Tina-quant IgM Gen.2 08057923190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613336121337), received between Aug 21, 2024 and Jun 27, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K040431); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Low Test Results1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 27, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S SAMPLE TUBES TESTED WITH TINA-QUANT IGM GEN.2 (IGM-2) ASSAY ON A COBAS C503 ANALYTICAL UNIT. TWO SAMPLE TUBES (SAMPLE TUBE 1 AND SAMPLE TUBE 2) WERE DRAWN FROM THE PATIENT AND TESTED. SAMPLE TUBE 1: INITIAL RESULT: 34.3 G/L (ACCOMPANIED BY A DATA FLAG). REPEAT RESULT: 30.80 G/L (AUTOMATICALLY RERUN WITH DILUTION). SAMPLE TUBE 2 WAS TESTED IN A DIFFERENT LABORATORY: RESULT OBTAINED FROM A DIFFERENT COBAS C503: 18 G/L. RESULT OBTAINED FROM ELECTROPHORESIS: 18.0 G/L. THE PHYSICIAN QUESTIONED THE RESULTS AS THEY WERE INCONSISTENT, PROMPTING THE REPEAT OF THE SAMPLE TUBES ON (B)(6) 2025: SAMPLE TUBE 2 WAS TESTED ON A COBAS C303 AT A DIFFERENT LABORATORY, AND THE RESULT WAS 14.6 G/L. SAMPLE TUBE 1 WAS REPEATED ON THE ORIGINAL C503, RESULTING IN AN IGM-2 VALUE OF 18.0 G/L AND ACCOMPANIED BY A DATA FLAG. SAMPLE TUBE 1 WAS RERUN AFTER DILUTION, RESULTING IN AN IGM-2 VALUE OF 15.4 G/L. THE RESULT OF 18 G/L OBTAINED FROM A DIFFERENT COBAS C503 AT A DIFFERENT LABORATORY WAS DEEMED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22344595 ↗MalfunctionMar 3, 2025High Test Results; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE TINA-QUANT IGM GEN.2 RESULTS FOR QC MATERIAL AND PATIENT SAMPLES FROM THE COBAS C 503 ANALYTICAL UNIT. ONE EXAMPLE WAS PROVIDED OF AN INITIAL RESULT OF 29 MG/DL AND A REPEAT RESULT OF 14.4 MG/DL. THE REPEAT RESULT WAS OBTAINED AFTER RECALIBRATING THE ASSAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21503249 ↗MalfunctionAug 21, 2024Low Test Results; Low Test Results
THE INITIAL REPORTER COMPLAINED OF A DISCREPANT LOW RESULT FOR 1 PATIENT SAMPLE TESTED FOR ROCHE DIAGNOSTICS TINA-QUANT IGM GEN.2 (IGM-2) ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 5.47 G/L. THE REPEAT RESULT WAS >63.0 G/L WITH AN INVALIDATING DATA FLAG. THE SAMPLE WAS REPEATED WITH A MANUAL DILUTION OF 1:21 AND THE RESULT WAS 59.85 G/L. THE QUESTIONABLE RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20020694 ↗
FDA clearance and classification
Cleared through Special 510(k) K040431 on Mar 10, 2004 (decision: Substantially Equivalent). Submitted by Roche Diagnostics Corp. as “Roche Diagnostics Tina-Quant Igm Gen.2”. The FDA review took 20 days. FDA record ↗
The FDA regulates product code DAO (igm (mu chain specific), antigen, antiserum, control) under 21 CFR 866.5550 as a Class II device. 21 CFR 866.5550 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K040431. One submission often covers many models, so check the product and lot details in each record.
- Z-1147-2019 · Dec 17, 2018 · TerminatedTina-quant IgM Gen.2, Catalog Nos. 03507190190, 05220726190, 03507041190, 03507149190 Product Usage: Immunoturbidimetric assay for the quantitative in vitro determination of IgM i
Common questions about the Tina-quant IgM Gen.2 08057923190
Has the Tina-quant IgM Gen.2 08057923190 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613336121337), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the Tina-quant IgM Gen.2 08057923190?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Aug 21, 2024 and Jun 27, 2025. In those reports, the most common specific device problems are “High Test Results” (2), “Low Test Results” (1) and “Non Reproducible Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Tina-quant IgM Gen.2 08057923190?
None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Tina-quant IgM Gen.2 08057923190 FDA approved?
It was cleared through Special 510(k) K040431 on Mar 10, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Tina-quant IgM Gen.2 08057923190, and is it still sold?
The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand Tina-quant IgM Gen.2, model 08057923190). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IGM (MU CHAIN SPECIFIC), ANTIGEN, ANTISERUM, CONTROL devices
- Tinaquant IgM Gen2 5220726190Roche Diagnostics GmbH · IGM (MU CHAIN SPECIFIC), ANTIGEN, ANTISERUM, CONTROL3 reports
- Tinaquant IgM Gen2 3507190190Roche Diagnostics GmbH · IGM (MU CHAIN SPECIFIC), ANTIGEN, ANTISERUM, CONTROL1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page