Skip to content

Tina-quant Ferritin Gen.4 08057648190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336121238), received between Jul 12, 2023 and Mar 8, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Low Test Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 8, 2024High Test Results; High Test Results

    THE INITIAL REPORTER COMPLAINED OF DISCREPANT HIGH RESULTS FOR 1 PATIENT SAMPLE TESTED FOR FERR4 ON A COBAS C 303 ANALYTICAL UNIT COMPARED TO TWO COMPETITOR METHODS. ON 14-FEB-2024 THE INITIAL RESULT FROM THE C303 ANALYZER WAS 46.7 NG/ML. ON 15-FEB-2024 THE REPEAT RESULT FROM COMPETITOR METHOD 1 WAS 2.95 NG/ML. THE REPEAT RESULT FROM COMPETITOR METHOD 2 WAS < 3 NG/ML. A NEW SAMPLE WAS OBTAINED AND THE RESULT FROM THE C 303 ANALYZER WAS 49 NG/ML.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18866633 ↗
  • MalfunctionJul 12, 2023Low Test Results; Low Test Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH FERR4 (FERRITIN) ON A COBAS PRO C 503 ANALYTICAL UNIT. THE CUSTOMER NOTED THAT CONTROLS WERE RUNNING HIGH WITH A NEW LOT OF FERRITIN REAGENT. THE SAMPLE INITIALLY RESULTED IN A FERRITIN VALUE OF 7 NG/ML. CONTROLS TESTED AFTER THE INITIAL SAMPLE RUN WERE OUTSIDE OF RANGE, SO THE SAMPLE WAS REPEATED. THE SAMPLE REPEATED WITH A FERRITIN VALUE OF 11 NG/ML. THIS REPEAT VALUE WAS DEEMED CORRECT AND A CORRECTED REPORT WAS ISSUED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17309096 ↗

Other FERRITIN, ANTIGEN, ANTISERUM, CONTROL devices

All FERRITIN, ANTIGEN, ANTISERUM, CONTROL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page