Skip to content

ProCuity 3009PX-L-100Recalls & reported problems

Stryker Corporation · ProCuity L + ProForm · Class 2

All models of ProCuity (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327541168), received between Apr 22, 2024 and Jan 30, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2023Open, Classified

    Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.

    Z-0425-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Electrical /Electronic Property Problem1
  • Incorrect Measurement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 30, 2025Incorrect Measurement; Incorrect Measurement

    THIS REPORT SUMMARIZES 0 MALFUNCTION EVENTS, INSTEAD OF 1.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21274468 ↗
  • MalfunctionApr 22, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19159991 ↗

Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices

All BED, AC-POWERED ADJUSTABLE HOSPITAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page