ONLINE DAT Benzodiazepines II 08056927190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336121054), received between May 23, 2025 and Jul 14, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- False Positive Result2
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 14, 2025False Positive Result; False Positive Result
THE INITIAL REPORTER STATED THEY RECEIVED FALSE POSITIVE RESULTS FOR THREE PATIENT SAMPLES TESTED WITH ONLINE DAT BENZODIAZEPINES II ON TWO COBAS C 503 ANALYTICAL UNIT ANALYZERS. THE FIRST PATIENT'S SAMPLE RESULTED IN A BENZODIAZEPINES VALUE OF 76.5 MABS (POSITIVE) WHEN TESTED ON THE FIRST C 503 ANALYZER AND IT RESULTED IN A VALUE OF 59.5 MABS (POSITIVE) WHEN TESTED ON THE SECOND C 503 ANALYZER. THE SAMPLE WAS SENT TO ANOTHER LABORATORY FOR GC/MS CONFIRMATION TESTING AND THE RESULT WAS NEGATIVE (SEE ATTACHMENT FOR FULL RESULTS). THE SECOND PATIENT'S SAMPLE RESULTED IN A BENZODIAZEPINES VALUE OF 12.5 MABS (POSITIVE) WHEN TESTED ON THE FIRST C 503 ANALYZER ON (B)(6) 2025 AND IT RESULTED IN A POSITIVE WHEN TESTED ON THE SECOND C 503 ANALYZER ON (B)(6) 2025. THE SAMPLE WAS SENT TO ANOTHER LABORATORY FOR GC/MS CONFIRMATION TESTING AND THE RESULT WAS NEGATIVE (SEE ATTACHMENT FOR FULL RESULTS). THE THIRD PATIENT'S SAMPLE RESULTED IN A BENZODIAZEPINES VALUE OF 10.8 MABS (POSITIVE) WHEN TESTED ON THE FIRST C 503 ANALYZER ON (B)(6) 2025 AND IT RESULTED IN A VALUE OF 20.9 MABS (POSITIVE) WHEN TESTED ON THE SECOND C 503 ANALYZER ON (B)(6) 2025. THE SAMPLE WAS SENT TO ANOTHER LABORATORY FOR GC/
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22493339 ↗MalfunctionMay 23, 2025False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings; False Positive Result; Incorrect, Inadequate or Imprecise Result or Readings
THERE WAS AN ALLEGATION OF A QUESTIONABLE FALSE POSITIVE ONLINE DAT BENZODIAZEPINES II RESULT FROM THE COBAS PRO SSU. THE PATIENT HAD AN INITIAL POSITIVE RESULT. THE REPEAT RESULT BY LC/MS WAS NEGATIVE FOR BENZODIAZEPINES INCLUDING SO-7-AMINOCLONAZEPAM, SO-7-AMINOFLUNITRAZEPAM, SO-ALPHA HYDROXYALPRAZOLAM, SO-LORAZEPAM, SO-NORDIAZEPAM, SO-OXAZEPAM, AND SO-TEMAZEPAM. THIS RESULT THAT WAS DEEMED TO BE CORRECT. THE RESULTS WERE NOT RELEASED OUTSIDE OF THE LABORATORY. THE SAMPLE WAS REPEATED AS THE CUSTOMER HAS HAD OTHER INITIALLY POSITIVE DRUG SCREENINGS THAT WERE REPEATED AND FOUND TO BE NEGATIVE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22080634 ↗
Other Enzyme immunoassay, benzodiazepine devices
- Benzodiazepine Intercept 5 Plate Kit 1110IBOrasure Technologies, Inc. · Enzyme Immunoassay, Benzodiazepine1 recall0 reports
- ONLINE DAT Benzodiazepines Plus 08056919190Roche Diagnostics GmbH · Enzyme immunoassay, benzodiazepine5 reports
- Benzodiazepines Plus 4490789190Roche Diagnostics GmbH · Enzyme immunoassay, benzodiazepine5 reports
- ONLINE DAT Benzodiazepines II 04939417190Roche Diagnostics GmbH · Enzyme immunoassay, benzodiazepine3 reports
- VITROS 6801989Ortho-Clinical Diagnostics, Inc. · Enzyme immunoassay, benzodiazepine3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page