Skip to content

PANGEA 542021SRecalls & reported problems

Stryker GmbH · Drill Bit, AO · Class 1

All models of PANGEA (53)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327614596), received between Jun 30, 2026 and Aug 28, 2026 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Foreign Body In Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 28, 2026Fracture

    As reported: "During a Pangea PH procedure, drilling was performed through the posterior screw hole in the second row from the proximal side. After screw insertion, fluoroscopic/X-ray confirmation revealed that the tip of the drill bit had fractured and remained within the bone.The surgeon determined that removal of the fractured drill bit tip was not feasible and therefore completed the procedure with the fragment left in situ."

    Patient problems: Foreign Body In Patient

    MAUDE report 50500183 ↗
  • InjuryJun 30, 2026Fracture

    AS REPORTED: "DURING DRILLING, THE TIP OF THE DRILL BROKE OFF DUE TO INTERFERENCE WITH AN ALREADY INSERTED SCREW, LEAVING ABOUT 2MM INSIDE THE BODY. IT COULD NOT BE REMOVED, SO THE WOUND WAS CLOSED IN THAT STATE."

    Patient problems: Foreign Body In Patient

    MAUDE report 25668249 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page