GAMMA 8325-0240SRecalls & reported problems
Stryker GmbH · Intermediate Nail, LEFT · Class 2 · Implant
All models of GAMMA (249)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07613327593877), received between Jan 31, 2024 and Jul 28, 2026 — classified as 4 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Positioning Failure” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Positioning Failure4
- Fracture3
- Device Damaged by Another Device2
- Improper or Incorrect Procedure or Method1
- Malposition of Device1
- Unstable1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information4
- Perforation1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJul 28, 2026Fracture
As reported: "Primary surgery was performed on (b)(6)2026. The nail broke off at the lag screw hole during postoperative follow-up.Revision surgery was performed on (b)(6)-2026, and Gamma4 was removed and DC nail of Hansson was implanted."
Patient problems: Insufficient Information
MAUDE report 50061358 ↗MalfunctionJul 20, 2026Positioning Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25881382 ↗InjuryJul 6, 2026Unstable
It was reported that, postoperatively, the distal screw came loose. A revision surgery was performed to replace the nail with a long nail. No further information was provided.
Patient problems: Insufficient Information
MAUDE report 25724687 ↗MalfunctionApr 30, 2026Positioning Failure
AS REPORTED: "THE GAMMA4 INTERMEDIATE NAIL WAS USED. DURING DRILLING FOR DISTAL LOCKING SCREW INSERTION, INTERFERENCE BETWEEN THE DRILL BIT AND THE NAIL WAS OBSERVED. DESPITE COMPLETING DRILLING TO THE CONTRALATERAL CORTEX AFTER MINOR ADJUSTMENT, ORBITING OCCURRED DURING SCREW INSERTION, RESULTING IN FAILURE TO INSERT ONE OF THE TWO DISTAL LOCKING SCREWS. DURING INSERTION OF THE DISTAL LOCKING SCREW, MISALIGNMENT FROM THE INTENDED POSITION OCCURRED, POTENTIALLY CAUSING UNNECESSARY ADDITIONAL BONE/TISSUE TRAUMA TO THE PATIENT."
Patient problems: Perforation
MAUDE report 25046345 ↗MalfunctionApr 20, 2026Malposition of Device
AS REPORTED: "AFTER INSERTING THE GAMMA4 INTERMEDIATE NAIL AND DRILLING FOR THE DISTAL LOCKING SCREW, THE DRILL BIT AND THE NAIL INTERFERED WITH EACH OTHER, AND AS A RESULT, WE WERE UNABLE TO INSERT ONE OF THE TWO LOCKING SCREWS."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24932110 ↗InjuryMar 19, 2026Fracture
AS REPORTED: "THE PATIENT UNDERWENT A REVISION SURGERY DUE TO THE NAIL BREAKING AT THE LAG SCREW JUNCTION".
Patient problems: Insufficient Information
MAUDE report 24640810 ↗MalfunctionMar 10, 2026Positioning Failure
AS REPORTED: "DURING THE PRE-DRILLING INTERFERENCE CHECK, THE DRILL BIT WAS FOUND TO HAVE SLIGHT INTERFERENCE WITH THE DISTAL LATERAL STOP HOLE. IT WAS USED AS IS, AND DURING DRILLING, THE BIT DEVIATED FROM THE HOLE, SO DRILLING WAS PERFORMED WHILE HOLDING THE SLEEVE. THE PRIMARY COMPLAINT IS THAT THE INTERMEDIATE TARGETING SLEEVE FELT STIFF DURING USE."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24560302 ↗MalfunctionDec 17, 2025Positioning Failure; Positioning Failure
AS REPORTED: "OUT OF BONE DURING DRILLING IN THE DISTAL DYNAMIC HOLE. RETRIED WITH THE DEVICE AND THE PROCEDURE WAS COMPLETED."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23834447 ↗InjuryDec 3, 2025Fracture; Device Damaged by Another Device; Fracture; Device Damaged by Another Device
AS REPORTED: "LEFT INTERMEDIATE GAMMA NAIL BROKE AT THE STATIC SCREW HOLE".
Patient problems: Insufficient Information
MAUDE report 23707111 ↗MalfunctionOct 18, 2024Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED AFTER THE FIRST USE OF GAMMA 4 WITHOUT THE PRESENCE OF THE SALES REP, THE DOCTOR NOTICED THAT HE FORGOT TO TIGHTEN THE SET SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20483716 ↗MalfunctionJan 31, 2024Device Damaged by Another Device; Device Damaged by Another Device
AS REPORTED: "G4 INTERMEDIATE NAIL MISS DRILL WAS OCCURRED ON LATERAL SIDE DURING DRILLING OF DISTAL DYNAMIC HOLE. PRE-CALIBRATION SHOWED CORRECT TARGETING. THE CORTICAL BONE AND NAIL WERE IN CONTACT AT THE LATERAL END, AND THE PHYSICIAN DETERMINED THAT THE MISS DRILL WAS DUE TO LOADING ON THE NAIL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18620127 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page