Atellica CH B2M 11097635Recalls & reported problems
Siemens Healthcare Diagnostics Inc. · ß2-Microglobulin (B2M) Atellica CH - RGT 400T · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Oct 21, 2024Class IIOpen, Classified
Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH 2-Microglobulin (B2M) reagent.
Z-0599-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 1, 2021Class IITerminated
2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L
Z-0440-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Other System, test, beta-2-microglobulin immunological devices
- Optilite Beta 2 Microglobulin Urine Kit LK043.L.OPT.AThe Binding Site Group Limited · System, Test, Beta-2-Microglobulin Immunological1 recall0 reports
- Tina-quant ß2-Microglobulin 08047430190Roche Diagnostics GmbH · System, test, beta-2-microglobulin immunological1 report
- Tinaquant 2Microglobulin 11660551216Roche Diagnostics GmbH · SYSTEM, TEST, BETA-2-MICROGLOBULIN IMMUNOLOGICAL1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page