Alinity 07P6530Recalls & reported problems

Abbott Ireland · Alinity i Ferritin Reagent Kit 1200 Tests · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00380740131487), received between Nov 21, 2018 and Apr 25, 2019 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Device Damaged Prior to Use” (7 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Nov 21, 2018Terminated

    Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.

    Z-2208-2019 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Device Damaged Prior to Use7
  • Unable to Obtain Readings7

Patient problems

  • No Patient Involvement6
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2019Unable to Obtain Readings; Device Damaged Prior to Use; Unable to Obtain Readings; Device Damaged Prior to Use

    THE CUSTOMER OBSERVED ERROR CODES 5652, 5662 AND 5819 RELATED TO REAGENT CARTRIDGE LOADING WHILE ATTEMPTING TO LOAD ALINITY I FERRITIN REAGENT LOT 91532UI00 ON THE ALINITY I PROCESSING MODULE. THERE WAS NO REPORT OF ANY PATIENT IMPACT DUE TO THE POTENTIAL DELAY IN GENERATING FERRITIN RESULTS.

    Patient problems: No Patient Involvement

    MAUDE report 8550231 ↗
  • MalfunctionDec 20, 2018Unable to Obtain Readings; Device Damaged Prior to Use; Unable to Obtain Readings; Device Damaged Prior to Use

    THE CUSTOMER OBSERVED ERROR CODE 5662 (REAGENT TRANSPORT FAILED TO LOAD REAGENT CARTRIDGE OR RACK IN REAGENT CAROUSEL) WHILE LOADING ALINITY I FERRITIN REAGENT LOT 91532UI00 ON THE ALINITY I PROCESSING MODULE. THERE WAS NO REPORT OF ANY PATIENT IMPACT DUE TO THE POTENTIAL DELAY IN GENERATING FERRITIN RESULTS.

    Patient problems: No Patient Involvement

    MAUDE report 8187145 ↗
  • MalfunctionDec 20, 2018Unable to Obtain Readings; Device Damaged Prior to Use; Unable to Obtain Readings; Device Damaged Prior to Use

    THE CUSTOMER OBSERVED ERROR CODE 5817 (UNABLE TO INITIALIZE MODULE DUE TO PREVIOUS LOAD OR UNLOAD ERROR. REMOVE REAGENT CARTRIDGE OR RACK WITH LOAD OR UNLOAD ERROR FROM REAGENT CAROUSEL) WHILE LOADING ALINITY I FERRITIN REAGENT LOT 93238UI00 ON THE ALINITY I PROCESSING MODULE. THERE WAS NO REPORT OF ANY PATIENT IMPACT DUE TO THE POTENTIAL DELAY IN GENERATING FERRITIN RESULTS.

    Patient problems: No Patient Involvement

    MAUDE report 8186942 ↗
  • MalfunctionDec 20, 2018Unable to Obtain Readings; Device Damaged Prior to Use; Unable to Obtain Readings; Device Damaged Prior to Use

    THE CUSTOMER WAS UNABLE TO TURN THE MICROPARTICLE PORTION OF THE ALINITY I FERRITIN REAGENT CARTRIDGE LOT 93238UI00 ON THE ALINITY I PROCESSING MODULE. THERE WAS NO REPORT OF ANY PATIENT IMPACT DUE TO THE POTENTIAL DELAY IN GENERATING FERRITIN RESULTS.

    Patient problems: No Patient Involvement

    MAUDE report 8186381 ↗
  • MalfunctionDec 20, 2018Unable to Obtain Readings; Device Damaged Prior to Use; Unable to Obtain Readings; Device Damaged Prior to Use

    THE CUSTOMER OBSERVED LOAD FAILURE ERROR CODES 5819, 5818, 5672 AND 5794 WHEN ATTEMPTING TO LOAD ALINITY I FERRITIN REAGENT CARTRIDGE LOT 91532UI00 ON THE ALINITY I PROCESSING MODULE. THERE WAS NO REPORT OF ANY PATIENT IMPACT DUE TO THE POTENTIAL DELAY IN GENERATING FERRITIN RESULTS.

    Patient problems: No Patient Involvement

    MAUDE report 8186096 ↗
  • MalfunctionDec 14, 2018Unable to Obtain Readings; Device Damaged Prior to Use; Unable to Obtain Readings; Device Damaged Prior to Use

    THE CUSTOMER OBSERVED ERROR CODES 5818 (REAGENT TRANSPORT ERROR), 5819 (REAGENT CARTRIDGE LOAD ERROR AT REAGENT CAROUSEL POSITION 27) AND 5794 (REAGENT CAROUSEL LOAD ERROR) WHILE LOADING ALINITY I FERRITIN REAGENT LOT: 91532UI00 ON THE ALINITY I PROCESSING MODULE. THERE WAS NO REPORT OF ANY PATIENT IMPACT DUE TO THE POTENTIAL DELAY IN GENERATING FERRITIN RESULTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8165574 ↗
  • MalfunctionNov 21, 2018Unable to Obtain Readings; Device Damaged Prior to Use; Unable to Obtain Readings; Device Damaged Prior to Use

    ABBOTT LABORATORIES HAS BECOME AWARE OF A MANUFACTURING ISSUE THAT CAN RESULT IN DAMAGE TO A BOTTLE NECK IN SOME REAGENT CARTRIDGES. IF A BOTTLE NECK IS DAMAGED, IT MAY CAUSE A FAILURE DURING THE REAGENT LOADING PROCESS ON THE ALINITY I INSTRUMENT. WHEN THE FAILURE OCCURS, ERROR CODES WILL BE GENERATED WHICH MAY LEAD TO THE POTENTIAL OF DELAYED PATIENT RESULTS. NO PATIENT INJURY HAS BEEN REPORTED DUE TO THIS ISSUE.

    Patient problems: No Patient Involvement

    MAUDE report 8097185 ↗

Other FERRITIN, ANTIGEN, ANTISERUM, CONTROL devices