Skip to content

iBed Wireless 521200380100Recalls & reported problems

Stryker Corporation · Secure Connect Locator · Class 2

Adverse event reports
20
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 07613327520453), received between Jun 4, 2021 and Jan 27, 2026 — classified as 20 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem17
  • No Audible Alarm3

Patient problems

  • No Clinical Signs, Symptoms or Conditions19
  • Fall1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 27, 2026No Audible Alarm

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL. THERE WAS A PATIENT FALL, BUT NO SERIOUS ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: Fall

    MAUDE report 24189144 ↗
  • MalfunctionOct 7, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THE POWER CORD PRONGS ARE GETTING STUCK IN THE OUTLETS. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23234719 ↗
  • MalfunctionApr 28, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 46 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21921701 ↗
  • MalfunctionApr 28, 2025No Audible Alarm; No Audible Alarm

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21921501 ↗
  • MalfunctionJan 30, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21275383 ↗
  • MalfunctionOct 29, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 20 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20556465 ↗
  • MalfunctionJul 29, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; NO ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19850293 ↗
  • MalfunctionApr 23, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 9 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19163125 ↗
  • MalfunctionJan 24, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 11 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18572078 ↗
  • MalfunctionOct 24, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES (B)(4) MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17999045 ↗
  • MalfunctionJul 24, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17378210 ↗
  • MalfunctionApr 26, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 149 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16817456 ↗
  • MalfunctionOct 24, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15659077 ↗
  • MalfunctionJul 27, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15112242 ↗
  • MalfunctionApr 26, 2022No Audible Alarm; No Audible Alarm

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE CLINICIAN WAS NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14209528 ↗
  • MalfunctionApr 26, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14206982 ↗
  • MalfunctionJan 20, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 40 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13306771 ↗
  • MalfunctionOct 26, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 32 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED EXPOSED BARE WIRES OR ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12697025 ↗
  • MalfunctionJul 28, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 32 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12238229 ↗
  • MalfunctionJun 4, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THE CHARGING BRICK FOR THE SECURE CONNECT LOCATOR BROKE OFF AT THE POWER PRONGS AFTER BEING BUMPED, LEADING TO ACCESSIBLE AC CURRENT. THERE ARE NO REPORTS OF PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11940857 ↗

Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices

All BED, AC-POWERED ADJUSTABLE HOSPITAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.