ProCuity 3009PX-ZMX-450Recalls & reported problems
Stryker Corporation · ProCuity ZMX + IsoTour Premium · Class 2
All models of ProCuity (17)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327541304), received between Oct 24, 2025 and Oct 24, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “No Audible Alarm” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 30, 2023Class IIOpen, Classified
Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.
Z-0434-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- No Audible Alarm1
Patient problems
- Fall1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2025No Audible Alarm; No Audible Alarm
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED CLINICIAN NOT ALERTED/AWARE OF POSSIBLE PATIENT FALL CONDITION. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT THAT EXPERIENCED A FALL; NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: Fall
MAUDE report 23384144 ↗
Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,937 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,377 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,193 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death374 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths362 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths358 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL317 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page