Skip to content

ProCuity 300900000000Recalls & reported problems

Stryker Corporation · Class 2

All models of ProCuity (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Stryker ProCuity is an adjustable hospital bed. Reports describe alarms that did not sound, unexpected movement or drops, power loss and some patients falling out of bed. There are 111 reports, including 5 injuries and no deaths, and reports in the last 12 months fell to 10 from 25. There is 1 open recall, which began in August 2023 over missing electrical test values that could lead to burns or electric shock.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
111
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 111 adverse event reports that list this exact device (DI 07613327512656), received between Jan 20, 2021 and Jul 28, 2026 — classified as 5 injuries, 106 malfunctions.

The most frequently reported specific device problem is “No Audible Alarm” (22 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 7 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 6 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 4 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 6 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 9 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 9 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 4 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 3 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2023Open, Classified

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

    Z-0424-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • No Audible Alarm22
  • Electrical /Electronic Property Problem16
  • Incorrect Measurement14
  • Intermittent Loss of Power14
  • Unintended System Motion14
  • Device Fell11
  • Patient Device Interaction Problem7
  • Sharp Edges7

Patient problems

  • No Clinical Signs, Symptoms or Conditions100
  • Fall9
  • Laceration(s)2
  • Low Blood Pressure/ Hypotension2
  • Abrasion1
  • Bone Fracture(s)1
  • Bruise/Contusion1
  • Concussion1

Recent adverse event reports

The 25 most recent of 111 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2026Patient Device Interaction Problem

    Patient problems: Physical Entrapment

    MAUDE report 50062278 ↗
  • MalfunctionJul 27, 2026Device Fell

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50045834 ↗
  • InjuryJul 21, 2026No Audible Alarm

    IT WAS ALLEGED THAT A PATIENT FELL OUT OF BED AND THE BED ALARM DID NOT ALERT. IT WAS REPORTED THAT THE PATIENT SUSTAINED A PELVIC FRACTURE THAT WAS TREATED WITH SURGERY.

    Patient problems: Bone Fracture(s)

    MAUDE report 25895807 ↗
  • MalfunctionApr 23, 2026Patient Device Interaction Problem

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED TRIPPING HAZARD; UNEXPECTED PERIMETER. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; THE PATIENT EXPERIENCED A FALL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: Fall

    MAUDE report 24966910 ↗
  • MalfunctionApr 23, 2026No Audible Alarm

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES, EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; THE PATIENT EXPERIENCED A FALL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: Fall; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24966660 ↗
  • MalfunctionApr 23, 2026Device Fell

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24966658 ↗
  • MalfunctionJan 27, 2026Patient Device Interaction Problem

    THIS REPORT NOW SUMMARIZES 2 MALFUNCTION EVENT(S), INSTEAD OF 1. AFTER FURTHER EVALUATION, IT WAS FOUND THAT THE USER WAS FOUND TO HAVE SUSTAINED A LACERATION AND CONCUSSION, BUT NO SERIOUS ADVERSE CONSEQUENCE OCCURRED.

    Patient problems: Concussion; Laceration(s); No Clinical Signs, Symptoms or Conditions

    MAUDE report 24187990 ↗
  • MalfunctionJan 27, 2026Unintended System Motion

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED UNINTENDED DIRECTION OF THE ZOOM; UNINTENDED MOTION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24187989 ↗
  • MalfunctionJan 27, 2026Device Fell

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 - DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24187984 ↗
  • MalfunctionJan 27, 2026No Audible Alarm

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24187983 ↗
  • MalfunctionOct 24, 2025Unintended System Motion; Unintended System Motion

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23384014 ↗
  • MalfunctionOct 24, 2025Device Fell; Device Fell

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23383862 ↗
  • MalfunctionOct 24, 2025No Audible Alarm; No Audible Alarm

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES, EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23383857 ↗
  • MalfunctionJul 23, 2025Unintended System Motion; Unintended System Motion

    THIS REPORT NOW SUMMARIZES 1 MALFUNCTION EVENT(S), INSTEAD OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22578971 ↗
  • MalfunctionJul 23, 2025Device Fell; Device Fell

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED SIDERAIL UNEXPECTED DROP/COLLAPSE OR FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22578731 ↗
  • MalfunctionJul 23, 2025No Audible Alarm; No Audible Alarm

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22578682 ↗
  • MalfunctionJul 23, 2025Intermittent Loss of Power; Intermittent Loss of Power

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ZOOM SUDDENLY STOPS. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22578442 ↗
  • MalfunctionApr 29, 2025Intermittent Loss of Power; Intermittent Loss of Power

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED UNINTENDED DIRECTION OF THE ZOOM. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925782 ↗
  • MalfunctionApr 29, 2025Sharp Edges; Sharp Edges

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP METAL EDGES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21925769 ↗
  • MalfunctionApr 28, 2025Unintended System Motion; Intermittent Loss of Power; Unintended System Motion; Intermittent Loss of Power

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919977 ↗
  • MalfunctionApr 28, 2025Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED INCORRECT MEASUREMENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919845 ↗
  • MalfunctionApr 28, 2025Unintended System Motion; Unintended System Motion

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED UN-COMMANDED MOTION OF THE ELECTRONIC POSITIONING FUNCTIONS. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919814 ↗
  • MalfunctionApr 28, 2025No Audible Alarm; No Audible Alarm

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919643 ↗
  • MalfunctionApr 28, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919642 ↗
  • MalfunctionApr 28, 2025Positioning Failure; Positioning Failure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED LIFT DOES NOT STOP LOWERING. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919641 ↗

Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices

All BED, AC-POWERED ADJUSTABLE HOSPITAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page