Total Protein Gen.2 08058652190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 07613336121528), received between Nov 22, 2022 and Mar 17, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Non Reproducible Results” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Non Reproducible Results9
- Low Test Results6
- High Test Results3
- Output Problem2
- Chemical Problem1
- Incorrect, Inadequate or Imprecise Result or Readings1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2026Low Test Results; Non Reproducible Results
THE INITIAL REPORTER COMPLAINED OF A DISCREPANT LOW RESULT FOR 1 PATIENT SAMPLE TESTED FOR TOTAL PROTEIN GEN.2 (TP2) ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 3.41 G/L. THE REPEAT RESULT WAS 68.1 G/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24619910 ↗MalfunctionFeb 11, 2026High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE TINA-QUANT C-REACTIVE PROTEIN IV (CRP), TOTAL PROTEIN GEN.2 (TP), AND ALKALINE PHOSPHATASE ACC. TO IFCC GEN.2 (ALP) RESULTS FOR 2 PATIENT SAMPLES ON A COBAS C 503 ANALYTICAL UNIT. THIS MEDWATCH WILL COVER TP. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE CRP RESULTS AND MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE ALP RESULTS. THE CUSTOMER WAS PROMPTED TO REPEAT THE PATIENT SAMPLES AS THEY LOOKED LIKE THEY HAD NOT BEEN CENTRIFUGED. SAMPLE 1 HAD AN INITIAL CRP RESULT OF 0.362 MG/DL AND AN INITIAL TP RESULT OF 10.6 G/DL. THE REPEAT CRP RESULT WAS 0.62 MG/DL AND THE REPEAT TP RESULT WAS 7.0 G/DL. SAMPLE 2 HAD AN INITIAL ALP RESULT OF 159 U/L. THE REPEAT RESULT WAS 88 U/L. THE REPEAT RESULTS WERE DEEMED CORRECT. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24315633 ↗MalfunctionMay 14, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE TOTAL PROTEIN GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 703 ANALYTICAL UNIT. THE INITIAL RESULT WAS 0.119 G/DL. THE REPEATED RESULT WAS 7.81 G/DL. THE REPEATED RESULT MATCHED THE PATIENT'S CLINICAL HISTORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22017820 ↗MalfunctionMar 24, 2025High Test Results; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE TOTAL PROTEIN GEN.2 RESULTS FROM THE COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 105 G/L. THE REPEAT RESULT WAS 73.5 G/L. THE REPEAT RESULT USING ANOTHER COBAS C503 ANALYZER WAS 73 G/L. THE QUESTIONABLE RESULT WAS NOT REPORTED TO THE PHYSICIAN OR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21676145 ↗MalfunctionFeb 19, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
THE INITIAL REPORTER ALLEGED A STOP ERROR OCCURRED ON A COBAS C 303 ANALYTICAL UNIT. REPORTEDLY, A SPECIMEN IN PROCESS STOPPED PROCESSING AND DID NOT FINISH TESTING BUT RESULTS WERE PRODUCED AND RELEASED. THE CUSTOMER NOTED FLAGS ON MANY OF THE RESULTS AND REPEATED THE SAMPLE. AFTER REPEAT TESTING, DISCREPANT RESULTS WERE IDENTIFIED FOR TOTAL PROTEIN GEN.2 (TP2). THE INITIAL RESULT WAS 6.36 G/DL. THE REPEAT RESULT WAS 7.97 G/DL. THE REPEAT RESULT WAS BELIEVED TO BE CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21410584 ↗MalfunctionJan 30, 2024Non Reproducible Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE TP2 TOTAL PROTEIN GEN.2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL TP2 RESULT WAS 6 G/L. THE CUSTOMER QUESTIONED THE RESULT AS IT DID NOT MATCH THE PATIENT'S PREVIOUS RESULT AND WAS PROMPTED TO REPEAT THE SAMPLE. THE REPEAT RESULT WAS 54 G/L. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18606696 ↗MalfunctionDec 11, 2023Output Problem; Output Problem
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S SAMPLE TESTED WITH TOTAL PROTEIN GEN.2 (TP2) ASSAY AND ALBUMIN BCP (ALBP) ASSAY ON A COBAS PRO C503 ANALYZER. THIS MEDWATCH WILL APPLY TO THE TP2 ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE ALBP ASSAY. TP2: INITIAL RESULT: 5.60 G/L. REPEAT RESULT: 75.3 G/L. ALBP: INITIAL RESULT: 4.83 G/L. REPEAT RESULT: 42.4 G/L. THE MEDICAL PERSONNEL QUESTIONED THE RESULTS AND THE SAMPLE WAS THEN REPEATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18305589 ↗MalfunctionDec 11, 2023Output Problem; Output Problem
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT'S SAMPLE TESTED WITH TOTAL PROTEIN GEN.2 (TP2) ASSAY AND ALBUMIN BCP (ALBP) ASSAY ON A COBAS PRO C503 ANALYZER. THIS MEDWATCH WILL APPLY TO THE ALBP ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6). FOR INFORMATION RELATED TO THE TP2 ASSAY. ALBP: INITIAL RESULT: 4.83 G/L. REPEAT RESULT: 42.4 G/L. TP2: INITIAL RESULT: 5.60 G/L. REPEAT RESULT: 75.3 G/L. THE MEDICAL PERSONNEL QUESTIONED THE RESULTS AND THE SAMPLE WAS THEN REPEATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18305588 ↗MalfunctionAug 2, 2023Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; No Apparent Adverse Event; Incorrect, Inadequate or Imprecise Result or Readings; Low Test Results; No Apparent Adverse Event
THE INITIAL REPORTER RECEIVED A QUESTIONABLE TP2 TOTAL PROTEIN GEN.2 RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE INITIAL RESULT DID NOT MATCH TYPICAL RESULTS FOR THIS PATIENT AS THEY USUALLY RUN BETWEEN 6.7 AND 7.2 G/DL. THE INITIAL RESULT WAS 3.4 G/DL. THE REPORTER RERAN THE PATIENT SAMPLE 5 TIMES. THE REPEAT RESULT OF ALL 5 RERUNS WAS 6.6 G/DL. THE REPEAT RESULTS WERE DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17444082 ↗MalfunctionApr 14, 2023Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
WE RECEIVED AN ALLEGATION OF QUESTIONABLE RESULTS FOR 7 PATIENT'S SAMPLE TESTED WITH TP2 (TOTAL PROTEIN) ASSAY ON A COBAS PRO C503 ANALYTICAL UNIT SERIAL NUMBER (B)(4) COMPARED TO ANOTHER COBAS PRO C503 ANALYTICAL UNIT SERIAL NUMBER (B)(4). ON (B)(6) 2023 THE CUSTOMER INITIALLY RAN THE SAMPLES ON COBAS PRO C503 ANALYTICAL UNIT SERIAL NUMBER 2176-06. THE QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE PHYSICIAN QUESTIONED THE RESULTS. THE NEXT DAY, THE SAMPLES WERE REPEATED ON A DIFFERENT COBAS PRO C503 ANALYTICAL UNIT SERIAL NUMBER 2169-06 AND THEY GOT CORRECTED. SAMPLE 1: INITIAL RESULT: 3.5 G/DL. REPEAT RESULT: 6.84 G/DL. SAMPLE 2: INITIAL RESULT: 4.8 G/DL. REPEAT RESULT: 7.21 G/DL. SAMPLE 3: INITIAL RESULT: 3.6 G/DL. REPEAT RESULT: 7.31 G/DL. SAMPLE 4: INITIAL RESULT: 0.6 G/DL. REPEAT RESULT: 7.65 G/DL. SAMPLE 5: INITIAL RESULT: 3.6 G/DL. REPEAT RESULT: 7.3 G/DL. SAMPLE 6: INITIAL RESULT: 3.9 G/DL. REPEAT RESULT: 6.71 G/DL. SAMPLE 7: INITIAL RESULT: 0.2 G/DL. REPEAT RESULT: 7.56 G/DL. THE TP2 REAGENT LOT NUMBER WAS 693361. THE EXPIRATION DATE WAS NOT PROVIDED. QC AND CALIBRATION WERE WITHIN THE ACEPTABLE RANGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16742920 ↗MalfunctionFeb 16, 2023Chemical Problem; Non Reproducible Results; Chemical Problem; Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE TP2 (TOTAL PROTEIN GEN. 2) ASSAY ON A COBAS C 503 ANALYTICAL UNIT. THE SAMPLE INITIALLY RESULTED IN A TP2 VALUE OF 0.3 G/DL, ACCOMPANIED BY A DATA FLAG. THE SAMPLE WAS REPORTED OUTSIDE OF THE LABORATORY AND QUESTIONED. THE SAMPLE WAS REPEATED RESULTING IN A VALUE OF 7.0 G/DL, WHICH WAS DEEMED CORRECT. THE TP2 REAGENT LOT WAS 63565201 WITH AN EXPIRATION DATE OF 31-JUL-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16388841 ↗MalfunctionJan 18, 2023Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED A QUESTIONABLE TP2 TOTAL PROTEIN GEN.2 RESULT FROM ONE PATIENT SAMPLE TESTED ON THE COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE TECHNICIAN QUESTIONED THE RESULT AFTER REPORTING IT, WHICH PROMPTED THE RERUN OF THE PATIENT SAMPLE. THE INITIAL RESULT WAS 3.2 G/DL. THE REPEAT RESULT WAS 6.4 G/DL. THE REPEAT RESULT WAS DEEMED CORRECT. THE REAGENT LOT NUMBER IS 65383601. THE EXPIRATION DATE IS 31-OCT-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16197740 ↗MalfunctionNov 22, 2022High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THE INITIAL REPORTER RECEIVED A QUESTIONABLE TOTAL PROTEIN RESULT FOR ONE PATIENT SAMPLE TESTED ON THE COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. ON (B)(6) 2022, THE INITIAL RESULT WAS 97.4 G/L. ON (B)(6) 2022, THE REPEAT RESULT WAS 57.3 G/L. THE REAGENT LOT NUMBER IS 59947001. THE EXPIRATION DATE IS 31-MAR-2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15844240 ↗
FDA clearance and classification
The FDA regulates product code CEK (biuret (colorimetric), total protein) under 21 CFR 862.1635 as a Class II device. 21 CFR 862.1635 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Total Protein Gen.2 08058652190
Has the Total Protein Gen.2 08058652190 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613336121528), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Total Protein Gen.2 08058652190?
The FDA's MAUDE database lists 13 adverse event reports for this device, received between Nov 22, 2022 and Mar 17, 2026. In those reports, the most common specific device problems are “Non Reproducible Results” (9), “Low Test Results” (6) and “High Test Results” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Total Protein Gen.2 08058652190?
None of the reports are classified as deaths, 0 as injuries and 13 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Total Protein Gen.2 08058652190 FDA approved?
Devices of this type (product code CEK) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Roche Diagnostics GmbH.
Who makes the Total Protein Gen.2 08058652190, and is it still sold?
The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand Total Protein Gen.2, model 08058652190). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BIURET (COLORIMETRIC), TOTAL PROTEIN devices
- Atellica CH UCFP 11097543Siemens Healthcare Diagnostics Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN2 recalls18 reports
- ADVIA Chemistry UCFP 11319151Siemens Healthcare Diagnostics Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN2 recalls10 reports
- Atellica™ CH TP 11097604Siemens Healthcare Diagnostics Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN1 recall0 reports
- VITROS 8392292Ortho-Clinical Diagnostics, Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN1 recall0 reports
- Total Protein Gen2 3183734190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN11 reports
- Total Protein Gen2 5171385190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN5 reports
- Atellica CH UPro 11097524Siemens Healthcare Diagnostics Inc. · BIURET (COLORIMETRIC), TOTAL PROTEIN1 report
- Total Protein Gen2 monocromatic 4657586190Roche Diagnostics GmbH · BIURET (COLORIMETRIC), TOTAL PROTEIN1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.