TMJ Bilateral Implant recallZ-0488-2026 · TMJ Solutions Inc · initiated Oct 13, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from TMJ Solutions Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Patient-fitted temporomandibular joint implant contained discrepancy in screw hole position.
FDA root cause category: Process control
What the company is doing
On February 25, 2026 Stryker (TMJ Solutions) issued a " Urgent: Medical Device Recall" Expansion notification to additional, affected consignees. Initial notification and previous expansion was issued on September 30, 2025 and December 4, 2025. TMJ Solutions asked consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product (Attachment B). Segregate the affected products, quarantine and discontinue use. 3. Please complete the Business Reply Form or use the QR Code on Attachment A. Response is required. a. If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility .
Recall details
- Recall number
- Z-0488-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 13, 2025
- Posted by the FDA
- Nov 14, 2025
- Recalling firm
- TMJ Solutions Inc
- Quantity
- 63 units
- Product code
- LZD
- FDA submissions
- P980052
- Distribution
- US: IL, OR, MS, TX, OK, UT, MA, MD, NY, PA, VA, FL, CA, MN, GA OUS: Canada, Spain, Brazil,
Product description
TMJ Bilateral Implant REF: CHG020 Sterile EO, Rx Only
Code information (models, lots, serial numbers)
UDI: 07613327626551,07613327626568,07613327626575,07613327626605,07613327627749 07613327627848/ Lot #s: 2410301062 2306091054 2312121078 2401181023 2401231058 2403191046 2404291064 2405021079 2407221023 2407241032 2408081047 2408131037 2408161043 2409031033 2409241016 2410011097 2410071065 2410091014 2410111030 2410291062 2411111012 2411121044 2411201015 2412061031 2412121062 2412201009 2501021011
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Patient-Fitted TMJ Reconstructive Prosthesis System CHG020TMJ Solutions, Inc. · Joint, Temporomandibular, Implant2 recalls172 reports
- Patient-Fitted TMJ Reconstructive Prosthesis System CHG010TMJ Solutions, Inc. · Joint, Temporomandibular, Implant1 recall94 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.