Ortho-Clinical Diagnostics, Inc. · VITROS Chemistry Products BENZ Reagent · Class 2

All models of VITROS (71)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10758750001750), received between Nov 8, 2024 and May 13, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2025False Negative Result; False Negative Result

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A DISCORDANT NEGATIVE VITROS BENZODIAZEPINE (BENZ) RESULT WAS OBTAINED FROM A 10X DILUTION OF A PATIENT SAMPLE USING VITROS CHEMISTRY PRODUCTS BENZ REAGENT PROCESSED ON A VITROS 5600 INTEGRATED SYSTEM. THE VITROS BENZ RESULT WAS CONSIDERED DISCORDANT WHEN COMPARED TO A POSITIVE RESULT OBTAINED USING THE SAME PATIENT SAMPLE TESTED UNDILUTED ON THE SAME VITROS 5600 INTEGRATED SYSTEM. PATIENT SAMPLE (TEST EVENT 2) RESULT OF < 850 NG/ML (NEGATIVE) VS AN EXPECTED RESULT OF POSITIVE. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. IT WAS UNKNOWN IF THE DISCORDANT, NEGATIVE VITROS BENZ RESULT WAS REPORTED FROM THE LABORATORY. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22014514 ↗
  • MalfunctionNov 8, 2024False Negative Result; False Negative Result

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT FALSE NEGATIVE VITROS BENZODIAZEPINE (BENZ) RESULTS WERE OBTAINED FROM SAMPLES FROM TWO DIFFERENT PATIENTS PROCESSED USING VITROS CHEMISTRY PRODUCTS BENZ REAGENT LOT: 1523-67-2046 IN COMBINATION WITH A VITROS XT 7600 INTEGRATED SYSTEM. THE RESULTS WERE CONSIDERED TO BE FALSE NEGATIVES WHEN COMPARED TO RESULTS USING THE SAME SAMPLES FROM CONFIRMATORY TESTING USING GC/MS AND AN UNKNOWN REFERENCE METHOD. PATIENT 1 SAMPLE RESULT OF 143 NG/ML (NEGATIVE) VS AN EXPECTED RESULT OF POSITIVE PATIENT 2 SAMPLE RESULT OF 129 NG/ML (NEGATIVE) VS AN EXPECTED RESULT OF POSITIVE BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE FALSE NEGATIVE VITROS BENZ RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY. THERE WAS NO REPORTED ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF TWO MDRS FOR THIS EVENT. TWO 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20639940 ↗
  • MalfunctionNov 8, 2024False Negative Result; False Negative Result

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT FALSE NEGATIVE VITROS BENZODIAZEPINE (BENZ) RESULTS WERE OBTAINED FROM SAMPLES FROM TWO DIFFERENT PATIENTS PROCESSED USING VITROS CHEMISTRY PRODUCTS BENZ REAGENT LOT 1523-67-2046 IN COMBINATION WITH A VITROS XT 7600 INTEGRATED SYSTEM. THE RESULTS WERE CONSIDERED TO BE FALSE NEGATIVES WHEN COMPARED TO RESULTS USING THE SAME SAMPLES FROM CONFIRMATORY TESTING USING GC/MS AND AN UNKNOWN REFERENCE METHOD. PATIENT 1 SAMPLE RESULT OF 143 NG/ML (NEGATIVE) VS AN EXPECTED RESULT OF POSITIVE PATIENT 2 SAMPLE RESULT OF 129 NG/ML (NEGATIVE) VS AN EXPECTED RESULT OF POSITIVE BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE FALSE NEGATIVE VITROS BENZ RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY. THERE WAS NO REPORTED ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ONE OF TWO MDRS FOR THIS EVENT. TWO 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABI

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20639756 ↗