Bicarbonate Liquid 08057494190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613336121146), received between Jun 1, 2023 and Sep 28, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Incorrect Measurement1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026High Test Results
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50914632 ↗MalfunctionSep 10, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50717073 ↗MalfunctionJun 1, 2023Incorrect Measurement; Incorrect Measurement
THE INITIAL REPORTER QUESTIONED THE RESULTS OF 1 PATIENT SAMPLE TESTED FOR BICARBONATE LIQUID (CO2-L) ON A C503 INSTRUMENT. THE INITIAL RESULT WAS 2.59 MMOL/L. AN ADDITIONAL RESULT OF < 2 MMOL/L WAS ALSO PROVIDED. A ½ DILUTION WAS PERFORMED AND THE RESULT WAS 4.73 MMOL/L. AN ADDITIONAL RESULT OF 4 MMOL/L WAS ALSO PROVIDED. THE SAMPLE WAS TESTED ON A DIFFERENT C503 INSTRUMENT AND THE SAME RESULTS WERE OBTAINED. THE SPECIFIC RESULTS WERE NOT PROVIDED. A SECOND SAMPLE WAS OBTAINED FROM THE PATIENT AND THE SAME RESULTS WERE OBTAINED. THE SPECIFIC RESULTS WERE NOT PROVIDED. THE SAMPLE WAS TESTED ON A RADIOMETER ABL800 FLEX BLOOD GAS INSTRUMENT WITH ELEVATED RESULTS. THE PCO2 RESULTS FROM THE BLOOD GAS INSTRUMENT FROM THE PLASMA SAMPLE WERE 20.0 MMHG AND 23.9 MMHG. THE PCO2 RESULTS FROM THE BLOOD GAS INSTRUMENT USING A WHOLE BLOOD SAMPLE WERE 20.4 MMHG AND 19.3 MMHG. ADDITIONAL SAMPLES WERE OBTAINED FROM THE PATIENT AND TESTED ON DIFFERENT C503 INSTRUMENTS. SEE ATTACHED DATA FOR THE RESULTS. A LIPOGRAM WAS PERFORMED AND THE RESULTS SUGGEST A HIGHER LEVEL OF CHYLOMYCRON IN THE SAMPLE. THE CUSTOMER SUSPECTS AN INTERFERENCE AFFECTING THE RESULTS. THE C503 INSTRUMENT SERIAL NUMBER WAS NOT PR
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17040412 ↗
Other ENZYMATIC, CARBON-DIOXIDE devices
- AU OSR6237Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls1 report
- AU OSR6637Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls0 reports
- AU OSR6137Beckman Coulter, Inc. · Enzymatic, Carbon-Dioxide2 recalls0 reports
- Alinity 07P7220Abbott GmbH · ENZYMATIC, CARBON-DIOXIDE90 reports
- VITROS 8262396Ortho-Clinical Diagnostics, Inc. · ENZYMATIC, CARBON-DIOXIDE1 recall0 reports
- ARCHITECT 3L80-22Abbott GmbH · ENZYMATIC, CARBON-DIOXIDE31 reports
- Atellica CH CO2_c 11097521Siemens Healthcare Diagnostics Inc. · ENZYMATIC, CARBON-DIOXIDE21 reports
- Dimension Vista® Flex® reagent cartridge CO2 10445163Siemens Healthcare Diagnostics Inc. · ENZYMATIC, CARBON-DIOXIDE15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page