difficult to Open or Close; Retraction Problem; Difficult to Open or Close

    ACCORDING TO THE REPORTER, DURING LOBECTOMY, THE TIP BECAME UNABLE TO OPEN WELL HALFWAY, THE CUTTER BECAME DIFFICULT TO RETURN TO INITIAL POSITION. THEY OPENED A NEW DEVICE BUT THE SITUATION WAS THE SAME. THE THIRD DEVICE WAS ABLE TO BE USED WITHOUT PROBLEM. THE PRODUCT DID NOT LOCK ON TISSUE AND WAS REMOVED FROM THE TISSUE WITHOUT DAMAGING IT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10283847 ↗

FDA clearance and classification

Cleared through 510(k) K180499 on Apr 23, 2018 (decision: Substantially Equivalent). Submitted by Stryker Sustainability Solutions as “Reprocessed LigaSure Maryland Jaw Sealer/Divider”. The FDA review took 56 days. FDA record ↗

The FDA regulates product code NUJ (electrosurgical, cutting & coagulation accessories, laparoscopic & endoscopic, reprocessed) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the N/A LF1923

Has the N/A LF1923 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613327504880), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the N/A LF1923?

The FDA's MAUDE database lists 13 adverse event reports for this device, received between Jul 16, 2020 and Jul 3, 2024. In those reports, the most common specific device problems are “Use of Device Problem” (6), “Energy Output Problem” (5) and “Activation Problem” (3); the most common patient problem is “Hematoma” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the N/A LF1923?

None of the reports are classified as deaths, 3 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the N/A LF1923 FDA approved?

It was cleared through 510(k) K180499 on Apr 23, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the N/A LF1923, and is it still sold?

The FDA's device database lists Stryker Sustainability Solutions, Inc. as the company responsible for it (brand N/A, model LF1923). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, REPROCESSED devices

All ELECTROSURGICAL, CUTTING & COAGULATION ACCESSORIES, LAPAROSCOPIC & ENDOSCOPIC, REPROCESSED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page