ProCuity 3009PX-LEX-400Recalls & reported problems
Stryker Corporation · ProCuity LEX + IsoTour Std · Class 2
All models of ProCuity (17)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Stryker ProCuity LEX hospital bed is an adjustable bed with a Lex positioning system. Most of the 67 reports describe the bed's backrest dropping unexpectedly, the bed-to-bed exit alarm not sounding, the bed rolling or moving on its own, and exposed sharp metal edges, along with incorrect measurements and electrical problems. The reports include 1 death and 2 injuries. One recall, opened in August 2023 and still open, concerns beds missing electrical safety test values, which the company says poses a risk of tissue burn or electrical shock.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 67
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 67 adverse event reports that list this exact device (DI 07613327541229), received between Apr 23, 2021 and Apr 23, 2026 — classified as 1 death, 2 injuries, 64 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (14 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 30, 2023Class IIOpen, Classified
Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.
Z-0429-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect Measurement14
- No Audible Alarm13
- Electrical /Electronic Property Problem12
- Device Fell10
- Unintended System Motion8
- Sharp Edges3
- Mechanical Problem2
- Patient Device Interaction Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions62
- Fall4
- Pain2
- Bone Fracture(s)1
- Cardiac Arrest1
- Hematoma1
- Laceration(s)1
Recent adverse event reports
The 25 most recent of 67 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24966912 ↗DeathApr 9, 2026No Audible Alarm
IT WAS REPORTED THAT THE BED EXIT FAILED TO ALARM WHEN THE PATIENT GOT OUT OF THE BED. THE PATIENT WAS FOUND BY STAFF ON THE FLOOR IN A STATE OF CARDIAC ARREST. THE PATIENT REPORTEDLY PASSED AWAY.
Patient problems: Cardiac Arrest; Fall
MAUDE report 24832929 ↗MalfunctionJan 27, 2026No Audible Alarm
NO NEW INFORMATION.
Patient problems: Fall; No Clinical Signs, Symptoms or Conditions
MAUDE report 24187995 ↗MalfunctionJan 27, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24187994 ↗MalfunctionOct 24, 2025Sharp Edges; Sharp Edges
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP EDGES EXPOSED (METAL). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23384146 ↗MalfunctionOct 24, 2025Positioning Failure; Positioning Failure
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23384137 ↗MalfunctionOct 24, 2025Device Fell; Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 9 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23384136 ↗MalfunctionJul 25, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED INCORRECT MEASUREMENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608938 ↗MalfunctionJul 23, 2025Device Fell; Device Fell
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22578953 ↗MalfunctionJul 23, 2025Unintended System Motion; Unintended System Motion
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED UN-COMMANDED MOTION OF THE ELECTRONIC POSITIONING FUNCTIONS. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22578884 ↗MalfunctionJul 23, 2025No Audible Alarm; No Audible Alarm
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22578737 ↗MalfunctionApr 28, 2025Device Fell; Device Fell
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919847 ↗MalfunctionApr 28, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INCORRECT MEASUREMENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919815 ↗MalfunctionApr 28, 2025No Audible Alarm; No Audible Alarm
THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES, EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT WITH A FALL; NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: Fall; No Clinical Signs, Symptoms or Conditions
MAUDE report 21919647 ↗MalfunctionApr 28, 2025Grounding Malfunction; Grounding Malfunction
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED GROUND PATH COMPROMISED. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919645 ↗MalfunctionApr 28, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919640 ↗MalfunctionJan 30, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE A PATIENT HAD SUSTAINED A MINOR INJURY AFTER CUTTING THEIR LEG ON A SMALL METAL TAB THAT IS PART OF THE PRODUCT'S DESIGN.
Patient problems: Laceration(s); Pain
MAUDE report 21274356 ↗MalfunctionJan 30, 2025No Audible Alarm; No Audible Alarm
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED A FAILURE TO ALARM. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21274355 ↗MalfunctionJan 30, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED INACCURATE SCALE READINGS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21274354 ↗MalfunctionJan 28, 2025Unintended System Motion; Unintended System Motion
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED UN-COMMANDED MOTION OF THE ELECTRONIC POSITIONING FUNCTIONS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21251427 ↗MalfunctionJan 28, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21251425 ↗MalfunctionJan 28, 2025Device Fell; Device Fell
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SIDERAIL OR FOWLER COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21251423 ↗MalfunctionOct 29, 2024Positioning Failure; Positioning Failure
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT, WHERE IT WAS REPORTED THE DEVICE EXPERIENCED LIFT DOES NOT STOP LOWERING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560262 ↗MalfunctionOct 25, 2024Device Fell; Device Fell
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FOWLER UNEXPECTED DROP/COLLAPSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20540007 ↗MalfunctionOct 25, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE AC CURRENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20536228 ↗
FDA clearance and classification
The FDA regulates product code FNL (bed, ac-powered adjustable hospital) under 21 CFR 880.5100 as a Class II device. 21 CFR 880.5100 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ProCuity 3009PX-LEX-400
Has the ProCuity 3009PX-LEX-400 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0429-2024 (Class II), began on Aug 30, 2023: Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ProCuity 3009PX-LEX-400?
The FDA's MAUDE database lists 67 adverse event reports for this device, received between Apr 23, 2021 and Apr 23, 2026. In those reports, the most common specific device problems are “Incorrect Measurement” (14), “No Audible Alarm” (13) and “Electrical /Electronic Property Problem” (12); the most common patient problems are “Fall” (4), “Pain” (2) and “Bone Fracture(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ProCuity 3009PX-LEX-400?
1 report is classified as a death, 2 as injuries and 64 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ProCuity 3009PX-LEX-400 FDA approved?
Devices of this type (product code FNL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker Corporation.
Who makes the ProCuity 3009PX-LEX-400, and is it still sold?
The FDA's device database lists Stryker Corporation as the company responsible for it (brand ProCuity, model 3009PX-LEX-400). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,831 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,320 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,149 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths333 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death330 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths354 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL299 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page