N/A 6002-810-000Recalls & reported problems

Stryker Leibinger GmbH & Co. KG · Spine Guidance 5 Software · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 07613327616118), received between Apr 2, 2025 and Sep 16, 2026 — classified as 1 injury, 13 malfunctions.

The most frequently reported specific device problem is “Imprecision” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Imprecision12
  • Fail-Safe Did Not Operate1
  • Image Orientation Incorrect1
  • Incorrect Measurement1
  • Program or Algorithm Execution Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • Hemorrhage/Blood Loss/Bleeding1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Image Orientation Incorrect

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50785058 ↗
  • MalfunctionMar 25, 2026Incorrect Measurement

    IT WAS REPORTED DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE, THE SCREWS SHOWED ON NAVIGATION TO BE IN PLACE AND ON TRAJECTORY BUT WHEN XRAYS WERE TAKEN THE ENTIRE LEFT SIDE WAS MEDIAL, RIGHT SIDE SCREWS WERE FINE. NO SURGICAL DELAY OR ANY ADVERSE CONSEQUENCES WERE REPORTED AT INTAKE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24686702 ↗
  • MalfunctionFeb 12, 2026Imprecision

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE, AFTER THE LEFT L5 SCREW WAS PLACED, NEUROMONITORING INFORMED THE SURGEON THAT THE MOTORS DROPPED BY A SIGNIFICANT AMOUNT. THE SURGEON PROCEEDED TO REMOVE THE LEFT L5 SCREW AND PLACED A PEDICLE FINDER THROUGH THAT CANNULATED HOLE USING NAVIGATION AND IT WAS SHOWN THAT THE INSTRUMENT WAS THROUGH THE FOREMEN. THE SURGEON PROCEEDED TO DO A DECOMPRESSION AND CONTINUE WITH PLACING THE REMAINING SCREWS, USING ONLY THE INTRAOPERATIVE SCANS FOR THE REMAINING SCREWS. THE REMAINDER OF THE SCREWS WERE ACCURATE WHICH WAS CONFIRMED BY X-RAYS. THERE WERE NO ADVERSE CONSEQUENCES AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24324181 ↗
  • MalfunctionFeb 12, 2026Imprecision; Program or Algorithm Execution Failure

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE, THE NAVIGATION PLAN DISPLAYED THE SCREW AT T3 DESPITE IT BEING AT T4. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24324055 ↗
  • MalfunctionFeb 12, 2026Imprecision

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE, THE TRAJECTORY WAS DISPLAYING LATERAL AND DID NOT ALIGN WITH THE PRE-PLANNED SCREW. THE SCREW APPEARED LATERAL ON IMAGING AND THE SCREW AS REMOVED. THERE WERE NO ADVERSE CONSEQUENCES AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF OVER 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24322643 ↗
  • MalfunctionNov 6, 2025Imprecision; Imprecision

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE, THE SCREWS WERE MISPLACED AND REQUIRED AN IMMEDIATE REVISIONS AND THERE WAS ANTERIOR DISPLACEMENT OF THE ALIF CAGE. A NEW SCAN WAS PERFORMED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23493196 ↗
  • MalfunctionOct 7, 2025Imprecision; Imprecision

    IT WAS REPORTED THAT DURING A CADAVER LAB USING THE SPINE GUIDANCE 5 SOFTWARE, THE NAVIGATED INSTRUMENTS WERE INACCURATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23240644 ↗
  • MalfunctionSep 18, 2025Imprecision; Imprecision

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE SOFTWARE, THERE WERE INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES AND A SURGICAL DELAY OF LESS THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23089093 ↗
  • MalfunctionAug 15, 2025Imprecision; Imprecision

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE, WHEN MERGING THE VERTEBRAL LEVEL IT WAS COMPLETELY OFF. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22820303 ↗
  • MalfunctionAug 4, 2025Imprecision; Imprecision

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE THERE WAS AN INACCURACY OF 3.7 MILLIMETERS FROM THE DISTAL TIP OF THE POINTER TO THE FIDUCIAL SCREW. THERE WAS A SURGICAL DELAY OF LESS THAN 30 MINUTES AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22710196 ↗
  • InjuryJun 9, 2025Imprecision; Imprecision

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE, THE SCREWS WERE MISPLACED CAUSING AN ANTERIOR BREACH. THERE WAS A SURGICAL DELAY OF OVER 120 MINUTES AND THE PATIENT EXPERIENCED BLEEDING AND BLOOD CLOTS.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Thrombosis/Thrombus

    MAUDE report 22171583 ↗
  • MalfunctionApr 17, 2025Imprecision; Imprecision

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21858541 ↗
  • MalfunctionApr 15, 2025Imprecision; Fail-Safe Did Not Operate; Imprecision; Fail-Safe Did Not Operate

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21840498 ↗
  • MalfunctionApr 2, 2025Imprecision; Imprecision

    IT WAS REPORTED THAT DURING A CASE USING THE SPINE GUIDANCE 5 SOFTWARE, AN X-RAY DISPLAYED THAT THE SCREWS WERE MEDIAL AND BREACHED. THE SCREWS WERE REMOVED AND PLACED AGAIN USING NAVIGATION. THERE WERE NO ADVERSE CONSEQUENCES AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21757860 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page