Atellica CH B2M 11097635Recalls & reported problems

Siemens Healthcare Diagnostics Inc. · ß2-Microglobulin (B2M) Atellica CH - RGT 400T · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Oct 21, 2024Class IIOpen, Classified

    Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH 2-Microglobulin (B2M) reagent.

    Z-0599-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 1, 2021Class IITerminated

    2 Microglobulin (B2M) Reagent Lot 211446 Bias with Quality Control and Patient Sample Results. Positive bias on patient samples of up to +20.9% at a B2M concentration of approximately 1.5 mg/L and a negative bias on patient samples of up to -10.5% at a B2M concentration of approximately 15 mg/L

    Z-0440-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other System, test, beta-2-microglobulin immunological devices

All System, test, beta-2-microglobulin immunological devices →