Patient-Fitted TMJ Reconstructive Prosthesis System CHG010Recalls & reported problems

TMJ Solutions, Inc. · TMJ Unilateral Implant · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This TMJ Solutions patient-fitted jaw joint implant has 94 reports, including 72 injuries and no deaths. Common reports describe poor fit of guides and splints during surgery, screws coming loose, unwanted bone growth requiring revision, and infections. The 1 open recall, from October 2025, concerns screw holes in the wrong position on some implants. Reports rose from 15 in the prior 12 months to 31 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
94
Deaths reported
0
Injuries reported
71
Recalls naming it
1

Summary

The FDA's MAUDE database contains 94 adverse event reports that list this exact device (DI 07613327626551), received between May 30, 2023 and Sep 23, 2026 — classified as 71 injuries, 23 malfunctions.

The most frequently reported specific device problem is “Nonstandard Device” (16 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 3 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 5 reportsApr 2024: 3 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 3 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 12 reportsJan 2026: 0 reportsFeb 2026: 9 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 2 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 13, 2025Class IOpen, Classified

    Patient-fitted temporomandibular joint implant contained discrepancy in screw hole position.

    Z-0488-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem49
  • Nonstandard Device16
  • Device Appears to Trigger Rejection13
  • Inadequacy of Device Shape and/or Size9
  • Device Dislodged or Dislocated6
  • Fracture6
  • Patient-Device Incompatibility3
  • Detachment of Device or Device Component2

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • Unspecified Infection28
  • Pain22
  • Swelling/ Edema12
  • Joint Dislocation10
  • Ankylosis9
  • Bacterial Infection9
  • Insufficient Information8

Recent adverse event reports

The 25 most recent of 94 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Device Dislodged or Dislocated

    It was reported that a revision surgery for a right side non-Stryker TMJ implant is required due to pain. During the revision planning of the non-Stryker TMJ implant on the right mandible, the surgeon observed that the contralateral left mandibular prosthesis was close to lateral dislocation due to a loose screw. The left-side implant was not the reason to request the revision surgery, since it does not create any symptoms for the patient and was considered as an incidental finding.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50864544 ↗
  • MalfunctionSep 15, 2026Inadequacy of Device Shape and/or Size

    It was reported that the fossa component could not be properly seated intraoperatively on either the bone or the manufacturing model. The component was unstable and rocked, requiring modification of the surrounding bone to achieve acceptable seating. Postoperative imaging showed areas where full adaptation was not achieved. In addition, remarks were reported regarding screw selection.

    Patient problems: Insufficient Information

    MAUDE report 50768321 ↗
  • MalfunctionJul 14, 2026Malposition of Device

    IT WAS REPORTED THAT TARGET OCCLUSION AND PROSTHESIS POSITION COULD NOT BE ACHIEVED AS PLANNED, RESULTING IN MALOCCLUSION.

    Patient problems: Unspecified Musculoskeletal problem

    MAUDE report 25809200 ↗
  • InjuryJun 24, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING REVISION SURGERY, THE PLANNED EXPLANTATION OF THE MANDIBULAR COMPONENT WAS NOT POSSIBLE DUE TO BONE GROWTH, SO THE SURGICAL PLAN WAS CHANGED TO REMOVAL OF HETEROTOPIC BONE ONLY AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Ankylosis

    MAUDE report 25606357 ↗
  • InjuryJun 16, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE FOSSA COMPONENT SCREWS ARE COMING LOOSE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 25514685 ↗
  • InjuryMay 28, 2026Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT DISCREPANCIES CAUSED INACCURATE SURGICAL CUTS AND IMPROPER FIT OF THE TMJ AND CUSTOM PLATES, ULTIMATELY REQUIRING ABANDONMENT OF THE CUSTOM ORTHOGNATHIC PORTION TO ENSURE STABILITY AND ACCURATE JOINT POSITIONING.

    Patient problems: Insufficient Information; Unspecified Musculoskeletal problem

    MAUDE report 25315526 ↗
  • MalfunctionMay 26, 2026Noise, Audible

    IT WAS REPORTED THAT THE JOINT REPLACEMENT SQUEAKS, BUT THE PATIENT HAS NO PAIN OR OTHER PROBLEMS.

    Patient problems: Discomfort; Unspecified Musculoskeletal problem

    MAUDE report 25278853 ↗
  • InjuryMar 2, 2026Device Appears to Trigger Rejection

    No new information.

    Patient problems: Unspecified Infection

    MAUDE report 24489324 ↗
  • MalfunctionFeb 26, 2026Nonstandard Device

    WITHIN THE SCOPE EVALUATION FOR A PRODUCT FIELD ACTION, THIS CASE WAS IDENTIFIED TO HAVE THE POTENTIAL TO EXHIBIT THE HAZARD OF THE RECOMMENDED SCREW LENGTH BEING GREATER THAN THE BONE THICKNESS FOR IMPLANTS WITH SCREW HOLES LOCATED IN AN AREA OF THIN BONE. NO ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24455058 ↗
  • MalfunctionFeb 25, 2026Nonstandard Device

    WITHIN THE SCOPE EVALUATION FOR A PRODUCT FIELD ACTION, THIS CASE WAS IDENTIFIED TO HAVE THE POTENTIAL TO EXHIBIT THE HAZARD OF THE RECOMMENDED SCREW LENGTH BEING GREATER THAN THE BONE THICKNESS FOR IMPLANTS, WITH SCREW HOLES LOCATED IN AN AREA OF THIN BONE. NO ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24447087 ↗
  • MalfunctionFeb 25, 2026Nonstandard Device

    WITHIN THE SCOPE EVALUATION FOR A PRODUCT FIELD ACTION, THIS CASE WAS IDENTIFIED TO HAVE THE POTENTIAL TO EXHIBIT THE HAZARD OF THE RECOMMENDED SCREW LENGTH BEING GREATER THAN THE BONE THICKNESS FOR IMPLANTS WITH SCREW HOLES LOCATED IN AN AREA OF THIN BONE. NO ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24446927 ↗
  • MalfunctionFeb 25, 2026Nonstandard Device

    WITHIN THE SCOPE EVALUATION FOR A PRODUCT FIELD ACTION, THIS CASE WAS IDENTIFIED TO HAVE THE POTENTIAL TO EXHIBIT THE HAZARD OF THE RECOMMENDED SCREW LENGTH BEING GREATER THAN THE BONE THICKNESS FOR IMPLANTS WITH SCREW HOLES LOCATED IN AN AREA OF THIN BONE. NO ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24446925 ↗
  • MalfunctionFeb 25, 2026Nonstandard Device

    WITHIN THE SCOPE EVALUATION FOR A PRODUCT FIELD ACTION, THIS CASE WAS IDENTIFIED TO HAVE THE POTENTIAL TO EXHIBIT THE HAZARD OF THE RECOMMENDED SCREW LENGTH BEING GREATER THAN THE BONE THICKNESS FOR IMPLANTS WITH SCREW HOLES LOCATED IN AN AREA OF THIN BONE. NO ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24446924 ↗
  • MalfunctionFeb 25, 2026Nonstandard Device

    WITHIN THE SCOPE EVALUATION FOR A PRODUCT FIELD ACTION, THIS CASE WAS IDENTIFIED TO HAVE THE POTENTIAL TO EXHIBIT THE HAZARD OF THE RECOMMENDED SCREW LENGTH BEING GREATER THAN THE BONE THICKNESS FOR IMPLANTS WITH SCREW HOLES LOCATED IN AN AREA OF THIN BONE. NO ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24446915 ↗
  • MalfunctionFeb 25, 2026Nonstandard Device

    WITHIN THE SCOPE EVALUATION FOR A PRODUCT FIELD ACTION, THIS CASE WAS IDENTIFIED TO HAVE THE POTENTIAL TO EXHIBIT THE HAZARD OF THE RECOMMENDED SCREW LENGTH BEING GREATER THAN THE BONE THICKNESS FOR IMPLANTS WITH SCREW HOLES LOCATED IN AN AREA OF THIN BONE. NO ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24446910 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE TMJ DEVICES REQUIRE REVISION FOR AN UNKNOWN REASON.

    Patient problems: Unspecified Musculoskeletal problem

    MAUDE report 24377324 ↗
  • InjuryFeb 4, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE ORIGINAL IMPLANTS HAVE BEEN EXPLANTED AND A NEW REQUEST HAS BEEN CREATED.

    Patient problems: Bacterial Infection; Fistula; Insufficient Information

    MAUDE report 24258342 ↗
  • InjuryDec 17, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT THE SURGEON WAS UNABLE TO PLACE THE FOSSA COMPONENT AND DECIDED TO OMIT IT, PACKING THE JOINT WITH FAT INSTEAD. THE MANDIBULAR COMPONENT WAS SUCCESSFULLY IMPLANTED. A NEW FOSSA COMPONENT IS NEEDED.

    Patient problems: Insufficient Information; Loss of Range of Motion

    MAUDE report 23827575 ↗
  • InjuryDec 16, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    IT WAS REPORTED THAT THE PATIENT HAD TO STAY IN HOSPITAL FOR FEW DAYS LONGER THAN NORMAL AS THE PROSTHESIS WAS NOT FITTING 100% DUE TO LACK OF SPLINTS AT THE SURGERY PHASE. BASED ON A POST OP CT SCAN OF THE PATIENT, THE MANDIBULAR PART IS ANGULATED SLIGHTLY FORWARD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23816125 ↗
  • MalfunctionDec 16, 2025Patient-Device Incompatibility; Patient-Device Incompatibility

    THE CASE WAS NOT IMPLANTED BECAUSE THE SURGEON HAD CONCERNS REGARDING THE SCREW LOCATION. PREOPERATIVE IMPLANT PLANNING DOCUMENTS PROVIDED TO THE SURGEON INCLUDED A WARNING ABOUT THIS SCREW HOLE POSITION. AFTER REVIEWING THIS INFORMATION, THE SURGEON ELECTED NOT TO PROCEED WITH THE STRYKER IMPLANT AND USED AN ALTERNATIVE DEVICE. NO ADVERSE CONSEQUENCES WERE REPORTED FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23816124 ↗
  • InjuryDec 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NO NEW INFORMATION.

    Patient problems: Insufficient Information; Unspecified Infection

    MAUDE report 23741830 ↗
  • MalfunctionDec 4, 2025Nonstandard Device; Nonstandard Device

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23719824 ↗
  • MalfunctionDec 4, 2025Nonstandard Device; Nonstandard Device

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23719786 ↗
  • MalfunctionDec 4, 2025Nonstandard Device; Nonstandard Device

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23719709 ↗
  • MalfunctionDec 4, 2025Nonstandard Device; Nonstandard Device

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23716796 ↗

Other Joint, Temporomandibular, Implant devices