Tina-quant ß2-Microglobulin 08047430190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336119990), received between Oct 16, 2024 and Oct 16, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2024Low Test Results; Low Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE TINA-QUANT 2-MICROGLOBULIN RESULTS FOR 2 PATIENT SAMPLES ON A COBAS 8000 C 502 MODULE. PATIENT 1: THE INITIAL RESULT WAS INVALIDATED DUE TO HAVING A DATA FLAG. THE SAMPLE WAS AUTOMATICALLY DILUTED AND THE RESULT WAS 6.4 MG/L. THE SAMPLE WAS MANUALLY DILUTED AND THE RESULT WAS 16.8 MG/L. THIS RESULT WAS BELIEVED TO BE CORRECT. PATIENT 2: THE INITIAL RESULT WAS INVALIDATED DUE TO HAVING A DATA FLAG. THE SAMPLE WAS AUTOMATICALLY DILUTED AND THE RESULT WAS 7.2 MG/L. THE SAMPLE WAS MANUALLY DILUTED AND THE RESULT WAS 16.8 MG/L. THIS RESULT WAS BELIEVED TO BE CORRECT. THE SAMPLES WERE REPEATED DUE TO THE INITIAL RESULT HAVING A DATA FLAG, AND HIGHER RESULTS WERE EXPECTED. THE QUESTIONABLE RESULTS WERE NOT REPORTED OUTSIDE THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20461807 ↗
Other System, test, beta-2-microglobulin immunological devices
- Atellica CH B2M 11097635Siemens Healthcare Diagnostics Inc. · System, test, beta-2-microglobulin immunological2 recalls0 reports
- Optilite Beta 2 Microglobulin Urine Kit LK043.L.OPT.AThe Binding Site Group Limited · System, Test, Beta-2-Microglobulin Immunological1 recall0 reports
- Tinaquant 2Microglobulin 11660551216Roche Diagnostics GmbH · SYSTEM, TEST, BETA-2-MICROGLOBULIN IMMUNOLOGICAL1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.