Skip to content

PANGEA 542021Recalls & reported problems

Stryker GmbH · Drill Bit, AO · Class 1

All models of PANGEA (53)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327614541), received between Jun 21, 2024 and Sep 10, 2026 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture6
  • Break4

Patient problems

  • Foreign Body In Patient4
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Perforation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2026Fracture

    As reported: "A drill bit for the Pangea small frag system broke today during use.The tip of the drill bit was reported and broke off in the patient's bone. Was not removed from patient."

    Patient problems: Foreign Body In Patient

    MAUDE report 50715194 ↗
  • InjuryMay 29, 2026Fracture

    THE DRILL BIT BROKE DURING A PANGEA TIBIAL SURGERY. IT BROKE WHEN IT CAME INTO CONTACT WITH THE PLATE ON THE OPPOSITE SIDE DURING A DOUBLE-PLATE PROCEDURE. THE SURGERY WAS COMPLETED WHILE THE TIP OF THE BROKEN DRILL BIT REMAINED INSIDE THE PATIENT'S BODY.

    Patient problems: Foreign Body In Patient

    MAUDE report 25332368 ↗
  • InjurySep 22, 2025Break; Fracture; Break; Fracture

    AS REPORTED: "MULTIPLE DRILL BITS BROKE CAUSING SKIVING AND DRILLING LARGER HOLES. A DELAY OCCURRED IN ORDER TO GET ANOTHER DRILL BIT; HOWEVER THE LENGTH OF THE DELAY WAS NOT PROVIDED".

    Patient problems: Perforation

    MAUDE report 23116428 ↗
  • MalfunctionJul 8, 2025Break; Fracture; Break; Fracture

    AS REPORTED VIA VOLUNTARY MEDWATCH REPORT: TWO DIFFERENT DRILL BITS USED DURING ORTHOPEDIC SURGERY BROKE IN PATIENT'S BONE.

    Patient problems: Foreign Body In Patient; Insufficient Information

    MAUDE report 22441305 ↗
  • MalfunctionNov 4, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT WHILST THE PATIENT WAS BEING IMPLANTED WITH A PLATE, THE DRILL BIT BROKE OFF INTO THE BONE. SURGEON DECIDED TO LEAVE THE BIT FRAGMENT IN PLACE. THE PROCEDURE WAS COMPLETED, AND PATIENT IS RECOVERING.

    Patient problems: Foreign Body In Patient

    MAUDE report 20604022 ↗
  • MalfunctionJun 21, 2024Break; Fracture; Break; Fracture

    IT WAS REPORTED: THE SURGEON WAS USING THE 2.5MM DRILL BIT, SKU 542021, AND WHILE HE WAS DRILLING IN THE FAR CORTEX OF THE MEDIAL DISTAL TIBIA, THE DRILL BIT BROKE AT THE JUNCTION OF THE CUTTING FLUTES AND THE SHAFT OF THE DRILL. IT SEEMED THAT THE DRILL BIT WAS TOO FLEXIBLE AND WOULD NOT PENETRATE THROUGH THE CORTEX. FROM THAT FORCE AND CURVATURE OF THE BONE, THE DRILL BIT BROKE AT THE SITE LISTED ABOVE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19580674 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →