8000-021-002, Scopis ENT Software with TGS with Sofware… recallZ-1885-2024 · Stryker Leibinger GmbH & Co. KG · initiated Apr 4, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Leibinger GmbH & Co. KG or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
FDA root cause category: Software Manufacturing/Software Deployment
What the company is doing
On April 22, 2024 an URGENT MEDICAL DEVICE FIELD ACTION letters were sent to customers. Next Steps: 1. Locate any ENU. Quarantine and discontinue use until software on device has been confirmed. - Locate software version and serial number (see pg. 2 for instructions). - Your Stryker Sales Representatives may help, if needed. 2. Return the enclosed Business Reply Form (BRF) via email to [email protected]. - If Scopis ENT Software with TGS version NOVA 3.6.0 RC16 is installed, Stryker will arrange shipment of compliant software. 3. Once compliant software is received and installed, return the enclosed Software Update Form via email to [email protected] to confirm completion of the recall action steps. 4. Maintain this notification internally until all required actions have been completed. Please contact Lacey Marshall at 269-270-5950 or [email protected] with questions or concerns. Report any serious adverse events or product quality problems to Stryker Instruments: 1-800-253-3210. Health care professionals and consumers may report serious adverse events or product quality problems with the use of this product to the FDA's MedWatch Adverse Event Reporting program either online, or by fax or phone. Online: ww.fda.gov/Safety/MedWatch/HowToReport/default.htm Fax: (800) FDA-0178 Phone: (800) FDA-1088.
Recall details
- Recall number
- Z-1885-2024
- Status
- Open, Classified
- Initiated by the company
- Apr 4, 2024
- Posted by the FDA
- May 23, 2024
- Recalling firm
- Stryker Leibinger GmbH & Co. KG
- Quantity
- 705 units
- Product code
- PGW
- FDA submissions
- K221098
- Distribution
- US Nationwide distribution.
Product description
8000-021-002, Scopis ENT Software with TGS with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
Sofware Version: NOVA 3.6.0 RC16, Product Code: 8000-021-002; UDI/DI: 07613327514179; Serial Numbers: 10984, 10110, 10164, 10454, 10762, 10387, 10120, 10252, 10414, 10612, 10408, 10673, 11013, 11032, 10566, 10974, 0310-B, 10326, 10442, 10729, 10681, 10552, 10088, 10476, 10307, 10121, 10371, 0252-B, 10262, 10602, 10613, 10381, 10234, 200774, 10135, 0304-B, 10303, 10257, 10639, 10445, 10684, 10740, 10986, 10413, 10079, 10528, 10565, 10482, 10209, 10574, 10606, 10229, 10214, 10797, 0335-B, 10284, 10465, 10295, 10308, 0282-B, 83B750AD, 0287-B, 10763, 10434, 11025, 10578, 10473, 10596, 0291-B, 10624, 10591, 11092, 10758, 10097, 10294, 0251-B, 10597, 10769, 0309-B, 10765, 0313-B, 0281-B, 10513, 11088, 10057, 10337, 10464, 10621, 10699, 10509, 10421, 10549, 10003, 10499, 10661, 10668, 0292-B, 10416, 0263-B, 10654, 10571, 0329-B, 10934, 10798, 11022, 10610, 10766, 10471, 10437, 10648, 10410, 10403, 10159, 10192, 10372, 10189, 10193, 10642, 10390, 10771, 10202, 10806, 10237, 10296, 0244-B, 0307-B, 10034, 10440, 10704, 10101, 10153, 10098, 10388, 10720, 10943, 10944, 10994, 10695, 10608, 10615, 10054, 10076, 10015, 10089, 10208, 10776, 10780, 11111, 10586, 10759, 10431, 10772, 10376, 10563, 10978, 10989, 10369, 10641, 10796, 10558, 10660, 10071, 10987, 10953, 10982, 10744, 10026, 0325-B, 10959, 11000, 10453, 0332-B, 10055, 10335, 10742, 10649, 10983, 10512, 10958, 10268, 10640, 0280-B, 11075, 11117, 10356, 10429, 10432, 10659, 10677, 10319, 10311, 10332, 10757, 10809, 10523, 10382, 10451, 10495, 10997, 11116, 11018, 10083, 10672, 10775, 10053, 10195, 10330, 10161, 10007, 10328, 10949, 11031, 10781, 10990, 10350, 10750, 10191, 10033, 10056, 10446, 10078, 10516, 10373, 10722, 10136, 10210, 10091, 10255, 10253, 10060, 10427, 10162, 10297, 10731, 10935, 0285-B, 10336, 10760, 10386, 10347, 10286, 10165, 10041, 10456, 10312, 10417, 10500, 10133, 10224, 10031, 10040, 10100, 10784, 10248, 10423, 10215, 10951, 10092, 0260-B, 10185, 10127, 10112, 10790, 10299, 10316, 10435, 10459, 10075, 10479, 10357, 10627, 10732, 10734, 0343-B, 0320-B, 10517, 10455, 10590, 11030, 10383, 10043, 10096, 10789, 10631, 10489, 10791, 10220, 10658, 10348, 10143, 10114, 10625, 10799, 11108, 10103, 10522, 10126, 10066, 10980, 11102, 0318-B, 11005, 10177, 10662, 10166, 10197, 10300, 10406, 10181, 0326-B, 10443, 10331, 10364, 10368, 10592, 10062, 10366, 10349, 10353, 11097, 10155, 10218, 10227, 10230, 10315, 10618, 10021, 10787, 11038, 0271-B, 10365, 10637, 10141, 10525, 10341, 10433, 10696, 0333-B, 10401, 10633, 10211, 10231, 10052, 10794, 10685, 10157, 10213, 10081, 10123, 10090, 10199, 10678, 11020, 10216, 10377, 10424, 10212, 10361, 10059, 10109, 10736, 10344, 10693, 10977, 11004, 10016, 10168, 10171, 10691, 10006, 10458, 10324, 10156, 10190, 10221, 10363, 10475, 0322-B, 0339-B, 10692, 10679, 10082, 10573, 10603, 10267, 10541, 10999, 10169, 10628, 10576, 10345, 10738, 10354, 10238, 0259-B, 10077, 0306-B, 10461, 10172, 10243, 10233, 10327, 0327-B, 11023, 11114, 10450, 10398, 10404, 0303-B, 10400, 10412, 10593, 10072, 10514, 10239, 10069, 10244, 10249, 10761, 10359, 10739, 10352, 10163, 10374, 10698, 10342, 10607, 10635, 10058, 10425, 10419, 10950, 10405, 10422, 10223, 0308-B, 10010, 10594, 10205, 10370, 10474, 10247, 10020, 10174, 10184, 10188, 10128, 11003, 10689, 10701, 10795, 10074, 10266, 10288, 10008, 10203, 10966, 10438, 10955, 10073, 10080, 10333, 10777, 10508, 10325, 10086, 10104, 10469, 10755, 10810, 10671, 10152, 10393, 10657, 10105, 10298, 10524, 10805, 10375, 10402, 10094, 10358, 10498, 10501, 10384, 10391, 10993, 10064, 10175, 10397, 10560, 10570, 10630, 10132, 10579, 10439, 10415, 10306, 10343, 10477, 10323, 10737, 10589, 10304, 10338, 10460, 10598, 10779, 10988, 10346, 11082, 10436, 10362, 10334, 10682, 10653, 10583, 10724, 10725, 10407, 10380, 10409, 10595, 10137, 10656, 10389, 10611, 10638, 11036, 11037, 10577, 10609, 10587, 10626, 10965, 11017, 11112, 10547, 10954, 11029, 11026, 10979, 10985, 11080, 10981, 11099, 10207, 10049, 10580, 10144, 10723.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Scopis 8000-021-002Stryker Leibinger GmbH & Co. KG · Ear, Nose, and Throat Stereotaxic Instrument1 recall19 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page